An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. New Technology for Sulfur Detection Otto Prohaska Science and Technology Group Otto.Prohaska@PerkinElmer.com

  2. Analytical Instrument Manufacturer GC, GC/MS, HPLC ICP-OES, ICP-MS, AA UV/Vis, FTIR, FT Imaging Research new technologies New instrument development for total and speciated sulfur monitoring New sulfur detection technology – patents filed PerkinElmer Instruments

  3. Resolution - 0.15 pgmS/s; sample used for the study: thiophene Response Time - PWHM: 0.95 sec Selectivity - versus HCs is > 106 Precision - < 3% for 10 runs Stability - drift < 5% per day Interferents - < 0.3% for nitrogen, phosphorous, chlorine compounds Linear Dynamic Range - 105(concentration range with linear response) Summary of Key Technical Results

  4. Thiophene Response with PKI and SCD Detectors Amount Injected 0.5 uL Concentration: 200 ppm Thiophene Split Ratio: 100:1 PWHM=1.2 s Signal/Noise = 5400 Peak Asymmetry factor at 5 %=1.3 Amount Injected 0.5 uL Concentration: 200 ppm Thiophene Split Ratio: 100:1 PWHM=1.4 s Signal/Noise = 4700 Peak Asymmetry factor at 5 % =1.22 200 600 500 SCD Signal(mV) PKI Detector Response 100 400 300 2.22 2.26 2.30 2.20 2.25 2.30 Time (min)

  5. Sulfur Compound Standard from Sievers: Results with PKI Detector 3 1 Dimethyl Sulfide 2 Ethyl Methyl Sulfide 3 Thiophene 4 Toluene Solvent PKI Detector Response 2 1 4 Time(min)

  6. 0 380 360 5 340 2 PKI Detector Response 3 320 1 6 7 8 4 300 280 2 4 6 8 10 12 Gasoline Sample 0 Unidentified Sulfur Compound 1 Thiophene 2 2-Methyl Thiophene 3 3-Methyl Thiophene 4 2-Ethyl Thiophene 5 Alkyl Thiophene 6 BenzoThiophene 7 Methyl Benzothiophene 8 Dimethyl Benzothiophene Time(min)

  7. Gasoline Response with PKI and SCD Detector 7 1 1 Thiophene 2 2-Methyl Thiophene 3 3-Methyl Thiophene 4 Alkyl Thiophene 5 Benzo-Thiophene 6 Methylbenzo-Thiophene 7 Dimethyl Benzo- Thiophene 5 6 PKI Detector Response 2 4 3 1 Thiophene 2 2-Methyl Thiophene 3 3-Methyl Thiophene 4 Alkyl Thiophene 5 Benzo-Thiophene 6 Methylbenzo-Thiophene 7 Dimethyl Benzo- Thiophene 5 7 1 6 2 4 SCD Signal(mV) 3 Time(min)

  8. Long –Term Stability - Gasoline

  9. Total Sulfur Content:195 ppm Sample size: 0.005 uL Analysis Time: 0.1 min 500 400 300 0.05 0.10 0.15 0.20 Time(min) Total Sulfur Measurement in Gasoline PKI Detector Response

  10. Resolution - 0.15 pgmS/s; sample used for the study: thiophene Response Time - PWHM: 0.95 sec Selectivity - versus HCs is > 106 Precision - < 3% for 10 runs Stability - drift < 5% per day Interferents - < 0.3% for nitrogen, phosphorous, chlorine compounds Linear Dynamic Range - 105(concentration range with linear response) Summary of Key Technical Results