Evaluation of the New Thermo- Fisher QMS Everolimus Assay

Evaluation of the New Thermo- Fisher QMS Everolimus Assay Julie Tran , J. Ray, K. Gao , P. Williams, M.James. Everolimus is a mammalian target of rapamycin inhibitor ( mTOR ) and it is used in the prevention of transplant rejections by preventing T-cell activation.

zia
Download Presentation

Evaluation of the New Thermo- Fisher QMS Everolimus Assay

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Evaluation of the New Thermo- Fisher QMS Everolimus Assay Julie Tran, J. Ray, K. Gao, P. Williams, M.James

  2. Everolimusis a mammalian target of rapamycin inhibitor (mTOR) and it is used in the prevention of transplant rejections by preventing T-cell activation. • Performed on the Thermo Fisher CDx90 analyser using QMS ( Quantitative Microparticle System) Everolimus Reagents • Turbidimetric immunoassay • Assay time: 10 mins per sample • Evaluation included: • Reagent stability under various conditions • Comparison with HPLC-MSMS • Sensitivity • Specificity • Precision • Specimen stability • Evaluation was conducted according to the CLSI guidelines (formerly NCCLS)

  3. Comparison with HPLC-MSMS

  4. Reagent Stability • Ran 3 levels of QC, twice a day, on consecutive days ( except wkends) for 2 weeks or until end of reagents. • 3 tests were set up: • Reagents to be left uncapped at all times • Reagents to be capped at the end of the day • Reagents to be capped at the end of each run • Each test was initially calibrated and whenever necessary. Ie, when at least 1 QC levels exceeded the 2 SD • Mean, SD and CV were calculated

  5. Cumulative CV

  6. Analytical Sensitivity Limit of Detection was 0.086ng/mL Functional Sensitivity Limit of Quantitation was 0.237ng/mL Intra-assay Precision Level 1 mean : 2.70ng/mL SD: 0.09 CV: 3.30% Level 2 mean: 5.77ng/mL SD: 0.14 CV: 2.43% Level 3 mean: 16.97ng/mL SD: 0.33 CV: 1.96% Inter-assay Precision Level 1 mean : 3.99ng/mL SD: 0.26 CV: 6.42% Level 2 mean: 8.02 ng/mL SD: 0.29 CV: 3.64% Level 3 mean: 16.18 ng/mL SD: 0.48 CV: 2.99%

  7. Specimen Stability

  8. QMS Everolimus on the UKNEQAS External QAP Program Sample 54- 59