1 / 40

Sudden jumps in the absolute energy scale and their origin

Sudden jumps in the absolute energy scale and their origin. FF Mn-K Pos Quadrant 0. ± 5 eV. Effect of the particle background. FF Mn-K Pos Quadrant 0. + 41.5 eV. ± 5 eV. 2005 Dec 20 – 2006 Jan 01. FF AL-K Pos Quadrant 0. ± 5 eV. + 10.5 eV.

Download Presentation

Sudden jumps in the absolute energy scale and their origin

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Sudden jumps in the absolute energy scale and their origin EPIC pn: Jumps in the energy scale

  2. FF Mn-K Pos Quadrant 0 ± 5 eV EPIC pn: Jumps in the energy scale

  3. Effect of the particle background EPIC pn: Jumps in the energy scale

  4. FF Mn-K Pos Quadrant 0 + 41.5 eV ± 5 eV 2005 Dec 20 – 2006 Jan 01 EPIC pn: Jumps in the energy scale

  5. FF AL-K Pos Quadrant 0 ± 5 eV + 10.5 eV 2005 Dec 20 – 2006 Jan 01 EPIC pn: Jumps in the energy scale

  6. Al-K: 1.486 keV, +10.5 eV  +0.707% Mn-K: 5.894 keV, +41.5 eV  +0.704%  Change in the gain, not in the offset EPIC pn: Jumps in the energy scale

  7. low background high background Energy resolution somewhat improved during high background • reduced charge loss noise  reduced charge loss • increased trap saturation Energy shift occurs also for first singles • precursors ‘invisible’ • precursors below lower threshold or in preceding frame(s) EPIC pn: Jumps in the energy scale

  8. The CTI during low and high background rev 709, low background Al-K Mn-K  decreased CTI during high background (Mn-K) rev 711, high background Al-K Mn-K EPIC pn: Jumps in the energy scale

  9. low background EPIC pn: Jumps in the energy scale

  10. high background EPIC pn: Jumps in the energy scale

  11. During episodes of high particle background the CTI is reduced. • This is most likely caused by saturation of traps due to precursors. • The charge of the precursors is below the 20 adu threshold and thus not available. • The usual CTI correction cannot be used. • But there may be a correlation with the background flux which can be used for the correction.. EPIC pn: Jumps in the energy scale

  12. Effect of the quadrant box temperature EPIC pn: Jumps in the energy scale

  13. Gain/temperature correlation: earlier investigations EPIC pn: Jumps in the energy scale

  14. Mn-Kα & Al-K position and quadrant box temperature EPIC pn: Jumps in the energy scale

  15. Quadrant box temperatures (yellow circles: no FF mode) EPIC pn: Jumps in the energy scale

  16. Mn-Kα position vs. temperature: CCD 1 & 2 EPIC pn: Jumps in the energy scale

  17. Mn-Kα position vs. temperature: CCD 3 & 4 EPIC pn: Jumps in the energy scale

  18. Mn-Kα position vs. temperature: CCD 5 & 6 EPIC pn: Jumps in the energy scale

  19. Mn-Kα position vs. temperature: CCD 7 & 8 EPIC pn: Jumps in the energy scale

  20. Mn-Kα position vs. temperature: CCD 9 & 10 EPIC pn: Jumps in the energy scale

  21. Mn-Kα position vs. temperature: CCD 11 & 12 EPIC pn: Jumps in the energy scale

  22. Mn-Kα position vs. Al-K position EPIC pn: Jumps in the energy scale

  23. Deviations of the Al-K and Mn-Kα positions EPIC pn: Jumps in the energy scale

  24. Mn-Kα position vs. Al-K position EPIC pn: Jumps in the energy scale

  25. Mn-Kα positions (after gain and CTI correction) EPIC pn: Jumps in the energy scale

  26. Mn-Kα positions after temperature (and dCTI/dt) correction EPIC pn: Jumps in the energy scale

  27. “CCD offsets” EPIC pn: Jumps in the energy scale

  28. Quadrant box temperatures Doy 2000 EPIC pn: Jumps in the energy scale

  29. EPIC pn: Jumps in the energy scale

  30. EPIC pn: Jumps in the energy scale

  31. Standard correction EPIC pn: Jumps in the energy scale

  32. with 1 adu / 2000 d drop EPIC pn: Jumps in the energy scale

  33. with 1 adu / 2000 d drop EPIC pn: Jumps in the energy scale

  34. Mn-Kα position [adu] (in quadrant 0, after 1 adu / 2000 d drop) slope: +0.43 adu / C Mean quadrant box temperature [C] (F1576..F1876) EPIC pn: Jumps in the energy scale

  35. standard correction EPIC pn: Jumps in the energy scale

  36. after temperature.. correction EPIC pn: Jumps in the energy scale

  37. standard correction EPIC pn: Jumps in the energy scale

  38. after temperature.. correction EPIC pn: Jumps in the energy scale

  39. Multiply all energies with a factor F: F = ( 1 - A(t) - B(Tq) ) C A = 4.24 *10-7 (t-t0)/[d] B = 3.65 * 10-4 Tq/[oC] C = 1.0057 t0 : 2000-Jan-01 Tq: mean of T(F1576)..T(F1876) EPIC pn: Jumps in the energy scale

  40. EPIC pn: Jumps in the energy scale

More Related