THEMIS INSTRUMENT SUITE • FM4/FM5 PRE-SHIP REVIEW • FGM • Ellen Taylor for FGM Team
Overview • Personnel • FGM Flight Hardware Status • FGM Open Items • FGM Backup Slides • Changes since FM4/5 PER • Environmental Qualification Summary • Performance Testing Summary • MRD Verification Summary
FGM Personnel • Personnel • TU Braunschweig and IWF Graz • Wolfgang Baumjohann • Karl-Heinz Glassmeier (Head) • Ulrich Auster (FGM TM) • Werner Magnes (FGM CoTM) • Olaf Hillenmaier • Ronald Kroth • Markus Wiedemann
FGM Status • Hardware Status • FGM calibration verified all requirements • 6 Flight Units of FGM Sensor (FGS) and FGM Electronics (FGE) have been delivered to UCB • Design verification through IDPU system occurs periodically during Instrument Suite Comprehensive Performance Tests. FGM sensor is quiet magnetic environment in THEMIS Calibration Unit (TCU) for these tests.
FGM Open Items • FM1 Open Item: None • FGM FM1 Sensor (SN 04) was replaced with Flight Spare (SN 12) • Flight Spare sensor has completed the additional 6 cycles of Thermal Vacuum testing that was missed during Suite testing • FM2/FM3 Open Items: None • FM2 (SN01) and FM3 (SN02) FGM Sensors have been integrated to the Instrument Suite, gone through Suite Environmental Test, and are ready for Integration with the Probe • FM4/FM5 Open Items: None • FM4 (SN03) and FM5 (SN05) FGM Sensors have been integrated to the Instrument Suite and are ready for Suite Environmental Test
BACK-UP: • THEMIS FM1 FGM PSR/PER Presentation • Changes since FM1 PER • Environmental Qualification Summary • Performance Testing Summary • MRD Verification Summary
FGS qualification • Vibration Specs. • Acceptance performed in Berlin • Sine wave: up to 6.4g @ 50Hz • Random noise: 11.7g RMS • TV-cycle • Performed in Graz • Two cycles between -120°C and 90°C • Parameter check (before, between and after tests) • Resistance of all coils / Sensor output signals • Coil constants and mechanical alignment of feedback coil system
Test performed with SN4 • digital functional test • check of board population, FPGA and current sources • Nov. 01. Berlin, THM_FGM_F6StartupTest.doc • analogue adjustment of FGS and FGE • Test of excitation,input filter, phase adjustment, sensitivity and noise • Nov. 05. Berlin, THM_FGM_F4AnalogTest.doc • Parameters vs. electronics temperature • Test of offset, sensitivity, inrush current, phase in dependency on e-temp. • Nov. 06. Braunschweig, THM_FGM_Cal_F4e-temp.doc • Parameters vs. sensor temperature • Test of offset, sensitivity and noise in dependency on sensor temperature • Nov. 14. Graz, THM_FGM_Cal_F4s-temp.doc
Test performed with SN4 • test of non-linearity • Test of DAC non linearity and filter characteristics • Nov. 18. Graz, THM_FGM_Cal_F4&F6linearity.doc • Earth field registration • Test of resolution and stability by comparison of F4 and F6 measuring Earth field variations • Nov. 19. Jeserigerhütten, THM_FGM_Cal_F4&F6RegJH.doc • determination of sensor axis direction • Direction of magnetic axis determined with respect to mechanical ref. frame • Nov. 22. Jeserigerhütten, THM_FGM_Cal_F4&F6AlignmentJH.doc • determination of scale value and orthogonality • Scale values and orthogonality check vs. sensor temp. in coil system • Nov. 24 Magnetsrode, THM_FGM_Cal_F4&F6_MR.doc
MRD requirements • The absolute stability of the FGM shall be less than 1nT Dependency on sensor temp. 2nT/170°C !!! Dependency on electronics temp. without feedback circuitry: 1nT/80°C with feedback circuitry: 10nT/80°C& (noisy at 60°C) • Action: • AD648 identified as not ideal, replaced by RH1013 • Temperature drift at maximum values of data sheet • noise at 60°C not specified in data sheet
FGM Parameter • The relative stability of the FGM shall be less than 0.2nT/12hrs
FGM Parameter • The FGM noise level @ 1Hz shall be less than 0.03nT/sqrt(Hz) 0.03nT/Sqrt(Hz) 0.01nT/Sqrt(Hz)
FGM Parameter • The FGM digital resolution shall be less than 0.1nT • The FGM science range shall exceed 0-1000nT Test shows Earth field variation with a resolution better 0.1nT @ a back ground field of 12000nT
FGM SN4 Test Report • FGE modification and re-tests done in Berkeley • Exchange of amplifiers + Analogue checkout • AD648 replaced by RH1013 • No noise dependency on temperature anymore • Temperature dependency with and without feedback circuitry about 1nT/80°C • Design verification • By SFT before and after each test step • Check of commanding, different modes, noise and scale values • Remaining tasks • Verification of housekeeping values • Spot shielding / removing of test points