1 / 13

Georg Gassner Robert Ruland, SLAC February, 2008

Georg Gassner Robert Ruland, SLAC February, 2008. Laser Tracker Test Facility at SLAC. Introduction Rotary Table Calibration Laser Tracker Investigation ADM Tests Conclusion. Introduction.

taffy
Download Presentation

Georg Gassner Robert Ruland, SLAC February, 2008

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Georg Gassner Robert Ruland, SLACFebruary, 2008 Laser Tracker Test Facility at SLAC • Introduction • Rotary Table Calibration • Laser Tracker Investigation • ADM Tests • Conclusion 1

  2. Introduction • Particle Physics experiments at SLAC require high accuracy positioning, e. g. 100 µm over a distance of 140 m or 25 µm in a 7 x 10 x 3 meter room • Realization -> Laser Tracker Measurements • Rotary calibration table (Kugler GmbH) providing an accuracy of better than 0.2 arcsec • Interferometer bench 2

  3. Rotary Table • Renishaw Signum RESM angle encoder system • 1 arcsec graduation accuracy • 0.01 arcsec resolution 3

  4. Calibration of the rotary table (1) • Rosette technique • calibrate precision polygon prisms • sum of the angles measured must result in 360 degrees • Moeller-Wedel electronic autocollimator (0.1arcsec accuracy, 0.05arcsec reproducibility) 4

  5. Calibration of the rotary table (2) • Derived Technique • set the fixture with the mirrors in a way that mirror ‘a’ is in line with autocollimator ‘a’ • rotate the table together with the fixture until mirror ‘b’ is in line with autocollimator ‘b’ 5

  6. Calibration of the rotary table (3) • Absolute value of an angle between two positions of the rotary table can be determined with ±0.2 arcsec std. 6

  7. Laser Tracker Horizontal Angle Measurement System Investigation • Test Setup • mirror as a target for laser tracker • parallelism of the two rotation axes • axial displacement of the rotation axes Mirror Faro Laser Tracker Rotary table 7

  8. Comparison SMR vs. SMM • Test of Mirror (SMM) and Retroreflector (SMR) • Only small deviations can be found 8

  9. Calibration Results Two calibration runs with Faro Laser Tracker Xi Two calibration runs with Faro Laser Tracker Xi V2 9

  10. Carriage Interferometer Dd Horizontal Bench Laser Tracker ADM Calibration on Interferometer Bench • Calibration of the scale factor of the ADM in the field. • Tests of the general performance and tests to detect malfunctions in the laboratory. 10

  11. Calibration Results Calibration run with Faro Laser Tracker Xi Calibration runs with Faro Laser Tracker Xi V2 11

  12. Conclusion • Rotary table - absolute angle accuracy < 0.2 arcsec • Rotary table to test any horizontal angle measurement instrument • Faro laser trackers angle measurement system deviations stable and correctible • With corrections, angle accuracies of 1 arcsec are possible. • Test of the ADM showed its usability as a replacement for the interferometer mode 12

  13. End of Presentation 13

More Related