1 / 23

Cobas 6000

생화학파트 신현진. Cobas 6000. Index. 도입 배경 Cobas 6000 – Module Overview Cobas 6000 의 검사항목 Performance Evaluation of cobas 6000 PAM 의 재배치도 Cobas 6000 의 장단점 Summary. 1) E-170 장비의 노후 장비 교체의 필요성 2) 24 시간 검사 요구 항목의 증가 임상의 요구 수용으로 내부 고객 만족도 향상 기대 3) 검사 건수의 지속적인 증가 & 신규검사 종목 추가로 업무량 증가

tjuanita
Download Presentation

Cobas 6000

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. 생화학파트 신현진 Cobas 6000

  2. Index • 도입 배경 • Cobas 6000– Module Overview • Cobas 6000의 검사항목 • Performance Evaluation of cobas 6000 • PAM의재배치도 • Cobas 6000의장단점 • Summary

  3. 1) E-170장비의 노후 장비 교체의 필요성 2) 24시간 검사 요구 항목의 증가임상의 요구 수용으로 내부 고객 만족도 향상 기대 3) 검사건수의 지속적인 증가 & 신규검사 종목 추가로 업무량 증가 인력 충원 필요 4) 자동화 검사 system의 재정비 4호기 증설로 인한 인력 부족 해소 1. Cobas 6000의 도입배경

  4. 2. cobas 6000 –module overview sampler c501 e601

  5. Product information of sampler • CU 150: • Up to 120 racks/hr • Loading: 150 samples/30 racks • Unloading: 150 samples/30 racks • Rack Rotor: • 20 rack positions • Definable positions for STAT • and auto QC

  6. Product information of c501 • Up to 600 photometric tests/h. • Up to 600 ISE tests/h. (200 samples) Hitachi 7600 검사 • Analysing time: 10 min • Sample pipetting volumes :1.5 µl - 35 µl • 60 positions for cobas c packs

  7. Product information of c501 A E B C F G D H A: Rack sampling positionB: Sampling area C: Reaction diskD: ISE area with ISE probe E: Reagent areaF: Second Reagent probeG: First Reagent probeH: Reagent cassette loading port

  8. ◈ 검사 도중 시약 교체 가능 불필요한 maintenance방지 ◈Ultra Sonic Mixer 검체 간의 carry over 방지 Product information of c501

  9. Product information of e601 • Up to 170 tests/h • Analysing time: 18 and 27 min • Sample pipetting volumes : 15 µl - 50 µl • 25 positions for cobas e packs • Electro-Chemiluminescence (ECL) Technology for heterogeneous Immunoassays

  10. Product information of e601 A: Prewash station B: Sample probe C: Bead mixer D: Assay tip station with rejection station E: Vortex mixing station F: Reagent probe G: Reagent disk with 25 positions

  11. Cobas 6000 rackflow

  12. 3. 검사 항목 cobas 6000 – c501 차후 24시간 검사 예정 정규시간 내 검사 (월~금) Primidone

  13. 검사 항목 cobas 6000 – e601 24시간 검사 정규시간 내 검사 (월~금) 요일별 검사(화,목) Osteocalcin

  14. 이전 예정 검사 항목 cobas 6000 – c501 정규시간 내 검사 예정 (월~금) 차후 24시간 검사 예정 (3월말 예정)

  15. 4. Precision test of c501

  16. Precision test of e601

  17. Comparison test of c501

  18. Comparison test of e601

  19. 5. PAM의 재배치도 Cobas6000 Cobas6000

  20. 6. Cobas 6000 의 장점 ◈ 검사 장비 일원화를 통한 효율성 증대 E-170 + Integra 800 + BNⅡ Cobas 6000으로 통합 ◈ 사용자에게 친근한 Operating 환경 제공 현재 사용중인 Hitachi 7600과 같은 User Interface를 사용하여 손쉽게 장비에 적응할 수 있음. ◈ Cobas link PC 활용 가능 e-library가 포함된 cobas link PC을 통해 web상으로 시약, calibrators, controls에 관한 최신 정보에 직접 접속 및 다운 로드 가능.

  21. Cobas 6000의 단점 ◈ Maintenance 상의 어려움 시약교체 또는 Error로 인해 c501이나 e601 장비 중 하나의 장비에 대한 Maintenance가 시행되는 동안, cobas6000전체 장비가 stop되므로 검사 불가능. ◈ Start up time C501의 경우, Integra 800보다 stand by Operation 으로 넘어가는 시간이 길어짐. ◈ Sampler 사용의 불편함 E-170에 비해 unloading 시 사용되는 tray의 길이가 짧아져 다소 불편함.

  22. Cobas 6000의 도입에 따른 업무변화 야간근무자 업무가중

  23. 7. Summary 업무 효율성 향상

More Related