An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. SWEA MAVEN Calibration Plan A. Fedorov

  2. Outline: • New electron gun presentation. • Electron beam properties • Another calibration facilities and control system • What and how to measure

  3. Overview of the new e- gun The goal: To create a wide uniform e- beam as close to the sensor as possible. The total current of the beam and its cross-section distribution should be measured.

  4. Overview of the new e- gun: external parts Hamamatsu Xenon lamp on the movable support Fused silica window

  5. Overview of the new e- gun: internal parts Palladium photocathode and an acceleration system: 1. Energy up to 10keV 2. Strictly parallel e- beam 3. No disturbed e- trajectories at the beam perifery

  6. Overview of the new e- gun: internal parts The photocathode at the end of the light tube. The beam current monitor. The beam monitor on the Y-Z axis stage. 1. Beam cross-section 2. Beam parallelism 3. Beam energy distribution

  7. Overview of the new e- gun: beam cross-section Flux distribution for close lamp position. 20% of variation within 40mm circle. We know the absolute beam for each sensor position

  8. Sensor positioning Y: 0.1mm Elevation (Thet) : +/- 90 deg, 0.1deg Azimuth (Phi) : +/- 170 deg, 0.1deg

  9. Calibration control

  10. Calibration plan: Main Variables Simulated (left) and measured (right) Elevation-K response For D = 0 Simulated (left) and measured (right) Elevation-K response For D = -0.59 K = E/Uan D = Udef/E Elevation, deg (mechanical) Virtual Elevation Azimuth, deg

  11. Calibration plan: Elevation-D profile GF determination 1. Define D – Elevation profile 2. For Az = const, for Elevations (step = 3 deg) : K-D response 3. Calculate dGF/dAz for this point (each anode)

  12. Calibration plan: GF azimuth profile Repeat the same for all Az (step = 2 deg) and all anodes