Susceptibility Weighted Imaging at 7T

Susceptibility Weighted Imaging at 7T. Brian Welch. MR Clinical Science. IACSM 2007. April 24, 2007. 7T human midbrain. T1W TFE TR/TE=19/9.6. TSE_IR TR/TI/TE=4000/60/10. 7T human brain – high resolution sagittal. High Res. Sagittal Protocol 3D T1W TFE inversion delay = 1816 msec

doris
Download Presentation

Susceptibility Weighted Imaging at 7T

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Susceptibility Weighted Imaging at 7T Brian Welch MR Clinical Science IACSM 2007 April 24, 2007

  2. 7T human midbrain T1W TFE TR/TE=19/9.6 TSE_IR TR/TI/TE=4000/60/10

  3. 7T human brain – high resolution sagittal High Res. Sagittal Protocol 3D T1W TFE inversion delay = 1816 msec TFE factor = 180 0.33 x 0.33 x 4.0 mm3 24 slices TR/TE=19/9.6 Scan Time = 7:23

  4. SWI Background and References Primarily associated with E. Mark Haacke, Ph.D. • Professor of Radiology, Wayne State University, Detroit, MI • Director, Magnetic Resonance Imaging Institute for Biomedical Research (http://www.mrimaging.com) U.S. Patents (http://www.uspto.gov) • 6,501,272 - MR Methods, some post-processing • 6,658,280 - post-processing, CNR optimization Selected SWI Publications • Small Vessels in the Human Brain: MR Venography with Deoxyhemoglobin as an Intrinsic Contrast Agent, JR Reichenbach, EM Haacke, et al. Radiology 204:272-277(1997) • High-resolution BOLD venographic imaging: a window into brain function, JR Reichenbach, EM Haacke. NMR Biomed 14:453-467(2001) • Susceptibility Weighted Imaging (SWI), EM Haacke, JR Reichenbach, et al. MRM 52(3):612-618(2004) • Magnetic Susceptibility-Weighted MR Phase Imaging of the Human Brain, A Rauscher, JR Reichenbach, et al. AJNR 26:736-743(2005)

  5. Mark Haacke’s website – www.mrimaging.com

  6. Why the interest in SWI at 7T? • Phase accrual per unit time is a function of static field strength • Shorter echo times mean shorter scan times or better coverage • 40-50 msec at 1.5T • 20-25 msec at 3.0T • 9-11 msec at 7.0T • Increased SNR • Popular topic at 2007 ISMRM Workshop on Advances in High Field MR • ~6 invited presentations • many posters

  7. Possible Clinical Applications for SWI • Small vessel (vein) imaging • Microhemorrhage • Mineralization, e.g. iron • Occult vascular disease • Stroke • Traumatic brain injury • Tumor vascularization • Multiple sclerosis • Anatomical imaging of the midbrain

  8. Typical SWI Protocol • 3D gradient echo • High resolution (0.5 mm – 1.0mm in-plane, 1.0-2.0 mm slice thickness) • RF spoiled (“T1 Enhancement”) • Full flow compensation • TE – long enough for susceptibility-induced effects to evolve • TR – shortest • Save magnitude, real and imaginary parts for post-processing

  9. SWI Post-Processing Steps • Low-pass filter original image • Produce high-pass image from complex division of original image by low-pass filtered image • Create phase mask from high-pass phase image with negative or positive weighting • Operate on original magnitude images with phase mask images • Perform minimum intensity projections (mIP’s) across neighboring slices

  10. Applying a phase mask

  11. Low pass filtering using pyramidal k-space window • Most SWI publications refer to convolution with a Hanning window - requires padding to avoid aliasing across image volume boundaries • Simplier approach is to apply a triangular/pyramidal window in k-space (borrowed from Jim Pipe’s PROPELLER processing) JG PIPE MRM 2004

  12. 7T 2D FFE high-pass phase image 2D FFE Protocol RF-spoiled, flow compensated SENSE factor 2 0.5 mm x 0.63 mm in-plane thick/gap = 3 mm/2 mm 20 slices TR/TE/FA/NSA=349/12/20/4 Scan Time = 3:25

  13. Negative phase mask – not so impressive 2mm mIP’s Control Negative Phase Mask

  14. Other phase mask options

  15. 7T human brain – SWI SWI Protocol 3D FFE RF-spoiled, flow compensation 0.5 x 0.5 x 1.0 mm3 30 slices TR/TE/FA=30/14.8/18 Scan Time = 7:18 High-pass filtered phase from original complex-divided by low-pass filtered (pyramidal filter in k-space). Positive phase mask multiplied with magnitude data 4 times.

  16. 7T human brain – SWI SWI Protocol 3D FFE RF-spoiled, flow compensation 1 x 1 x 1 mm3 80 slices TR/TE/FA=20/18/16 Scan Time = 6:07 High-pass filtered phase from original complex-divided by low-pass filtered (pyramidal filter in k-space). Positive phase mask multiplied with magnitude data 4 times. 20 mm mIP’s

  17. 7T SWI (multi-echo) with Buckeye06 3.9 6.9 9.9 12.9 15.9 18.9 21.9 Protocol 3D FFE RF-spoiled, flow compensation 7 echoes SENSE factor 2 1 x 1 x 1 mm3 128 slices TR/TE/ΔTE/FA=24/3.8/3.0/12 Scan Time = 4:44

  18. 7T T2* mapping T2* [msec] 100.0 SWI Protocol 3D FFE RF-spoiled flow compensation 7 echoes SENSE factor 2 1 x 1 x 1 mm3 128 slices TR/TE/FA=24/3.9+n*3.0/12 Scan Time = 4:44 0.0 M0 T2*

  19. 7T SWI (multi-echo) – full brain coverage 7.1 3.1 5.1 9.1 Protocol 3D FFE RF-spoiled, flow compensation 7 echoes SENSE factor 2 1 x 1 x 1 mm3 140 slices TR/TE/ΔTE/FA=20/3.1/2.0/10 Scan Time = 4:15 11.1 13.1 15.1

  20. mIP’s at varying TE TE/ΔTE=3.8/3.0 Control Negative phase mask Positive phase mask

  21. Full brain venograms Protocol 3D FFE RF-spoiled, flow compensation SENSE factor 2 1 x 1 x 1 mm3 140 slices TR/TE/ΔTE/FA=20/3.1/2.0/10 Scan Time = 4:15

  22. T2*W FFE – MultiVein MultiVane CARTESIAN MULTIVANE

  23. Related topic Positive Contrast SWI (PFL Hamburg, Hannes Dahnke)

  24. Conclusions • SWI data collection is straightforward using existing Philips protocol options • Full brain, high-res SWI data can be acquired at 7T in a short time • SWI post-processing consists of simple steps, though best results yielded using a method inconsistent with published approaches • SWI may be a method waiting for an application • Current plans to include SWI as part of an IRB-approved protocol to image pathology (patients recruited from hospital MR center)

  25. 7T “Clinical” ExamCard (WIP) 10 cm = 100 mm slice coverage for axial and sagittal orientations, except isotropic 1mm which covers 172 mm • SCOUT SHC16 00:29 • REF SHC16 4x4x4 01:55 • B1MAP 3DFFE 04:44 2.0 x 2.0 x 2.0 • T1_3D_iso1mm 07:14 1.0 x 1.0 x 1.0 SENSE 2.5 • T1_TRA_2mm 02:52 0.5 x 0.5 x 2.0 SENSE 2.0 • T2s_TRA_2mm 04:10 0.5 x 0.5 x 2.0 SENSE 2.0 • T2s_SAG_4mm 04:18 0.33 x 0.33 x 4.0 SENSE 2.0 • T2s_TRA_4mm 04:17 0.33 x 0.33 x 4.0 SENSE 2.0 • T2_TRA_GRASE 06:23 0.5 x 0.5 x 2.0 SENSE 2.0 • SWI 9echo 03:58 1.0 x 1.0 x 1.0 SENSE 2.0 • MULTI-FLIP 03:30 1.0 x 1.0 x 2.0 SENSE 2.0

  26. 4. T1_3D_iso1mm 07:14

  27. 5. T1_TRA_2mm 02:52

  28. 6. T2s_TRA_2mm 04:10

  29. 7. T2s_SAG_4mm 04:18

  30. 8. T2s_TRA_4mm 04:17

  31. 9. T2_TRA_GRASE 06:23

  32. 10. SWI 9echo 03:58

  33. 11. MULTI-FLIP 03:30

  34. 7T human brain – T1 mapping T1 [sec] 0.0 Multi-Flip Protocol 3D RF-spoiled FFE 5 angles {20,16,12,8,4} 1.5 x 1.5 x 2.5 mm3 40 slices TR/TE=14/3.9 Scan Time = 6:56 2.5 M0 T1

  35. Other future work…