Gilian Air Sampling Pump Field Calibration. A Step-By-Step Guide. Items required. Sampling Pump Collection Media Gilibrator Calibrator. A Field Calibration Consists of Two Parts. The Pre-Calibration (Pre-Cal) is done before the sample is taken
E N D
Gilian Air Sampling Pump Field Calibration A Step-By-Step Guide
Items required • Sampling Pump • Collection Media • Gilibrator Calibrator
A Field Calibration Consists of TwoParts • The Pre-Calibration (Pre-Cal) is done before the sample is taken • The Post-Calibration (Post Cal) is done after the sample is taken
The Pre-Cal • The Pre-Cal is used to set the desired sample flow rate, and to verify it on an air flow calibrator.
The Pre-Cal • Set the Flow Rate • Observe the Flow Rate • Record the Flow Rate • An average of 6 to 10 Gilibrator readings is recommended
Run The Sample • Pump is placed on the belt • Sample media is placed on the lapel (Breathing Zone) and pointed downward • Record the Start Time • Record the End Time
The Post-Cal • The Post-Cal is used to obtain the final flow rate and to verify that the flow rate stayed within +/- 5% of the set flow.
The Post-Cal • Observe the Final Flow Rate • Record the Final Flow Rate • An average of 6 to 10 Gilibrator readings is recommended
Average Flow Rate • The Pre-Cal flow rate and the Post-Cal flow rate may be averaged to obtain an average flow rate for the sample.
Pre-Cal Flow Rate + Post Cal Flow Rate = Avg Flow Rate 2 • Determining the Average Flow Rate
Cyclone Calibration • The Gilian 10mm nylon cyclone has only an exit fitting and must be placed inside a calibration jar to accept the Gilibrator.
Low Flow Modules • Calibrating with low flow modules
Constant Flow Modules • Set pump flow to 1500cc/min • Turn module switch on • Connect sample tubing to pump • Adjust desired flow rate at module • Read flow rate at Gilibrator
Multi-Flow Modules • Set pump to 1500 cc/min • Turn module switch on • Connect sample tubing to pump • Adjust flow rate at holder • Adjust each tube twice • Read the flow rates
Further Information • For detailed information on any of the Gilian products, please consult the product operation manuals. • Manuals may be accessed through our web site at www.sensidyne.com