MR Angiography. Hattie Dong MRSRL Study Group Friday, July 18, 2008. What and Why. MR angio: study of vessels and blood flow by generating contrast b/w blood and background Applications: Detect aneurysms, atherosclerosis, and other vessel diseases.
E N D
MR Angiography Hattie Dong MRSRL Study Group Friday, July 18, 2008
What and Why • MR angio: study of vessels and blood flow by generating contrast b/w blood and background • Applications: • Detect aneurysms, atherosclerosis, and other vessel diseases. • Visualize blood flow (e.g. during kidney transplant). • Guide surgeons when performing implantation or using a stent.
Overview • Black blood: dark blood signal • RF spin echo • Inversion recovery • Bright blood: bright blood signal • Time-of-flight (TOF) • Projection angiography • Contrast-enhanced MRA (CEMRA)
Spatial Saturation • Removes signals from a specific location • Can be easily added to angio sequences
RF Spin Echo • Complete outflow from imaging slice z when through-plane velocity exceeds: • Single echo: • Double echo: • Adv: obtained for “free” • Drawback: slow/recirculating flow not suppressed
Inversion Recovery • T1blood 1200ms • Choose TI such that blood signal recovers to approximately zero. • Single inversion drawback: static blood and tissues with T1 similar to blood also suppressed
Double IR • Non-selective 180 slice-selective 180 TI imaging sequence • Complete suppression condition: • Common to add DIR module to FSE
Multislice DIR-FSE • Yarnykh, JMRI 2003: Invert a slab and readout at different slice locations.
Triple Inversion Recovery • Simonetti, Radiology 1996: Add one more selective 180 inversion to suppress fat (T1=300ms)
Time-of-Flight • Imaging volume excited with short TR, reducing steady state Mz of static tissue • Inflowing blood experiences fewer excitations.
Flow-Related Enhancement • Incomplete T1 relaxation in between RF excitations, unless T1<<TR. • GRE steady state FID:
FRE • If spins experience j excitations: • j FRE j Ernst (Ernst is angle that maximizes FRE)
2D & 3D TOF • 2D: Fresh spins enter slice after each excitation • Spatial saturation to suppress venous flow • 3D: blood traverses entire slab, Mz lowers over time • Higher FRE at entrance than exit • MIP to recon
MOTSA • Tradeoff: image volume vs. blood “freshness” MOTSA • Parker, MRM 1991: Overlapping short 3D slabs acquired • Slice in overlap region: acquire as entry slice from one slab and exit slice from another, then take maximum
Projection Angiography • TOF recon uses MIP: contrast reduced at high resolution • If background signal sufficiently suppressed, can directly perform 2D projection of the 3D volume
Multiple Inversion Recovery Mani, MRM 1997
sm m-1 m Mz(m-1) Mz(m) MIR • Propagate through n tip pulses and image at a time Q after spatial saturation, =, Mz(0)=0:
MIR • 4 inversions, 3 species to null nonlinear underdetermined system • Quasi-Newton optimization (BFGS), least norm of final Mz’s.
MIR Implementation • Algorithm sensitive to B0 & B1 inhomogeneity • Saturation: CHESS (Moonen, MRM 1990) • Inversion: Adiabatic sech pulses
Contrast-Enhanced MRA • Contrast agent gadolinium chelate (molecular cage) • T1 reduction of blood: • Causes skin thickening in kidney failure patients
Discussions • Black blood methods: • RF spin echo and IR-related • Intravoxel dephasing due to complex flow further reduces blood signal • Pulsatile flow does not produce ghosting as blood nulled • Bright blood methods: • 2D & 3D TOF, MIR PA, CEMRA • Clear depiction of vessels without static signal distraction