A statistic approach to electronic moulding versus traditional plaster moulding

A statistic approach to electronic moulding versus traditional plaster moulding. Jean Claude de Mauroy , Frédéric Barral , Cyril Lecante. Background. Electronic moulding tends to gradually replace the plaster moulding. Is it as effective ?

wilmer
Download Presentation

A statistic approach to electronic moulding versus traditional plaster moulding

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. A statistic approach to electronic moulding versus traditional plaster moulding Jean Claude de Mauroy, FrédéricBarral, Cyril Lecante

  2. Background • Electronic moulding tends to gradually replace the plaster moulding. Is it as effective? • There is no study on the results obtained by this process for the scoliosis. • In the department of orthopaedic medicine of the Clinique du Parc - Lyon, the electronic moulding replaced the traditional plastered moulding in May, 2008. • Are the radiological angular corrections better in brace?

  3. Adjusting position

  4. Comparing virtual vertebral reconstruction and real X ray

  5. Full 3D Electronic Moulding

  6. automatic modification of the shape

  7. Methodology • The angular correction of 166 plaster moulding scoliosis was compared with the correction of 117 electronic moulding scoliosis (random). • Both mouldings were made by the same physician. • The electronic moulding has been produced using the full 3D system ORTEN. • The curves have been grouped into thoracic (n=127), thoracolumbar(n=65) and lumbar(n=206).

  8. Lumbar brace Plaster cast Init

  9. Results • The average initial angle is: 28,19 (+-9,21) for thoracic, 28,11 (+-9,34) for thoraco-lumbar and 25,86 (+-7,04) for lumbar curves. • The angular reducibility is 54 % in braces for the thoracic curvatures, 69 % for the thoraco-lumbar curvatures and of 73 % for the lumbar curvatures, which corresponds to the usual results of the Lyon management.

  10. Anatomo-radiological classification

  11. In Brace angular reducibility 73 % 69 % 54 %

  12. Comparative results

  13. Electronic versus Plaster Moulding But nonsignificant 5.44° 4.75° 3.89° 3.63° 3.02° 2.14°

  14. Conclusion • Such results are in favour of the electronic moulding which remains however delicate and require: • a precise position of the patient during the surface topography and • well trained orthotic technicians.