Fast BLM Functional Specs. 2 different inputs. Func Spec ( EDMS 1233009-Injectors )… Table of Content. 1) Introduction 4 2) Machine Parameters 4 3) Derived Parameter: Integrated Loss over the cycle 5 4) Derived Parameter: Integrated Loss during beam presence 6

yoshe
Download Presentation

Fast BLM Functional Specs

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Fast BLMFunctional Specs 2 different inputs

  2. Func Spec (EDMS 1233009-Injectors)… Table of Content 1) Introduction 4 2) Machine Parameters 4 3) Derived Parameter: Integrated Loss over the cycle 5 4) Derived Parameter: Integrated Loss during beam presence 6 5) Derived Parameter: Ambient Radiation 7 6) Derived Parameter: Maximum Loss Integral Values 8 7) Derived Parameter: Loss Evolution Buffer 9 8) Derived Parameter: Capture Buffer 11 9) Derived Parameter: Integrated Losses in the transfer-lines 12 10) Derived Parameter: Loss evolution in the transfer-lines (spill) 12 11) Derived Parameter: Post Interlock Buffer 13 12) Resolution of Measurements 14 13) Interlocking Functionality 15 14) Fast (<2 µs) Loss Observation 16 15) Related Instrument and Monitors 16

  3. … Fast (<2 µs) Loss Observation Expected Results In dedicated locations, e.g. the PSB injection dump and PS and SPS injection and extraction regions, BLM detectors able to measure losses with a time-resolution better than 2 µs should be installed for observation purposes. Such measurements are expected to be extremely beneficial during commissioning, operational optimisations and other situations where more detailed structure of the beam loss evolution is required. Remarks • Measurements from these detectors will be acquired via OASIS-type systems • No beam-loss interlocking should be made for these measurements on the front-ends

  4. … Monitors in Func. Spec

  5. Bernd TB’s Technical Specification of Fast DAQ… Fast BLM acquisition system, B.Dehning

  6. Bernd’s Specification • Sampling frequency, bandwidth • 500 MHz analogue bandwidth relates to a rise time of 0.7 ns (10 – 90%) • Rise time observation 2.3 ns, limited by 1GHz sampling frequency and cable, rise time contribution from analogue part of acquisition almost negligible Fast BLM acquisition system, B.Dehning

  7. ??? Objectives/Priorities ??? • I> Beam Instrumentation • Things we must do and are expected to • Request from OP • We should not miss that • Our Specs up to 500 kHz (2 us) • II> Beam Observation • Request from ABP • We should be proactive there and helpful • Fast BLMs up to 500 MHz via (or ‘a la’) OASIS. • III> Instrumentation R&D • This is for us (to be ready for Beam Observation) • One channel somewhere with more for us to test