430 likes | 645 Views
Biophotonics lecture 7. December 2011. Exam date ? Monday, 30 January 2012 or Wednesday, 1 February 2012. Last week: Stimulated emission depletion (STED) microscopy. Today: Imaging deep in tissue: 2-photon microscopy Enlarging the NA: 4Pi microscopy
E N D
Exam date ? Monday, 30 January 2012 or Wednesday, 1 February 2012
Last week: • Stimulated emission depletion (STED) microscopy
Today: • Imaging deep in tissue: 2-photon microscopy • Enlarging the NA: 4Pi microscopy • Super-resolution: Pointillism, STORM, PALM
Imaging deep in tissue:2-photon microscopy Refractive(andscattering)tissue Imaging singlecells Imaging deep in tissue objectivelens not welldefinedfocalspot welldefinedfocalspot Refractive(andscattering)tissue welldefinedfocalspot welldefinedfocalspot
The Problem: Scattering, aberrations, absorption Rayleigh scattering: ~ l-4 Blue: Bad! Red / Infrared: OK!
Imaging deep in tissue:2-photon microscopy Solution: imagingusinglongerwavelength objectivelens Refractive(andscattering )tissue not welldefinedfocalspot welldefined, but LARGER focalspot
x x z z Focal spot, l=500nm Focal spot, l=1000nm ATF ATF OTF OTF
Fluorescence Jablonski diagram Absorption… … and spontaneous emission
Fluorescence Jablonski diagram NO absorption…
2-photonfluorescence Jablonski diagram 2-photon absorption… … and spontaneous emission
2-photonfluorescence • 2-photon absorption requires two photons to be present simultaneously • The probability for this growsquadratically with intensity • It will only occur where the local intensity is high
x x x z z z Focal spot, l=500nm Focal spot, l=1000nm 2-photon, l=1000nm missingconefilledopticalsectioning
2-photonfluorescence Zipfel, Williams, Webb, Nature Biotechnology21, 1369 - 1377 (2003)
DichromaticReflector Wide Area Detector at close destance emissionphotons will still bemultiplyscatteredandcannotbefocussed on a pinhole Non-descanneddetectionneededtomaximizecapturearea
Two Photon Microscopy • Much lessabsorption • Much lessscattering • Feweraberrations • Less out-of-focus bleaching • Inherentopticalsectioning
Sample betweenCoverslips Detector Pinhole FluorescenceIntensity z Dichromatic Beamsplitter HighSidelobes z 2 Photon Effect 4 Pi Microscope (Type C) Stefan W. HellMax Planck Institute of Biophysical Chemistry Göttingen, Germany Aperture increase: Laser Illumination Emission
OTF ATF widefield 4Pi
widefield, l=500nm 4Pi, l=500nm 4Pi PSFs widefield, l=1000nm 2-photon, l=1000nm 4Pi, l=1000nm 4Pi, l=1000nm, 2-photon
Leica 4Pi http://www.leica-microsystems.com
DevidingEscherichia Coli 4Pi images From: Bahlmann, K., S. Jakob, and S. W. Hell (2001). Ultramicr. 87: 155-164.
4Pi images Confocal (2-Photon ) 4Pi (2-Photon) Thanks to: Elisabeth Ehler, Reiner Rygiel, Martin Fiala, Tanjef Szellas
Localization, not resolution If positions are know you can paint a picture! Seurat: Tiger Douthwaite: Lewis Hamilton If particles can be separated, their relative positions can be measured accurately
Localization, not resolution PSF position
Localization, not resolution position ?????
How to separate particles? Spectral precision distance microscopy Problems: Chromatic Aberrations, few dyes Using fluorescence lifetime for separation (FLIM) Problems: Lifetime depends on microenvironment Use the blinking characteristics P. Edelmann, A. Esa, H. Bornfleth, R..Heintzmann, M. Hausmann, and C. Cremer. Proc. of SPIE , 3568:89-95, 1999 M. Heilemann, D.P. Herten, R.Heintzmann, C. Cremer, C. Müller, P. Tinnefeld, K.D. Weston, J. Wolfrum and M. Sauer. Anal. Chem., 74, 3511-3517, 2002. K.A. Lidke, B. Rieger, T.M. Jovin, R. HeintzmannOptics Express13, 7052-7062, 2005.
How to separate particles? Better: Avoid overlap entirely by temporally separating the particles E. Betzig, "Proposed method for molecular optical imaging", Opt. Lett. 20, 237 (1995)
Earth at night Earth
Jena at night Task: Localizationoftheuniversitybuildings How: EachProfessor hasto turn on the light foroneminute Resoution Localizingismuchmoreprecisethanresolution
Separation over time Withoutlabelling:everythingisbright Labellingtheuniversitybuildingswidefield: badresolution Pointillistic: accuratemap
Pointillism, PALM, STORM PALM Photo-activation and localisation microscopy other techniques:STORM, FPALM http://jcs.biologists.org/cgi/reprint/123/3/309.pdf
EM Pointillism, PALM, STORM WF PALM E. Betziget al., Science, DOI: 10.1126/science.1127344, Aug. 2006 MitochondriaCOS-7 Zellen Cryo-Schnitte Cytochrom C Oxidase import Sequenz - dEosFP
Pointillism, PALM, STORM Hochauflösende Struktur der Podosomen (Vinculin) New, sophisticated algorithms, which can handle overlappingfluorophores
Pointillism, PALM, STORM 400nm Podosomenbildung Susan Cox, Edward Rosten, Marie Walde, James Moneypenny, Gareth Jones
Confocal microscopy dSTORM / B3 STED 1 m Comparing some methods Structured illuminationmicroscopy Widefield fluorescence
StochasticOptical Reconstruction Microscopy Science 319, 810 (2008); Bo Huang, et al. Three-Dimensional Super-Resolution Imaging by Stochastic Optical Reconstruction Microscopy Microtubules – (Cy3-Alexa647)
Localisation precision in pointilism:(for Gaussian PSFs) • N photons collected from 1 fluorophore • Positions of these photons are rn=rfluorophore ±with being the standard deviation defined by the PSF • The fluorophore position is determined as the mean of all photon positionsrfluorophore=rn/ N • This mean position has an error ofrfluorophore with rfluorophore = / • With N photons, the localisation precision is better than the resolution Problem: sparseness of labelling