1 / 28

ACS Optical Instability

ACS Optical Instability. Pam Sullivan Updated 5 July 99. Instability Introduction. Small Temperature Changes in ACS Components Cause Apparent Motion of All Detectors wrt Internal or External Optical Sources Specification is 10 mas peak-peak over 2 Orbits

thaddeus
Download Presentation

ACS Optical Instability

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. ACS Optical Instability Pam Sullivan Updated 5 July 99

  2. Instability Introduction • Small Temperature Changes in ACS Components Cause Apparent Motion of All Detectors wrt Internal or External Optical Sources • Specification is 10 mas peak-peak over 2 Orbits • Several Distinct Phenomena Identified: • “Warm-up” Transients • ASCS Interface Plate Related Transients • “Rolling” Transients Associated with Room Temperature • Shifts in Coronagraph Location with Orientation Change and with Different Holding Fixture

  3. Coronagraph Motion w/Orientation & Holding Fixture Changes COR. SPOT LOCATION DATE CONFIGURATION X CENTER Y CENTER FOC INN OUT 31 Mar TBF/Diamond 2465 57814 11939 14 Apr Dolly/Square 488 543 2345 57810 11946 2 Jun SIFIG/Aperture Down 453 603 2568 59046 10675 3 Jun Dolly/Square 480 566 2568 59049 10673 14 Jun Dolly/Square 477 566 2569 59046 10673

  4. Instability Theories • Troubleshooting has Confirmed that Optical Bench is not Isolated from Enclosure (XYZ Fittings should Isolate Bench) • This Anomalous Behavior May Cause/Contribute to All Phenomena through the Following Mechanisms: • Enclosure Shape Change Transmits Loads to Optical Bench • TBF/Dolly CTE Effects Load Enclosure which Load Bench • Optical Bench CTE Expansion Results in Bench Bending • Other Effects Which May be Contributing to Instability: • Heat Pipe CTE Expansion Imparts Loads to Optical Bench • Localized CTE Effects in Detector Housings & Other Components Expanding/Contracting CCD Heat Pipes Impart Load to CCDs and Optical Bench

  5. Introduction to Data • Organized Chronologically • All Data Collected at Ambient Temperature and Pressure • Optical Source was either: • RAS/Cal Point Source (External to ACS) • Coronagraphic Spot (Internal to ACS) • Motion is that of the Optical Source Position within the CCD Field of View • Data is Presented as Delta X (Diamonds) and Y (Triangles) Pixels from Arbitrary Starting Point • HRC Pixels are 21 um • SBC Pixels are 25 um • WFC Pixels are 15 um

  6. March 22: Initial Identification of Anomaly on HRC • Configuration: • Mounting: ACS in TBF/Diamond Configuration, in Chamber 225 • Cooling: ASCS Simulator with GN2 • Optical Source: HRC: RAS/Cal point source • Starting Temp: Warm, Instrument had been On for Several Hours • Results: • Optical motion is correlated with cool-down of ASCS simulator at 1730 • Change of 7C at ASCS caused ~ 1 pixel motion • Note that ACS Interface Plate and CCD Housing cools with ASCS • Note that Optical Bench temperature is rising slightly through test • No apparent motion is seen during ASCS warm-up • Implies anomaly is not a simple CTE expansion/contraction effect • Direction of HRC motion is: • Parallel to floor, as viewed looking into aperture • Side to side at the CCD, as viewed looking down on instrument

  7. March 30: Verify Reproduce-ability & Correlation w/WFC • Configuration: • Mounting: ACS in TBF/Diamond Configuration, in SSDIF • Cooling: Chiller Plate • Optical Source: WFC: RAS/Cal point source; HRC: Coronagraph • Starting Temp: Ambient, at Start of Day • Results: • Little to no motion seen during ACS “warm-up” • Chillers turned on to 14C causes Interface Plate temperature change with causes motion in both WFC and HRC • In this test, a 2C delta causes ~ 4 pixels of motion • HRC moves same direction as Mar 22 data set (parallel to floor) • WFC moves: • Perpendicular to floor (up and down) as viewed into aperture • Fore to Aft at CCD, as viewed looking down on instrument

  8. April 13 AM: Test Effect when ACS is Suspended • Configuration: • Mounting: ACS Suspended by Crane, in TBF w/B-Latch Disengaged • Cooling: Chillers • Optical Source: HRC: Coronagraph • Starting Temp: Ambient, at Start of Day • Results: • Small motion seen during warm-up • Turning chillers on to 14C causes motion • Same direction as when latched into TBF • Magnitude Reduced by 5X

  9. April 13 PM: Test Effect when ACS Held in Dolly/Diamond • Configuration: • Mounting: ACS in Handling Dolly in Diamond Orientation • Cooling: Chillers • Optical Source: HRC: Coronagraph • Starting Temp: Ambient, ACS off for few hours during relocation to dolly • Results: • Small motion seen during instrument warm-up • Chillers on to 14C causes noticeable transient, but otherwise does not seem to change warm-up trend • Chillers off does not change the warm-up trend • HRC motion is this test is different: downward as looking into aperture

  10. April 14: Test Effect when ACS Held in Dolly/Square • Configuration: • Mounting: ACS in Handling Dolly in Square Orientation • Cooling: Chillers • Optical Source: HRC: Coronagraph • Starting Temp: Ambient, at start of day • Results: • Square Orientation has radiator heat pipes parallel to floor, and thus allows them to operate; square is also the most heat-transfer-effective for the CCD heat pipes • Turn on HRC CCD and TECs only: ~0.5 pixel motion caused during warm-up • Turn on chillers: no apparent change in warm-up motion trend • Turn on WFC CCDs and TECs: causes small slope increase in warm-up trend • Turning off chillers changed the direction of the motion trend

  11. April 15 AM: Compare Dolly/Diamond Motion when HPs Disconnected • Configuration: • ACS: CCD Heat Pipes Disconnected at IF Plate • Mounting: ACS in Handling Dolly in Diamond Orientation • Cooling: Chillers • Optical Source: HRC: Coronagraph • Starting Temp: Ambient, at start of day • Results: • Note: data noisy due to operating CCDs at room temperature (TEC off) • Small motion ~0.5 pixel seen during warm-up • Motion perpendicular to that seen in dolly/diamond with HP connected • Chillers turned on to 14C caused no correlated motion • Note: large jump at 17.3 most likely due to optical source change

  12. April 15 PM: Compare Dolly/Square Motion when HPs Disconnected • Configuration: • ACS: CCD Heat Pipes Disconnected at IF Plate • Mounting: ACS in Handling Dolly in Square Orientation • Cooling: Chillers • Optical Source: HRC: Coronagraph • Starting Temp: Warm after several hours of Operation • Results: • Note: data noisy due to operating CCDs at room temperature (TEC off) • Start warm with chillers on at 14C • No apparent motion due to turning chillers off • Turning HRC TEC on at 21.9 causes a large jump due to image quality improvement (not a thermal-mechanical effect) • Correlation between motion & TEC on/off state identified (due to Housing temp?) • Turning HRC TEC off at 22.7 causes small ~0.3 pixel motion • Turning HRC TEC on at 23.0 causes TEC-off motion to reverse

  13. April 16: Compare TBF/Diamond Motion when HPs Disconnected • Configuration: • ACS: CCD Heat Pipes Disconnected at IF Plate • Mounting: ACS in TBF in Diamond • Cooling: Chillers • Optical Source: HRC: Coronagraph • Starting Temp: Ambient, at start of day • Results: • Warm-up effect of ~1.5 pixels seen • Direction of motion similar to that with heat pipes connected • Chillers-on seemed to restart warm-up trend which had stalled for ~15 minutes • Chillers off had no apparent effect • Motion correlated with TEC on/TEC off repeated

  14. May 7: Measure Thermal Stability Effect on SBC • Configuration: • ACS: Fully Assembled • Mounting: ACS in TBF in Diamond • Cooling: Chillers • Optical Source: SBC: RAS/Cal Point Source in N2 Purge • Starting Temp: Close to Ambient after being off for Lunch • Results: • Warm-up effect of ~1 pixel seen • Chillers-on caused motion of 4-5 pixels • Motion is different than HRC effect • Indicates that effect is not solely M1 or M2 motion • Chillers off reversed the chillers-on effect • WFC TECs off/on had no effect on warm-up or chiller trend

  15. June 2: Measure Stability with Aperture Down/Y-Fitting Free • Configuration: • ACS: Fully Assembled • Mounting: ACS attached to SIFIG by GSE Lifting Points ACS Clocked 10 deg off Vertical, Aperture Facing Floor Note: This Configuration Frees Y-Fitting but also Precludes HRC Heat Pipes from Transferring Heat to ASCS Plate • Cooling: Chillers • Optical Source: HRC Coronagraph Backlight • Add’l Sensors: Displacement Sensors Measuring Latch Motion & Theodolites Measuring In Plane Motion of IF Plate and Radiator Panel Relative to +P2 Enclosure Panel • Starting Temp: Ambient at Start of Day • Results: • Warm-up effect barely present at ~0.1 pixel • Chillers-on Caused no Perceptible Change in Warm-up Trend • HRC TECs Off (to Prevent Safing due to Overtemp Housing) Caused Drift Effect which is Not Related to Stability Effect • No Measurable Motion seen by Displacement or Angular Sensors (sensitivity is .0001 inch)

  16. June 3: Measure Stability with Radiator Panel Unbolted • Configuration: • ACS: +P2 Radiator Panel Bolts Loosened 1 Turn • Mounting: In Dolly, in Square with Radiator Facing Up & Parallel to Floor • Cooling: Chillers • Optical Source: HRC Coronagraph Backlight • Starting Temp: Ambient at Start of Day • Results: • Warm-up Effect of 1 pixel seen • Chillers-On & Off had No Effect • HRC Set Point Change from -90C to -67C Caused Noticeable Slope Change • New “Rolling” Effect in X-Axis only Started without Apparent Cause ~7 Hours Into Test and Continued through Remaining 8 Hours • Also Noticed an Unexplained Increase in Noise in Spot Location Data as Test Progressed • There was No Corresponding Increase in Image Noise

  17. June 4: Measure Stability with Radiator Panel Rebolted • Configuration: • ACS: Assembled with Panel Bolts Re-Tightened • Mounting: In Dolly, in Square with Radiator Facing Up & Parallel to Floor • Cooling: Chillers • Optical Source: HRC Coronagraph Backlight • Starting Temp: Ambient at Start of Day • Results: • Executed Test Sequence Identical to Previous Day for Apples-Apples Comparison • Resulting Pixel Motion Nearly Identical to Previous Day, Except: • Warm-up Effect Barely Present • Possibly because ACS Started 1C Warmer? • Possibly because Enclosure is More Resistant to Shape Changes with Panel? • HRC Set Point Change Caused Slope Change • This Effect Probably Occurred Yesterday but was Masked by Warm-up Effect • “Rolling” Effect Larger Magnitude • Rolling Appears Correlated with JIN2PANT Enclosure Panel Temperature • Except for Rolling, Stability is Better with Enclosure Panel Bolts Torqued

  18. June 7: Test “Rolling” Effect Theories • Configuration: • ACS: Assembled • Mounting: In Dolly, in Square with Radiator Facing Up & Parallel to Floor • Cooling: Chillers • Optical Source: HRC Coronagraph Backlight • Starting Temp: Ambient at Start of Day • Results: • Rolling was Correlated with Enclosure Panel Temp • Heat from 650W Lamp Used to Change Enclosure Panel Temperature • Applying Heat to O3 and O2 (Separately) Caused Apparent Motion in Y-Axis Only • Previous “Rolling” was in X-Axis Only • Imaging Interval Varied (30 min imaging, 5 min break, 30 min imaging, 10 minute break, 30 min imaging, 15 min break) to Determine if Breaks Causes Cooling • Conclusion: • Rolling is not Caused by Enclosure Panel Heating nor Imaging Operations

  19. June 9: Relieve Heat Pipe Installation Stress & Loosen ASCS Bolts • Configuration: • ACS: ASCS Plate Bolts to Radiator Panel Loosened • Mounting: In Dolly, in Square with Radiator Facing Up & Parallel to Floor • Cooling: Chillers • Optical Source: HRC Coronagraph Backlight • Starting Temp: Ambient at Start of Day • Results: • No “Warm-up” Effect Seen • No “Chillers-On” or Off Effect Seen • Changing HRC Set Point to -85C Apparently Caused Motion in X • Changing HRC Set Point back to -67C Produced Motion in Y, but did Not Reverse Y Trend

  20. June 10: Re-Bolt ASCS Interface Plate • Configuration: • ACS: Fully Assembled • Mounting: In Dolly, in Square with Radiator Facing Up & Parallel to Floor • Cooling: Chillers • Optical Source: HRC Coronagraph Backlight • Starting Temp: Ambient at Start of Day • Results: • No “Warm-up” Effect Seen • No “Chillers-On” Effect Seen • “Chillers Off” Caused ~0.3 pix/hr Motion in Both Axes • Changing HRC Set Point had No Effect • Conclusion: • No Significant Difference between ASCS IF Plate Loose vs Bolted

  21. June 11: Monitor Stability w/o Changing Detector or Chiller State • Configuration: • ACS: Fully Assembled • Mounting: In Dolly, in Square with Radiator Facing Up & Parallel to Floor • Cooling: Chillers • Optical Source: HRC Coronagraph Backlight • Starting Temp: Ambient at Start of Day • Results: • “Warmup” and/or Chiller Effect seen for ~7 Hours • Pixel Location Stabilizes for ~1 Hours at End of Data Collection

  22. June 14: Monitor Stability w/Room Temperature Change • Configuration: • ACS: Fully Assembled • Mounting: In Dolly, in Square with Radiator Facing Up & Parallel to Floor • Cooling: Chillers • Optical Source: HRC Coronagraph Backlight • Starting Temp: Ambient at Start of Day • Results: • “Warm-up” and/or Chiller Effect seen for ~4 Hours at Start of Day • Stability after 1700GMT Correlates with JO3APANT Enclosure Panel Temperature & Enclosure Panel Correlates Grossly with Room Temperature • Conclusion • Stability is Correlated with Room Temperature • Potential Mechanism is Dolly Length Change with Temperature Loads Enclosure

  23. June 17: Compression Test to Verify Presence of Y-Fitting “Stiction” • Configuration: • ACS: Fully Assembled; OFF for this Test to Eliminate Thermal Effects • Mounting: In TBF/Diamond • Cooling: Chillers • Optical Source: None (ACS Off) • Starting Temp: Ambient throughout Test (ACS Off) • Add’l Sensors: LVDTs to Monitor Motion of Bench Relative to Enclosure & Theodolites to Monitor TBF and ACS Optical Cube Motion • Results: • TBF B-Latch Used to Compress Enclosure from 0 to 800lb in 50lb Increments • Increasing Load Compresses Both Enclosure & Optical Bench by ~25 mil • ACS Optical Cubes’ Motion of ~6 to 18 as Indicates Bench is Bending w/Increasing Load • Conclusion: • Optical Bench Moves with Enclosure • Y-Fitting is not Behaving per Design

  24. June 18: Repeat Compression Test for Repeatability • Configuration: • ACS: Fully Assembled; OFF for this Test to Eliminate Thermal Effects • Mounting: In TBF/Diamond • Cooling: Chillers • Optical Source: None (ACS Off) • Starting Temp: Ambient throughout Test (ACS Off) • Add’l Sensors: LVDTs to Monitor Motion of Bench Relative to Enclosure & Theodolites to Monitor TBF and ACS Optical Cube Motion • Results: • Decompression Reverses Compression Effect with Little or No Hysteresis • Compression and Decompression Effects are Repeatable • Conclusion: • Optical Bench Moves with Enclosure • Y-Fitting is not Behaving per Design

  25. June 21 AM: Compression Test with ACS On to Correlate Spot Motion • Configuration: • ACS: Fully Assembled • Mounting: In TBF/Diamond • Cooling: Chillers • Optical Source: HRC Coronagraph Spot with Backlight • Starting Temp: Ambient at Start of Day • Add’l Sensors: LVDTs to Monitor Motion of Bench Relative to Enclosure & Theodolites to Monitor TBF and ACS Optical Cube Motion • Results: • TBF B-Latch Used to Compress/Decompress Enclosure • Both Enclosure & Bench Move with Changing Load • Coronagraph Moves ~7 Pixels in Each Axis with Changing Load • Compression Effect is Repeatable; Decompression is Repeatable • Compression vs Decompression Motion Shows Hysteresis • Conclusion: • Coronagraph Motion is Correlated with Enclosure/Bench Motion and is thus Caused (at least in part) by Stuck Y-Fitting

  26. June 21 PM: Compression Test with ACS On to Correlate Spot Motion • Configuration: • ACS: Fully Assembled • Mounting: In TBF/Diamond • Cooling: Chillers • Optical Source: SBC Coronagraph Spot with Backlight • Starting Temp: Warm after Morning Operations • Add’l Sensors: LVDTs to Monitor Motion of Bench Relative to Enclosure & Theodolites to Monitor TBF and ACS Optical Cube Motion • Results: • TBF B-Latch Used to Compress/Decompress Enclosure • Both Enclosure & Bench Move with Changing Load • Coronagraph Moves with Changing Load • Compression vs Decompression Motion Show Hysteresis • SBC Motion is Nearly Identical to HRC Magnitude and Direction • Conclusion: • Coronagraph Motion is Correlated with Enclosure/Bench Motion and is thus Caused (at least in part) by Stuck Y-Fitting • Similarity of SBC & HRC Motion Indicate Effect is Predominately Motion of the M2 Mirror

  27. June 23: Aborted “Baseline” & TBF Temperature Change Test • Configuration: • ACS: Fully Assembled • Mounting: In TBF/Diamond with New Linear Bearing B-Latch • Cooling: Chillers • Optical Source: HRC Coronagraph Spot with Backlight; WFC RAS/Cal • Starting Temp: Ambient at Start of Day • Add’l Sensors: LVDTs to Monitor Motion of Bench Relative to Enclosure • Results: • Had Planned to Run Long Duration Baseline for Comparison to RASHOMS Results • WFC Over-Temp Forced WFC TECs to be Turned Off; Baseline Aborted • Temp Change of TBF Bottom Support Post (20 to 30 to 20C) Caused WFC & HRC Motion • Temp Change of +V2 Support Post (20 to 30 to 20C) Caused WFC & HRC Motion • During Temp Changes B-Latch Load Cell Measured Changing Load (~30lb) and LVDTs Measured Motion of Enclosure & Bench Relative to TBF • Conclusion: • TBF Temp Changes Causes TBF Length Change Causes Load Change Causes Enclosure/Bench Motion • Y-Fitting “Stiction” Prevents Verification of Proper B-Latch Motion

  28. June 24: “Baseline” Stability in TBF for Comparison to RASHOMS • Configuration: • ACS: Fully Assembled • Mounting: In TBF/Diamond with New Linear Bearing B-Latch • Cooling: Chillers • Optical Source: HRC Coronagraph Spot with Backlight; WFC RAS/Cal • Starting Temp: Ambient at Start of Day • Add’l Sensors: LVDTs to Monitor Motion of Bench Relative to Enclosure • Test Sequence for Baseline: • T=0 (1300GMT) ACS On/WFC TECs On • T=4 Hr HRC TEC On • T=7 Hr Chillers On • T=11 Hr Chillers Off • T=14 Hr Test End • Results: • “Warmup” and Chillers On Effects Present at ~0.5 pixels &Chillers Off Effect Present at ~1 pixel • TBF Cube Motion of ~8 as and AC1 Motion of ~3 as Observed During Chillers On (Optical Cube Data was Taken Only During the Chillers On Segment from1540-1800) • LVDTs Show Small TBF Expansion Correlated w/Chillers On & Off • Cube & LVDT DataIndicates Chiller Effect is (Partially?) Caused by TBF Loading the Enclosure

More Related