First Preconditioning Test of New Batch of 33mm 20µm Pore MCPs

First Preconditioning Test of New Batch of 33mm 20µm Pore MCPs. ANL MCPs 164 and 163, 20µm, 60:1, 8° bias. INCOM serial # s , 11863-99, 11863-95 Resistances 159MΩ, 138MΩ ( NiCr over) XDL detector with MCPs gapped by 0.001” spacer. Completed 350°C bake, with RGA scans.

Download Presentation

First Preconditioning Test of New Batch of 33mm 20µm Pore MCPs

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. First Preconditioning Test of New Batch of 33mm 20µm Pore MCPs • ANL MCPs 164 and 163, 20µm, 60:1, 8° bias. • INCOM serial #s, 11863-99, 11863-95 • Resistances 159MΩ, 138MΩ (NiCr over) • XDL detector with MCPs gapped by 0.001” spacer. • Completed 350°C bake, with RGA scans. • Scrub in progress, with RGA scans. • Have gone from 1µA to 2µA, now 3µA >0.65 C cm-2 extracted.

  2. MCP 164-163 Scrub Record Up to 3µA Up to 2µA 1µA Minimal gain drop for 1µA to 0.06 C cm-2, some small H2 outgas (10-10Torr). Full functional test showed no substantive performance changes except for slightly higher background. Faster drop @ 2µA but stablized quickly with some H2 outgas (4 x 10-10Torr). At 3µA since then, with small drop then recovery to >100% gain – leveling off now @0.65 C Cm-2. 164-163 Scrub @ 3 x 105 gain, 700v per MCP

More Related