Shielding Measurements For A Proton Therapy Facility

Shielding Measurements For A Proton Therapy Facility. S. Avery, K. P. Risolo, M. Bartels, C. Ainsley, J McDonough, R. L. Maughan University of Pennsylvania. Introduction I/II. Penn’s Roberts Proton Therapy Center 5 Treatment rooms

stacey
Download Presentation

Shielding Measurements For A Proton Therapy Facility

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Shielding Measurements For A Proton Therapy Facility S. Avery, K. P. Risolo, M. Bartels, C. Ainsley, J McDonough, R. L. Maughan University of Pennsylvania

  2. Introduction I/II • Penn’s Roberts Proton Therapy Center • 5 Treatment rooms • Single Scattering, Double Scattering, Uniform Scanning, Modulated Scanning • Double scatting beam line components • First and second scatters • Modulator wheel • Primary jaws • Collimator (MLCs) • Compensator

  3. Introduction II/II • Shielding calculations performed using analytical methods including neutron spectra • Calculations performed for IBA double scattering mode • Max beam current • At any particular energy • Usage factors • Beam direction • Occupancy factors • Workload

  4. Methods and Material I/V

  5. Methods and Materials II/V

  6. Methods and Materials III/V • Both photon sensitive and neutron sensitive detectors required • Total instantaneous dose equivalent obtained from their sum • Neutron detection • Prescila detector for up to 100MeV • Photon detection • Ludum ion chamber

  7. Methods and Materials IV/V • Integrated dose equivalent readings • Neutrak Dosimeters inside a Luxel plastic blister pack • Two types of detectors used: • Those sensitive to 40 keV to 40 MeV with a dose range of 0.2 mSv to 250 mSv • Those sensitive to fast, intermediate, and thermal neutrons with a dose range of 0.1 mSv to 250 mSv

  8. Methods and Materials V/V • Detectors placed at various locations around facility • Primarily around treatment rooms 1 and 5 • Fast neutron detectors left in for one month • Fast, intermediate, and thermal neutrons left in place for an additional month • Control detectors for each type

  9. Results I/V • A modified form of NCRP Report No. 51 equation used to calculate instantaneous dose equivalent index rate • Φp derived from Monte Carlo using GEANT4 • Np is proton beam intensity at the neutron source, related to beam current • Measurements normalized to a beam current of 6 nA emerging from the nozzle

  10. Results II/V • Points outside the cyclotron and ESS showed shielding to be more than adequate with only one reading above background • Reading 4-5 times below the calculated value • Points outside TR1 showed 2 readings below background, with 6 other readings found below their calculated values • Measurements with the dose badges showed only photon irrdatiation

  11. Results III/V • Neutron dose equivalent transmission data from NCRP Report 144 used with the analytic method • Measurements made under identical gantry angle and proton energy conditions for both neutron and gamma ray dose • The instantaneous and integrated two month dose equivalent data allows for calculate of annual dose equivalents

  12. Results IV/V

  13. Results V/V

  14. Conclusions • Shielding in place at Roberts Proton Center is more than adequate • Conservative shielding calculations • Measured vs. Analytic • Film dosimeters questionable • Only saw photon component

  15. Concluding Remarks • Thanks to the authors Dr. Avery, Dr. Maughan, Dr. McDonough, Dr. Ainsley, and Mr. Bartels • Thanks to the HPS