521 likes | 1.11k Views
Small Fragment LCP Set Implants & Instruments. Objectives. Describe the implants found in the Small Fragment Locking Compression Plate (LCP) Set Identify the standard and new instruments that are used with locking screws & plates. Small Fragment LCP Set.
E N D
Small Fragment LCP Set Implants & Instruments
Objectives • Describe the implants found in the Small Fragment Locking Compression Plate (LCP) Set • Identify the standard and new instruments that are used with locking screws & plates
Small Fragment LCP Set • Used on small bone fractures -- ankle, forearm, humerus, pelvis • Instruments same as standard Small Fragment plus more • Available in graphic cases with stainless steel or titanium implants
ImplantsStandard Screws • 3.5mm Cortex • 3.5mm Shaft (Cortex) • 4.0mm Cancellous • Fully threaded • Partially threaded • All with standard small fragment hex head • 2.7mm Cortex Screw also in set (not shown)
Screw Characteristics Head Shaft Thread Core • Dimensions determine drill bit sizes • Core of all screws = 2.5mm drill bit • Thread diameter of cortex screws = 3.5mm drill bit
Implants3.5mm Locking Screw • StarDrive™ recess • Threaded underside of head • Locks screw to locking plate • Small thread pitch • Large 2.8mm core diameter
ImplantsStraight Plates 3.5mm Locking Compression • One Third Tubular 3.5mm Straight Reconstruction 3.5mm Curved Reconstruction
ImplantsSpecialty Plates 3.5mm Oblique T-Plate, L & R • 3.5mm T-Plate 3.5mm Locking Proximal Humeral Plate
Locking Plate FeaturesCombination “Combi” Hole • Smooth side accepts cortex, shaft & cancellous screws • Threaded side accepts locking screws • Plate is contourable
Standard Instruments Same as other Small Fragment Sets: • Exposure • Reduction • Screw insertion • Plate bending • Plate application
Exposure Instruments • Periosteal Elevator • Hohmann Retractors
Reduction Instruments • Forceps • With points -- narrow & broad • Serrated forceps
Screw Insertion Instruments • Drill Bits • 3.5mm • 2.5mm • 2.0mm (not shown)
Screw Insertion Instruments • Universal Drill Guide • Spring loaded side accepts 2.5mm drill bit • Opposite side accepts 3.5mm drill bit & taps • Must be used with locking plates 2.5mm 3.5mm
Screw Insertion Instruments • 3.5mm Universal Drill Guide • Spring loaded end allows precise placement of screw in non-threaded end of plate hole • Spring activated = “neutral” position • Spring not activated = “load” or “compression” position (away from fracture line)
Screw Insertion Instruments • Countersink • Use with quick connect Tap Handle
Screw Insertion Instruments • Depth Gauge • Calibrations in 2mm increments • Measures to 50mm length screws • Hook on same side as calibrations
Screw Insertion Instruments • Taps • 3.5mm cortex • 4.0mm cancellous • Used w/QC Tap Handle • Always pass with Tap Sleeve
Screw Insertion Instruments • Small Fragment Hex Screwdriver • Hints: • Screw driver shaft + tap handle = back-up screwdriver • Use holding sleeve to remove screws from screw rack
3.5mm Cortex Lag TechniqueSteps • Drill 3.5mm • Drill 2.5mm • Countersink • Measure • Tap sometimes – 3.5mm • Insert 3.5mm Cortex Screw
4.0mm Cancellous Lag TechniqueSteps • Drill 2.5mm • Measure • Tap – rarely • Insert 4.0mm Cancellous Screw
Plate Bending Instruments • Bending Pliers • Bending Irons • Bending Templates
Instruments for Locking Plates 2.8mm Drill Bit • Used to drill for 3.5mm Locking Screws
Correct Incorrect LCP Instruments 2.8mm Threaded Drill Guide • Ensures proper drilling angle • Not for use as a plate bender
LCP Instruments 1.6mm Wire Sleeve • Inserts into Threaded Drill Guide to accept 1.6mm Guide Wire • Screw placement can be checked under image
LCP Instruments Direct Measuring Device • Used over 1.6mm K-wires to measure for locking screw length
LCP Instruments StarDrive Screwdrivers • Self-retaining • StarDrive Shaft used with Torque Limiting Attachment
LCP Instruments Torque Limiting Attachment • Allows power insertion of locking screws • No further tightening of screws necessary
LCP Instruments Threaded Plate Holder • Non-cannulated • Used as handle to position plate and insert plate in percutaneous technique • Can be used to bend T-plates and One-third Tubular Plates when used in pairs
LCP Instruments Push-Pull Reduction Device (Whirly-Bird) • Secures plate to bone • Guards against “helicopter effect” of the plate • Hole can be used for cortex screw insertion (same 2.5mm diameter as drill bit)
Care & Maintenance • Keep drill bits & taps sharp • Clean cannulations • Disassemble instruments • Lubricate
Care & Maintenance • Tighten screwdriver holding sleeve • Clean graphic cases
Care & Maintenance • Inspect threads on Threaded Drill Guides
Quiz Question • True or false? You must remember to pull the Small Fragment locking screw rack in addition to the instrument and plate set. • False. Small fragment instruments, plates & screws are in the same graphic case.
What is the drill bit size used to drill for the core of the 3.5mm cortex screw? 2.5mm • What is the drill bit size used to drill for the core of the 3.5mm locking screw? 2.8mm
For what types of procedures might you have the LCP Small Fragment Set in the room? • Older patients (osteoporosis) • Re-operation on a non-union • Replacing broken hardware • Other ?
Now you should be able to: • Describe the implants (screws & plates) in the LCP Small Fragment Set • Identify the instruments that are used with standard and locking screws & plates