1 / 6

FilterWire-Ex in Primary PCI

Mechanical Prevention of Distal Embolization During Primary Angioplasty: Safety, Feasibility, and Impact on Myocardial Reperfusion.

lee-quinn
Download Presentation

FilterWire-Ex in Primary PCI

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Mechanical Prevention of Distal Embolization During Primary Angioplasty: Safety, Feasibility, and Impact on Myocardial Reperfusion Ugo Limbruno, MD, PhD; Andrea Micheli, MD; Marco De Carlo, MD; Giovanni Amoroso, MD, PhD; Roberta Rossini, MD; Caterina Palagi, MD, PhD; Vitantonio Di Bello, MD; Anna Sonia Petronio, MD; Gabriella Fontanini, MD; Mario Mariani, MD Circulation. 2003;108:171-176

  2. FilterWire-Ex in Primary PCI 106 patients with acute MI undergoing primary PCI Matched Case-Control Study • FilterWire-Ex (FW) • A distal embolic protection device • Successfully positioned in 89% of patients without complication • (n = 53) • Matched Control • Matched for infarct-related artery, TIMI flow grade, gender and age • (n = 53) • Endpoints: • Primary – Feasibility and safety of adjunctive FW use during primary PCI • Secondary – Markers of effective reperfusion (ST resolution, myocardial blush grade, and cTFC); peak CK and CK-MB release, change in LVEF and WMSI at 30 days compared with admission; MACE (death, reinfarction, and need for target vessel revascularization) at 30 days Circulation 2003; 108:171-176

  3. FilterWire-Ex in Primary PCI Corrected TIMI Frame Count p=0.005 Frames Circulation 2003; 108:171-176

  4. FilterWire-Ex in Primary PCI Myocardial Blush Grade 3 p=0.006 Circulation 2003; 108:171-176

  5. FilterWire-Ex in Primary PCI ST-segment Resolution p=0.006 Circulation 2003; 108:171-176

  6. FilterWire-Ex in Primary PCI MACE at 30 Days p=0.488 Circulation 2003; 108:171-176

More Related