1 / 29

Boron-GEM3 – Further measurements

Boron-GEM3 – Further measurements. Dorothea Pfeiffer. 19 .03.2014. Content. X-ray spectrum of 241 AmBe source Neutron spectrum measured with Boron -GEM3 with 8 mm drift Mesaurements with SRS What’s next?. Measurements with AMPTEK XR- 100CR Si -PIN X-Ray detector. 5.9 keV.

tahlia
Download Presentation

Boron-GEM3 – Further measurements

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Boron-GEM3 – Further measurements Dorothea Pfeiffer 19.03.2014

  2. Content • X-ray spectrum of 241 AmBe source • Neutron spectrum measured with Boron-GEM3 with 8 mm drift • Mesaurements with SRS • What’s next? Dorothea Pfeiffer

  3. Measurements with AMPTEK XR-100CRSi-PIN X-Ray detector 5.9 keV 6.49 keV

  4. Efficiency of AMPTEK XR-100CRSi-PIN X-Ray detector (500 um Si, 1 mm Be) Dorothea Pfeiffer

  5. 241 Am example spectrum AMPTEK XR-100CRSi-PIN X-Ray detector

  6. Measurements with AMPTEK XR-100CRSi-PIN X-Ray detector (efficiency corrected) 13.95 keV 17.74 keV 59.54 keV

  7. Triple GEM with Boron-10 converter (variant 2) Drift cathode: 18 um Al foil coated with 1.3 um10B4C one side 8mm 2 mm 2 mm 2 mm Dorothea Pfeiffer

  8. Boron GEM3: 18 um Al foilwith 1.3 um 10B4C y coordinate (APV chips not connected) x coordinate (APV chips not connected) Dorothea Pfeiffer

  9. 55 Fespectrum ADC channel 50 at preamp gain 10 = about channel 25 at preamp gain 5 ?

  10. 241 AmBespectrum, driftgap 8 mm ADC channel 2800(~ 672 keV)Li peak 840 keV) ADC channel 5000 (~ 1.2 MeV)alpha peak 1.47 MeV

  11. SpectrumfromBoron Blade DetectoratILL Looks similar to our spectrum, also only one peak (not two peaks for alpha and Li ion) Dorothea Pfeiffer

  12. Boron GEM3: twosectors:withandwithout 18 um Al foilwith 1.3 um 10B4C 1 cm of lead covering whole detector Normal cathode on top 1 2 y Sector 1: 1 cm lead Sector 2: 10B4C converter and 1 cm lead x Dorothea Pfeiffer

  13. Boron GEM3: twosectors – 55 Fesource Looks similar to our spectrum, also only one peak (not two peaks for alpha and Li ion) 1 2

  14. Boron GEM3: twosectors – 55 Fesource Looks similar to our spectrum, also only one peak (not two peaks for alpha and Li ion) 1 2

  15. Boron GEM3: twosectors – 55 Fesource 1: x 1: y Spectra in sector 2 have 50% of the hits 2: x 2: y

  16. Boron GEM3: twosectors – 55 Fesource 1: x 1: y Cut in sector 2, x=4: Kapton tape? But in y? 2: x 2: y

  17. Boron GEM3: twosectors – 241AmBe source • Due to limited dynamic range of the APV, measurement had to be carried out with very low gain of about 15 (I=565 uA, U=3250 V) • Even then saturation visible saturation Dorothea Pfeiffer

  18. Boron GEM3: twosectors – 241AmBe source Looks similar to our spectrum, also only one peak (not two peaks for alpha and Li ion) 1 2

  19. Boron GEM3: twosectors – 241AmBe source Looks similar to our spectrum, also only one peak (not two peaks for alpha and Li ion) 1 2

  20. Boron GEM3: twosectors – 241AmBe source Sector1 + 2 Looks similar to our spectrum, also only one peak (not two peaks for alpha and Li ion)

  21. Boron GEM3: twosectors – 241AmBe source sector1 Looks similar to our spectrum, also only one peak (not two peaks for alpha and Li ion)

  22. Boron GEM3: twosectors – 241AmBe source sector2 Looks similar to our spectrum, also only one peak (not two peaks for alpha and Li ion)

  23. Boron GEM3: twosectors – 241 AmBesource 1: x 1: y 2: x 2: y

  24. Boron GEM3: twosectors – 241 AmBesource 1: x 1: y Spectra in sector 2 have 2% of the hits compared to sector 1 2: x 2: y

  25. Boron GEM3: twosectors – 241 AmBesource Calculation of “neutron spectra”: sector 2 – sector 1 6Li peak and alpha peak? Energy calibration with APVs and SRS? Dorothea Pfeiffer

  26. Boron GEM3: twosectors – 241 AmBesource 1: x 1: y Source not perfectly centered on detector 2: x 2: y

  27. Boron GEM3: Foursectors Sector 1: free (normal GEM) Sector 2: 1 cm lead Sector 3: 10B4C converter 1 cm of lead 1 3 Sector 4: 10B4C converter and 1 cm lead y 2 4 x Dorothea Pfeiffer

  28. Boron GEM3: Foursectors 2 sector measurements 4 sector measurements Problem with one of the APVs apparent at the end of the measurement, but also variability very large in first part of dataset (event number < 150000) => repeat measurements Dorothea Pfeiffer

  29. TO DO • Measure X-ray spectrum of 241 AmBe source with lead in front • Measure gamma spectrum of 241 AmBe with Sodium Iodide chrystal • Repeat 4 sector measurement with SRS • Organize Gadolinium samples Dorothea Pfeiffer