BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0. Composition of Blood. Formed ElementsErythrocytes (RBC)Leukocytes (WBC)NeutrophilsEosinophilsBasophilsLymphocytesMonocytesThrombocytes (Platelets). Plasma92% water7 to 9 % of solutes are proteins55 to 60% Albumin, 15% Globulins, 4% FibrinogenNon-protein nitrogen substance, Enzymes, Antibodies, Electrolytes, etc.Serum: No fibrinogen or Factors II, V and VIII.
E N D
1. Coagulation Testing Diane Jette
BioMedica Diagnostics Inc.
2. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0
3. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0 Hemostasis is the arrest of bleeding from an injured blood vessel Vasoconstriction and compression of injured vessels
Platelets adhere to the site of injury and form a platelet plug
Platelets release factors to augment vasoconstriction and initial vessel wall repair
Platelets provide surface membrane sites and components for the formation of enzyme/cofactor complexes in blood coagulation reactions
4. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0 Coagulation Reactions Lead to the Formation of a Blood Clot Two pathways: Intrinsic and Extrinsic - Coagulation Cascade
Formation of a prothrombin activator - complex of Factor Xa, Factor Va and procoagulant phospholipid on surface of platelets.
Prothrombin activator cleaves prothrombin into two fragments to give Thrombin.
Thrombin cleaves small peptides from fibrinogen to form fibrin monomers that polymerize.
Thrombin activates Factor XIII to cross-link the fibrin to form an insoluble clot.
5. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0 Coagulation Cascade Intrinsic Pathway: (APTT)
Factors VIII, IX, XI, and XII.
Activated on surface of exposed endothelium.
Complexes form on platelet phospholipids.
Extrinsic Pathway: (PT)
Factors IV, V, VII, X
Activated by Tissue phospholipids (Tissue Factor or Tissue thromboplastin) released into blood as a result of tissue damage.
6. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0 Coagulation Cascade Common Pathway (Thrombin Time)
Factors I and II
Leads to the formation of Fibrin Clot
Thrombin time does not measure deficiencies in Intrinsic or Extrinsic pathway
7. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0 The Role of Calcium Ca ions are needed for most of the reactions in the Coagulation Cascade
Ca-chelating agents are used in vitro as anticoagulants (Citrate, EDTA, Oxalate)
When Coagulation Factors are synthesized without Vitamin K they cannot bind Ca and lose enzymatic function
8. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0 Regulatory Mechanisms Inhibition of Factor Activity
Plasma protease inhibitors: anti-thrombin III (ATIII), ?2-macroglobulin, ?1 - antiprotease
Heparin converts ATIII from a slow acting inhibitor to an instantaneous inhibitor of Thrombin, Factor Xa and Factor IXa
Protein C and Protein S are serine proteases that cleave Factors VIII and Factor Va rendering them inactive
9. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0 Fibrolysis Fibrin clot is degraded by protolytic enzymes and fragments dissolved in blood
Process is catalyzed by Plasmin
Plasminogen is converted to Plasmin
Activation by tissue plasminogen activator (tPA) and urokinase
Fibron degrades into large fragments X and Y then smaller fragments D and E
10. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0 Regulation of Fibrolysis Plasminogen activator inhibitors (PAIs) and plasmin inhibitors slow the fibrolysis process
tPA and urokinase have short half-lives and are rapidly cleared through the liver
Unbound plasmin is instantaneously neutralized by ?2-antiplasmin
11. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0 Hereditary Coagulation Disorders Hemophilia A
Factor VIII deficiency
80% of all Hemophilia cases
Hemophilia B
Factor IX deficiency
Prolonged ATPP
Recovered by dilution 1:1 with normal plasma
Normal PT and Normal Bleeding Time
Factor XI Deficiency
5 to 9% of European Jews
?2-antiplasmin Deficiency
12. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0 Acquired Coagulation Disorders Liver Disease
Impaired clotting Factor synthesis
Increased fibronolysis
Thrombocytopenia
Desseminated Intravascular Coagulation (DIC)
Something enters the blood that activates factors
Complication of obstetrics, infection, malignancy, shock, severe brain trauma
Elevated PT, APTT, D-Dimer and other fibron degradation products
13. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0 Circulating Anticoagulants Antibodies that neutralize clotting factor activity
Factor VIII Anticoagulants
Antibody
Same profile as Hemophilia A
Clotting time not restored by mixing with normal plasma
Life-threatening condition
14. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0 Lupus Anticoagulants Antibodies to phospholipid binding sites on clotting factors
Prevent factors from accumulating on phospholipid surfaces
Elevated APTT clotting times not corrected with mixing with normal plasma
PT normal or slightly elevated.
Non-specific depression of clotting factor activities (Factors VIII, IX, XI, XII)
Test sensitivity increased by using diluted reagent
Dilute ATPP reagent, Russell’s viper venom time, Kaolin time
Clotting times corrected with the addition of phospholipids
15. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0 Oral Anticoagulant Therapy Coumadin or Warfarin
Inhibitor of Vitamin K dependant Factor synthesis
Oral anticoagulant
Dose regulated by therapeutic effect
PT assay to measure INR
INR range established for optimum therapeutic effect (typically 2.0 to 3.0)
16. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0 Prothrombin Time: PT PT reagent contains Calcium ions and Thromboplastin from brain tissue (Rabbit).
Thromboplastin (Tissue Factor) protein-lipid complex found in tissues outside blood vessels.
Measures the function of the Extrinsic Pathway.
Sensitive to Factors IV, V, VII, X.
Provided as a lyophilized reagent.
Used to monitor oral anticoagulant therapy (Warfarin / Coumadin).
17. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0 PT Reagent Calibration Reagents are calibrated against standard PT reagent established by the WHO.
ISI = International Sensitivity Index.
ISI is assigned by the manufacturer for each lot of reagent using reference material traceable to WHO.
The lower the ISI the more sensitive the Reagent
ISI of 1.8 to 2.4 = Low sensitivity (North American Standard PT)
ISI of 1.4 to 1.8 = Average sensitivity
ISI 1.0 to 1.4 = High Sensitivity
18. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0 PT: INR Values INR = International Normalised Ratio.
MNP = Mean Normal Plasma.
INR = (PT / MNP)ISI
An INR of 1.0 means that the patient PT is normal.
An INR greater then 1.0 means the clotting time is elevated.
19. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0 INR Calculation Example 1
MNP = 12.0 s
ISI = 1.25
Patient Plasma = 20 s
INR = (20.0 / 12.0)1.25 = 1.9
Example 2
MNP = 12.0 s
ISI = 1.85
Patient Plasma = 17 s
INR = (17.0 / 12.0)1.85 = 1.9 Example 3
MNP = 12.0 s
ISI = 1.4
Patient Plasma = 20 s
INR = (20.0 / 12.0)1.4 = 2.0
Example 4
MNP = 12.0 s
ISI = 2.0
Patient Plasma = 20 s
INR = (20.0 / 12.0)2.0 = 2.8
20. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0 Expected PT Values Mean Normal Plasma = 10 to 14 seconds.
Mean Normal Plasma value varies with PT sensitivity. A high sensitivity (Low ISI) PT will give a high normal PT value (13 to 15 seconds).
Oral anticoagulant monitoring = Target INR of 2.0 to 3.0.
INR of greater than 5 or 5.5 = unacceptable high risk of bleeding.
%CV between duplicates less than 5%
21. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0 Performing a PT test Pre-warm PT reagent and sample to 37 oC
Add 100 ?L sample to cuvette
Add 200 ?L of PT reagent to cuvette
Start timer
Record time to clot in seconds
Calculate INR
see product insert for PT
22. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0 Activated Partial Thromboplastin Time APTT or PTT
Reagent contains phospholipids and a ‘surface activator’; (Ellagic Acid, Micronized Silica)
Calcium Chloride reagent added to start the reaction.
APTT reagent mimics the surface of a platelet.
Measures activity of clotting factors in the Intrinsic Pathway, factors VIII, IX, XI and XII
No WHO calibration standard
23. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0 Uses of APTT Sensitive to 30 to 40% deficiencies of all factors except VII and XIII.
Heparin inhibits the activity of clotting factors in the Intrinsic Pathway.
A standard curve (APTT time vs Heparin concentration) is prepared using a heparin standard.
APTT is also sensitive to other non-specific Factor inhibitors such as Lupis Anticoagulant.
Can be influenced by Vitamin K deficiency and Coumadin therapy.
Negative APTT result usually rules out Hemophilia.
24. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0 Expected APTT Values Normal Range: 26 to 40 seconds
Slightly Elevated: 45 to 65 seconds
Extremely Elevated = > 70 seconds
%CV less than 7%
25. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0 Performing an APTT Test Pre-warm Calcium Chloride reagent to 37 oC.
Add 100 ?L of sample to cuvette.
Add 100 ?L of APTT to cuvette and incubate for 3 minutes.
Add 100 ?L of Calcium Chloride reagent and start timer.
Record the time to clot in seconds.
See APTT product Insert
26. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0 Heparin Monitoring Prepare Heparin Standards
Prepare stock heparin (10 USP units/mL)
Prepare working heparin dilutions (0.1 to 0.8 U/mL) by diluting stock in normal plasma
Run heparin dilutions as samples in APTT assay
Plot the results as U/mL vs Log Clotting times
Run patient samples in APTT assay and determine heparin concentration from the plot.
27. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0 Heparin Calibration Curve
28. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0 Factor Substitution Tests: PT and APTT Dilute patient sample one to one with adsorbed plasma and serum to determine if normal clotting time is restored.
Serum: Source of Factors IX, X, XI & XII
Adsorbed Plasma: Source of Factors VIII, V, XI, & XII
29. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0 Factor Plots: PT and APTT Dilute a normal plasma sample or control 1 to 10 in saline
Dilute in saline: 100%, 50%, 25%, 12.5%. 6.25%, 3.12%,
Mix one to one with a factor deficient plasma.
Measure PT or APTT and plot Log % Factor vs Log clotting time.
Dilute patient plasma 1 to 10 in saline.
Mix one to one with a factor deficient plasma.
All factors will be restored except the deficient factor. If the factor is present in test sample no reduction in clotting will be seen. (Same clotting time as the 100% standard)
Determine percent factor from standard curve.
30. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0 Factor Calibration Curve
31. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0 Plasma Controls Used to monitor assay performance (QC)
Made from pooled normal human plasma with some factors selectively adsorbed.
Three levels available:
Level 1 control represents a normal plasma
Level 2 represents a slightly elevated plasma (INR ~ 1.5)
Level 3 represents a severely elevated plasma (INR ~ 2.5)
Each laboratory should establish expected ranges for PT and APTT.
32. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0 Thrombin Time (TT or TAT) Measures common pathway.
Fibrinogen --> Fibron Clot
Not sensitive to deficiencies in Intrinsic or Extrinsic pathways.
Reagent consists of animal thrombin.
Normal Clotting time is 15 seconds.
If elevated sample is mixed one to one with normal and re-tested. If normal clotting is not restored then an anti-coagulant is present.
33. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0 Bleed Time Test An incision is made
Time to stop bleeding is measured
Normal clotting time is 7.5 minutes
34. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0 D-Dimer Test D-Dimer is a fibrin monomer
Product of fibrolysis
Latex agglutination assay is used
Anti-D-Dimer antibody coated on micro-latex beads
Cardiac Infarction Marker
35. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0 Activated Clotting Time (ACT) Clotting time of whole blood in the presence of silica based activator.
Normal clotting times = 90 to 170 sec.
Used to monitor heparin doses from 1 to 10 U/mL (APTT is sensitive to heparin at 0.2 to 1 U/mL).
Used with invasive procedures that require on-site adjustment of heparin and protamine dosage. (ex. Cardiopulmonary bypass surgery).
Not amenable for use with an optical instrument, too cloudy.
Also called HMT, Heparin Management Test
36. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0 Fibrin-1 Clot Detection
37. BioMedica Diagnostics Inc. / March 2002 / Rev. 1.0 Clot Detection