100 likes | 307 Views
MOSFET Dosimeters for 50 kVp Electronic Brachytherapy with the Xoft Axxent Source. Steve Axelrod, PhD Tom Rusch, PhD Xoft, Inc., Fremont, CA www.xoftinc.com. Background and Motivation.
E N D
MOSFET Dosimeters for 50 kVp Electronic Brachytherapy with the Xoft Axxent Source Steve Axelrod, PhD Tom Rusch, PhD Xoft, Inc., Fremont, CA www.xoftinc.com
Background and Motivation • Xoft is introducing a novel miniature x-ray source which is FDA approved for use in HDR breast brachytherapy • Skin dose is increasingly being seen as important in this treatment • MOSFET dosimeters are possibly the most convenient means of measuring skin dose in this situation • The Xoft source operates at 50 kVp with a broad x-ray spectrum • It is inserted into a balloon of ~4cm diameter in the lumpectomy cavity • Response of MOSFETs in this energy range is known to be sensitive to the spectrum, and higher than at mega-voltages.
The Measurements • A Xoft source operating at 50 kVp and 300 μA illuminated an Exradin model A600 ion chamber and two Thomson-Nielsen dosimeters, all at 100 mm • MOSFETs were read out with the Thomson-Nielsen supplied program at 10 second intervals • The ion chamber was connected to a UniDos electrometer and read into a custom LabVIEW program on a standard PC, which also controlled the source • A sequence of aluminum absorbers was placed in the beam path to modify the spectrum, as would occur with tissue in clinical use
Analysis • Data from both sources was exported to a spreadsheet for comparison • Linearity was established • Calibration constants were assigned to the MOSFET readings to optimize agreement with dose as measured by the ion chamber • Variation in the calibration constants with absorber thickness was studied
Cumulative Dose and Dose Rate vs. Time • Unfiltered 50 kVp source Slope is calibration constant, 0.0031 Gy/mV MOSFET noise at 10 seconds ~9%
Calibration Slope Changes with Spectrum • MOSFET sensitivity is lower when beam is hardened by Al absorber • Line is a simple exponential Clinical region > 2 mm (~ 20 mm water)
Calibration Slope Varies Linearly with Absorption Ratio • Absorption ratio is dose ion in chamber with an absorber, divided by dose with no absorber
Conclusion • MOSFET dosimetry can be used to measure skin dose in Xoft Axxent breast brachytherapy • Effects of beam hardening by tissue on the MOSFET calibration should be taken into account • Real time readout in the Thomson-Nielsen system may have future applications