1 / 31

EPIC flux comparison from 2XMM sources

EPIC flux comparison from 2XMM sources. S. Mateos, R. Saxton, S. Sembay & A. Read. Sources Used. Point-like sources from 2XMM detected in 2+ cameras > 200 counts in each camera Off-axis angle 0-12 arcmins F 2-10 > 6E-12 have been excluded to avoid pile-up effects.

amalia
Download Presentation

EPIC flux comparison from 2XMM sources

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. EPIC flux comparisonfrom 2XMM sources S. Mateos, R. Saxton, S. Sembay & A. Read

  2. Sources Used • Point-like sources from 2XMM detected in 2+ cameras • > 200 counts in each camera • Off-axis angle 0-12 arcmins • F2-10> 6E-12 have been excluded to avoid pile-up effects

  3. Count rate – flux conversion Count rates converted to fluxes using energy conversion factors (ECF) which are based on a spectral model of an absorbed power-law with NH=3E20, slope=1.7 ECFs calculated using the detector matrices: MOS: On-axis RMF for revolution 375 + on-axis ARF PN: Latest, canned, on-axis, full-frame RMF for single and single+double events + on-axis ARF Count rates found with - MOS: pattern=0-12, PN: 0.2-0.5 keV, pattern=0; 0.5-12 keV, pattern=0-4

  4. PN v MOS-1: Band 3 (1-2 keV)

  5. PN v MOS-1: Band 1 (0.2-0.5)

  6. PN v MOS-1: Band 2 (0.5-1)

  7. PN v MOS-1: Band 4 (2-4.5)

  8. PN v MOS-1: Band 5 (4.5-12)

  9. PN v MOS-1: Flux comparison

  10. PN v MOS-2: Flux comparison

  11. MOS-1 v MOS-2: Flux comparison

  12. Flux Ratios (%) The Kirsch relation: mos = k * pn where k is an energy independent constant, ~1.05 – 1.08 CAL-TN-0052-5 (Stuhlinger et al. 2008)

  13. First Results • MOS cameras agree to better than 4% at all energies. • PN has a ~constant offset from MOS cameras of 7-9% from 0.5-4.5 keV • PN / MOS agreement much better (<3%) in 0.2-0.5 keV band • PN / MOS agreement worse at high energies at least for MOS-1 (12.5%)

  14. Low-Energy difference Why so good ?? Is the use of a single RMF ok ? Reminder: MOS flux conversion uses RMF for on-axis (i.e. on patch) at Rev 0375. PN: Uses on-axis (Y=9) RMF These approximations will mainly effect low energies.

  15. PN v MOS-1: Change with time

  16. PN v MOS-2: Change with time

  17. MOS-1 v MOS-2: Change with time

  18. PN v MOS-1: Off-axis angle Iufh Yth Tj Tyj e

  19. PN v MOS-2: Off-axis angle F F F

  20. Low-Energy summary • Ignoring sources which fall on the MOS patches, i.e. using Θ = 2 – 12 arcmins we get: (m1-pn)/m1 = 10 - 12% (m2-pn)/m2 = 2 - 7% Time variability makes these numbers unreliable but m2/pn looks to be less than ~8%

  21. PN v MOS-1: Flux comparison ?

  22. High-Energy difference (m1-pn) / m1=12.5%Why so high ?? Is the Kirsch relation wrong ??

  23. PN v MOS-1: Off-axis angle

  24. PN v MOS-2: Off-axis angle

  25. MOS-1 v MOS-2: Off-axis angle

  26. What depends on off-axis angle? • Vignetting (all cameras) • RGS obscuration (MOS) • PSF (all cameras) Azimuthal-angle dependent Azimuthal-angle dependent (MOS)

  27. MOS PSF A measure of the XMM PSFs, lighter colour means a sharper PSF.

  28. MOS CCDs 1 4 2 4 1 3 3 2 A measure of the XMM PSFs, lighter colour means a sharper PSF.

  29. PN v MOS-1: Azimuthal angle

  30. PN v MOS-2: Azimuthal angle

  31. Conclusions for MOS / PN • MOS = PN * 1.08 from 0.5 – 4.5 keV • With this analysis we can’t say what the relation is in the 0.2-0.5 keV band. • At high energies there is an extra off-axis angle, azimuthal-angle dependent effect which increases the MOS excess. This aligns with the RGS dispersion direction and probably means that the RGS absorption needs recalibrating.

More Related