Basic Instrument Tools & Tackle Introduction

Basic Instrument Tools & Tackle Introduction. Tools & Tackle: A tool or device is a piece of equipment that most commonly provides a mechanical advantage in accomplishing a physical task. the most basic tools are simple machines. Example: Hammer, calibrator etc.

libitha
Download Presentation

Basic Instrument Tools & Tackle Introduction

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Basic Instrument Tools & Tackle Introduction Tools & Tackle: A tool or device is a piece of equipment that most commonly provides a mechanical advantage in accomplishing a physical task. the most basic tools are simple machines. Example: Hammer, calibrator etc Instrumentation and Control

  2. Basic Instrument Tools and Tackles Lists List of Instrument Tools & Tackles • Multimeter • Milliamps Source • Pressure Calibrator • HART communicator • Wrenches • Pliers • Allen keys • Crimping tools • Fixed spanners and so on.., Instrumentation and Control

  3. What do meters measure • A meter is a measuring instrument. • Anammetermeasures current, • Avoltmetermeasures the potential difference (voltage) between two points, and • Anohmmetermeasures resistance. • Amultimetercombines these functions, and possibly some additional ones as well, into a single instrument. • Before going in to detail about multimeter, it is important to have a clear idea of how meters are connected into circuits. • Diagramsin Next Slides show a circuit before and after connecting an ammeter. Instrumentation and Control

  4. What do meters measure • To measure current, the circuit must be broken to allow the ammeter to be connected in series. • Ammeters must have a LOW resistance . • All the current flowing in the circuit must pass through the ammeter. • Meters are not supposed to alter the behavior of the circuit, or at least not significantly, and it follows that an ammeter must have a very LOW resistance. Instrumentation and Control

  5. What do meters measure • To measure potential difference (voltage), the circuit is not changed: the voltmeter is connected in parallel . • voltmeters must have a HIGH resistance . • The voltmeter is connected in parallel between the two points where the measurement is to be made. • Since the voltmeter provides a parallel pathway, it should take as little current as possible. • In other words, a voltmeter should have a very HIGH resistance. Instrumentation and Control

  6. What do meters measure • To measure resistance, the component must be removed from the circuit altogether • Ohmmeters work by passing a current through the component being tested • Ohmmeters work by passing a small current through the component and measuring the voltage produced. • If we try this with the component connected into a circuit with a power supply, the most likely result is that the meter will be damaged. • Most multimeters have a fuse to help protect against misuse. Instrumentation and Control

  7. Multimeters • A multimeter or a multitester is an electronic measuring instrument that combines several functions in one unit. • The most basic instruments include an ammeter, voltmeter and ohmmeter. • Analog multimeters are sometimes referred to as "volt-ohm-meters", abbreviated VOM. • Digital multimeters are usually referred to as "digital-multi-meters", abbreviated DMM. • A multimeter can be a handheld device useful for basic fault finding and field service work or a bench instrument • Such an instrument will commonly be found in a calibration lab and can be used to characterize resistance and voltage standards or adjust and verify the performance of multi-function calibrators. Instrumentation and Control

  8. Analogue Multimeter • An analogue meter moves a needle along a scale. Switched range analogue multimeters are very cheap. The meter movement is delicate and dropping the meter is likely to damage it. • Each type of meter has its advantages. Used as a voltmeter, a digital meter is usually better because its resistance is much higher, 1 M ohm or 10 M ohm , compared to 200 ohm for a analogue multimeter on a similar range. • On the other hand, it is easier to follow a slowly changing voltage by watching the needle on an analogue display. • Used as an ammeter, an analogue multimeter has a very low resistance and is very sensitive, with scales down to 50 µA. More expensive digital multimeters can equal or better than this performance. • Most modern multimeters are digital and traditional analogue types are destined to become obsolete. Instrumentation and Control

  9. Digital Multimeter • Multimeters are designed and mass produced for electronics engineers. • Digital meters give an output in numbers, usually on a liquid crystal display. • The central knob has lots of positions. • We must choose which one is appropriate for the measurement you want to make. • If the meter is switched to 20 V DC, for example, then 20 V is the maximum voltage which can be measured. • This is sometimes called 20 V fsd, where fsd is short for full scale deflection. • For circuits with power supplies of up to 20 V, which includes all the circuits you are likely to build, the 20 V DC voltage range is the most useful. DC ranges are indicated by on the meter. • Sometimes, we will want to measure smaller voltages, and in this case, the 2 V or 200 mV ranges are used. Instrumentation and Control

  10. Hart Communicator Introduction • The HART Communicator is a menu driven system. Each screen provides a menu of options that can be selected as outlined above, or provides direction for input of data, warnings, messages, or other instructions. • When the HART Communicator is turned on, one of two menus will appear. • If the HART Communicator is connected to an operating loop, the communicator will find the device and display the Online Menu (see below). • If it is not connected to a loop, the communicator will indicate that no device was found. When you press OK (F4), it will display the Main menu. • The Main Menu provides the following options: • Offline–saves or retrieves transmitter configuration information. • Online–connects the communicator to a compatible device. Instrumentation and Control

  11. Hart Communicator Introduction • Frequency Device–The Frequency Device option displays the frequency output and corresponding pressure output of current-to-pressure transmitters. • Utility–The Utility option provides access to the contrast control for the HART Communicator LCD screen and to the auto poll setting used in multidrop applications. • Once selecting a main menu option, the HART Communicator provides the information you need to complete the operation. • The Online Menucan be selected from the main menu as outlined above, or it may appear automatically if the HART Communicator is connected to an active loop and can detect an operating transmitter. • On-line mode is used for direct evaluation of a particular meter, reconfiguration, changing parameters, maintenance, and other functions. Instrumentation and Control

  12. Hart Communicator Keys • The keys of the HART Communicator include action, function, and alphanumeric and shift keys. Action Keys • ON/OFF Key Use this key to power the HART Communicator. • When the communicator is turned on, it searches for a transmitter on the 4–20 mA loop. • If a device is not found, the communicator displays the message, “No Device Found. Press OK.” • If a HART-compatible device is found, the communicator displays the Online Menu with device ID and tag. Instrumentation and Control

  13. Hart communicator Keys Directional Keys • Use these keys to move the cursor up, down, left, or right. • The right arrow key also selects menu options, and the left arrow key returns to the previous menu. HOT Key • Use this key to quickly access important, user-selectable options when connected to a HART-compatible device. • Pressing the Hot Key turns the HART Communicator on and displays the Hot Key Menu. • See Customizing the Hot Key Menu in the HART Communicator manual for more information. Instrumentation and Control

  14. Hart communicator Keys • Use the four software-defined function keys, located below the LCD, to perform software functions. On any given menu, the label appearing above a function key indicates the function of that key for the current menu. As you move among menus, different function key labels appear over the four keys. • For example, In menus providing access to on-line help, the label may appear above the F1 key. Simply press the key to activate the function. • In menus providing access to the On-line Menu, the label may appear above the F3 key. Simply press the key to activate the function. Instrumentation and Control

  15. Hart communicator Keys Alphanumeric and Shift Keys • The alphanumeric keys perform two functions: the fast selection of menu options and data entry. Data Entry • Some menus require data entry. Use the alphanumeric and shift keys to enter all alphanumeric information into the HART Communicator. If we press an alphanumeric key alone from within an edit menu, the bold character in the center of the key appears. These large characters include the numbers 0 through 9, the decimal point (.), and the dash symbol (—). • To enter an alphabetic character, first press the shift key that corresponds to the position of the letter you want on the alphanumeric key. Then press the alphanumeric key. • For example, to enter the letter R, first press the right shift key, then the “6” key (see Figure). Do not press these keys simultaneously, but one after the other. Instrumentation and Control

  16. Hart Communicator Fast Key Sequences • HART fast key sequences provide quick on-line access to transmitter variables and functions. Instead of stepping our way through the menu structure using the action keys, we can press a HART fast key sequence to move from the Online Menu to the desired variable or function. Onscreen instructions guide you through the rest of the screens. HART Fast Key Example • HART fast key sequences are made up of the series of numbers corresponding to the individual options in each step of the menu structure. • For example, from the Online Menu we can change the Date. Following the menu structure, press 1 to reach Device Setup, press 3 for Basic Setup, press 4 for Device Info, press 5 for Date. • The corresponding HART fast key sequence is 1,3,4,5. • HART fast keys are operational only from the Online Menu. If we use them consistently, we will need to return to the Online Menu by pressing HOME (F3) when it is available. If we do not start at the Online Menu, the HART fast key sequences will not function properly. Instrumentation and Control

  17. Hart Communicator Fast Key Sequences Instrumentation and Control

  18. Hart Communicator Fast Key Sequences Instrumentation and Control

  19. Hart Communicator Connections • The HART Communicator Model 275 / 375 can interface with a transmitter from the control room, the instrument site, or any wiring termination point in the loop through the rear connection panel as shown in Figure next slide. • To communicate, connect the HART Communicator in parallel with the instrument or load resistor. The connections are non-polarized. • Avoid contact with leads and terminals. • Do not make connections to the serial port or NiCad recharger jack in an explosive atmosphere. • Before connecting the HART Communicator in an explosive atmosphere, make sure the instruments in the loop are installed in accordance with intrinsically safe or non incendive field wiring practices. • Both transmitter covers must be fully engaged to meet explosion proof requirements. • The HART Communicator needs a minimum of 250 ohms resistance in the loop to function properly. The HART Communicator does not measure loop current directly. Instrumentation and Control

  20. Hart Communicator Connections Instrumentation and Control

  21. Hart Communicator Menu tree Instrumentation and Control

  22. Hart Communicator Menu tree Instrumentation and Control

  23. Milliamp Calibrator Introduction • Basic functions Sources & Reads Mill volts, Volts, Milliamps, Sources Frequency, Simulates RTD values Sink & Reads Millivolt / Volts, Milliamps, Indicates Continuity. • portable calibrator is the result of many years of combined field experience that sets a new standard of portable multifunction calibrators. • calibrator can be used to calibrate any instrument in the industry Its dual operation of source & sink of signals make it useful in Labs & Process Plants. Instrumentation and Control

  24. Calibrator Instruction Turn-on • Each time you turn on the Model 434-KP the LCD will display all segments for about 1 second. It then displays the most recently selected scaling for the function selected by the selector knob. Turn-off • Press the POWER push-button to turn the 434-KP off. If AUTO-OFF is enabled the 434-KP will turn itself off after 30 minutes of inactivity. Choosing milliamp or percent • You may choose to display from 0.000 to 24.000 mA or -25.00 to 125.00%. Press the MILLIAMP / PERCENT push-button to toggle the display. When scaled in percent, 100.00% corresponds to 20.000 mA and 0.00% corresponds to 4.000 mA. Auto cal • To maintain accuracy, the Model 434-KP periodically recalibrates its measuring circuitry against internal references. While this is occurring the word CAL will appear on the display for less than 2 seconds. Instrumentation and Control

  25. Calibrator General Details General • Milliamp accuracy: ±(0.015% of reading + 0.002 mA) • Temperature effect: ±0.008%/°C based on 23°C ±25°C • Batteries: Six "AA" batteries (Alkaline supplied and recommended) • Battery life: • Milliamp Source & 2-wire Modes: Nominal 44 hours at 12 mA, 30 hours at 20 mA into 250 Ohm load • Other Functions: Nominal 75 hours • AC Adapters: Optional, 120 or 240 VAC, 50/60 Hz Low battery indication: "BAT" indication on the display at 6.5 V nominal, approximately 4 hours left • Operating temperature range: -5 to +130°F (-20 to +55°C) • Storage temperature range: -13 to +130°F (-25 to +55°C) • Relative humidity: 10 to 90%, non-condensing for 24 hours • Warm up time: 10 seconds to rated accuracy, 2 minutes to full • accuracy • Overall size: 6.23 x 3.27 x 1.94 inches (158.1 x 83.1x 49.3 mm) • Weight: 1lb, 2oz (0.5 kg) Instrumentation and Control

  26. Calibrator Specification Milliamp source • Ranges: 0.000 to 24.000mA with 1 micro amp resolution -25.00 to 125.00% of 4-20mA with 0.01% resolution • Typical drive capability: 1200 Ohms @ 20.000 mA • Power to drive receiver: Nominal 24 VDC • Compliance voltage: 0 to 500 Ohm Load: 12 V nominal @ 20.000 mA 500 to 1200 Ohm load: 24 V nominal @ 20.000 mA Power & measure 2-wire transmitters • Ranges: Same as Milliamp Source • Output current: up to 24.000 mA • Typical drive capability: 1200 Ohms @ 20.000 mA • Power to external 2-wire transmitter: Nominal 24 VDC Read milliamps • Ranges: Same as Milliamp Source • Voltage burden: 0.9 V at 4 mA, 1.2 V at 20 mA,1.4 V at 24 mA Instrumentation and Control

  27. Calibrator Specification Read transmitter test jacks or across loop diodes • Ranges: Same as Milliamp Source • Over voltage protection: Fuse, 5 x 20 mm, 250 mA, 250 V • Input impedance: 15 ohms Simulate 2-wire transmitters • Ranges: 1.000 to 24.000 mA • Loop voltage limits: Minimum, 2 VDC; Maximum 50 VDC Measure DC Volts • Range: -100.0 to +100.0 DC Volts • Accuracy: ±0.5% of Full Scale Reading from 0.0 to 100.0V • Input resistance: > 1 Meg ohm • Source resistance effect: 0.01% per 100 Ohms Instrumentation and Control

  28. Universal Calibrator Applications APPLICATION : • Calibrations of instruments  • Commissioning of  field instruments • Repair & Maintenance of field instruments • Use as millimeter with limited facilities USER INDUSTRIES : • Chemical Process Industries • Cement Plant • Paper & Pulp Industries • Petrochemical Industries • Textile Industries  • Fertilizer Industries • Pharmaceutical Industries • Instrument manufacturing unit • Automobile Industries • Instrument Calibration Labs Instrumentation and Control

  29. Introduction about Calibrator • MC5-IS is an Intrinsically Safe, documenting, All-In-One Multifunction Calibrator with calibration capability of pressure, temperature, electrical and frequency signals. • MC5-IS performs automatic calibration of electrical and temperature process instruments. • MC5-IS also communicates with HART field instruments. • MC5-IS represents the state of the art in accuracy, adaptability and all-round usability. The Upper Panel The upper panel has 5 places for the following modules/connectors: External Pressure Modules MC5-IS has a connector for External Pressure Modules (EXTs). The connector is located on the right hand side of the upper panel and is marked with PX1 in a sticker on the upper panel. Instrumentation and Control

  30. Calibrator Introduction Internal Pressure Modules • Up to three Internal Pressure Modules may be installed in MC5-IS.One of them may be an internal barometric module. The connectors for Internal Pressure modules start from the second connector on the left. The possible Barometric Module is always located as second from right and it measures the barometric pressure through a connection in the back panel of MC5-IS. Normally nothing need to connected to the barometric pressure module’s connector. • Internal pressure modules are marked with P1 … P3. • The recommended pressure medium for all internal pressure modules is clean air. • Clean non-corrosive liquids may optionally be used in modules with a measuring range of 20 bar/300 psi or more. • Avoid spilling liquid on MC5-IS when connecting/disconnecting pressure hoses to/from pressure modules. • To avoid damaging the calibrator, use hand tightening only when connecting the pressure measurement hoses (max. torque 5 Nm,approx. 3.6 lbf ft). • If the use of tools is required to secure the connection (typically pressure modules with a pressure range higher than 20 bar), apply the counterforce with a spanner on the connector body’s hexagonal part. • The overpressure protection of the internal pressure modules vents to the back of the calibrator. Instrumentation and Control

  31. Operational section and connection Instrumentation and Control

  32. Calibrator Front Panel Section Electrical Module (E module) • The E module can measure the following quantities: voltage, current and frequency. • It can also be used when counting pulses or detecting the state of a switch. • Additionally there is a possibility to sink current. • The E module also includes the optional HART modem. • This allows communication with an instrument with HART capabilities. • To ensure proper HART communication, make sure that the loop also includes a resistor with a resistance between 250 to 600 ohm or • that the impedance of the loop itself is at least 250 ohms. Instrumentation and Control

  33. Calibrator Front Panel Section Electrical and Temperature Module (ET module) • The ET module is specially designed for temperature instrument calibration needs. • It is not however restricted to only temperature instrument use because it can also generate voltage, frequency and pulses. Additionally the ET module is able to sink current. Instrumentation and Control

  34. Calibrator Measuring capability Measuring capabilities: • Low Voltage measurement and T/C measurement using either the internal reference junction or the Low Voltage connectors. • Resistance and RTD measurement. Generation/simulation capabilities: • T/C simulation using either the internal reference junction or the Low Voltage connectors. • Resistance and RTD simulation. • Voltage, frequency and pulse generation. • Current sink. Instrumentation and Control

  35. Calibrator Reference Junction Module • The T/C measurement/ simulation internal reference junction is an optional addition to the ET module. • It is specially designed for MC5-IS and therefore best suited for reference junction compensation when calibrating thermocouples or instruments connected to a thermocouple. Instrumentation and Control

  36. Calibrator Reference Junction Module • The Reference Junction Module is suited for all standard T/C plugs and stripped wires. • Open the fixing screw on the left side of MC5-IS before connecting the wires/plug to the Reference Junction Module. • Make sure to connect the wires/plug as the polarity is indicated on the Reference Junction Module. • Remember to tighten the fastening screw when the wires/plug are connected. • Hand tightening is adequate. • Do not pull out the wires/plug without first opening the fixing screw. Otherwise you might damage the contact surface of the connectors. • T/C measurement and simulation may also be done without the internal reference junction by using the Low Voltage terminals in the ET module. Instrumentation and Control

  37. Calibrator Display • MC5-IS has a reflective display. The resolution of the display is 240 x 320 pixels. • To quickly tune the contrast of the display: • Press and hold the contrast button down. • Use the up and down arrow keys to change the contrast. • The changed setting is automatically saved as default settings. Instrumentation and Control

  38. Calibrator Keyboard functions • The Cursor keys and the Enter key are located close to the upper left corner of the display. • The Cursor keys are used when moving the cursor on the screen. • They also have several special functions in certain situations, e.g. when tuning the contrast of the display. • The Enter key finishes the entering of values. Instrumentation and Control

  39. Calibrator Keyboard functions • The Contrast key together with the and cursor keys are used when setting the contrast of the display. • The Help key displays case sensitive help. • The On/Off key switches MC5-IS on and off. Press the On/Off key for about half-a-second to switch on/off. • This delayed function prevents accidental on/off switching of MC5-IS. • Pressing the +/- key toggles the sign of the entered numeric value. Note. The +/- key is applicable only in numeric fields. • The Decimal key adds the decimal point to the numeric value that is currently edited. Instrumentation and Control

  40. Calibrator Keyboard functions The Numeric Keys • The Numeric keys are not only used when entering numbers: • Keys 1 to 7 are used as menu selector keys. • Keys 0 and 8 are used to scroll through several pages of menu options. They may also be used when browsing through options in a pop-up list. • Key 9 can be used when accepting a selection or when finishing a data entry. The functionality of the 9 key is almost similar to the Enter key, except for one situation: • When entering numbers, the 9 key produces the number 9. To finish entering a number, you will have to use the Enter key or use the D/OK Function Key when available. Instrumentation and Control

  41. Calibrator General Description Instrumentation and Control

  42. Calibrator General Description Startup Procedure • Every time MC5-IS is started the Startup Procedure checks the functionality of the device by performing a self test. • If the self-test is passed successfully, some basic calibrator data is displayed. • After that MC5-IS automatically proceeds to Basic Mode. Basic Mode • In Basic Mode you can measure and generate/simulate signals. There are two separately configurable windows available. Basic Mode is often used for testing connections before starting the actual calibration procedure of an instrument. • Stepping and Ramping tools enable generating/simulating signals that vary with time. Maintenance • This main function handles calibrator configuration settings. Additionally there is the possibility to recalibrate MC5-IS (requires a password). Instrumentation and Control

  43. Calibrator User interface Instrumentation and Control

  44. Calibrator User interface • The Status Bar at the top of the display is visible all the time. It is divided into four main sections. • The first (leftmost) section displays the charge level of the battery. The battery symbol is replaced by a plug symbol ( ) if you are using the battery charger. • The second section displays the time and date. • The third section displays the temperature measured with the optional environment sensor, if the sensor is connected to MC5- IS. • The fourth section (rightmost) section displays additional information in the form of symbols, like: 1.An hourglass when MC5-IS is working on something that takes time. 2.A question mark when an error occurred. Note: that the fourth section is empty for most of the time. The symbols are visible only when needed. Instrumentation and Control

  45. Calibrator User interface • The Function Keys are located below the display. • The meaning of each Function Key varies depending on the situation. • The lower part of the display indicates what the Function Key stands for at the moment. • The Function Key Bar at the bottom of the display is visible all the time. • The meaning of the Function Keys varies depending on the situation. Instrumentation and Control

  46. Calibrator Menus • The Function Key for opening the menu is always D/Menu. The same key is used when closing the menu. • If a menu is not needed for the current subject, the fourth Function Key is used for other needs. • If the opened menu has several pages, the menu’s uppermost and/or lowermost item includes an up/down triangle. • In that case, use the 0 and8 keys to browse through the available menu pages. • A menu option is selected with the numeric keys 1 to 7. Selecting a menu option results in one of the following events Instrumentation and Control

  47. Calibrator Menus • An immediate action follows and the menu closes automatically,e.g. when selecting the Zero Pressure Module option in the picture above. • A pop-up list opens for selecting one of the available options. The current selection is displayed inside brackets in the menu. • Use the up and down keys, the 0 and 8 keys or the same numeric key that opened the pop-up list to scroll the list. • To select an option in the pop-up list, use either the enter key or the 9 key. To close the pop-up menu without selecting anything, press the right side key or the D/Close Function Key. • Another menu with new options replaces the previous menu. Sometimes the Function Keys can also open another menu. • In the previous picture, the Window 1 setup menu is opened. In this case Function Key B/Window 2 Setup and Function Key C/Others can be used for opening other menus. • A new window opens for, e.g. viewing additional information or for configuring the selected task. Instrumentation and Control

  48. Calibrator Display Area • The layout of the display area varies according to the needs of the active tasks/settings. The following pictures give an overview of typical elements seen in different display area layouts. Basic Measurement/Generation: • The display area is divided into two windows with informative texts and numeric measurement/generation values. • A border surrounding a numeric value indicates that the field is editable. It is, e.g. a generation field for entering generation values. • If several editable fields are visible, choose the active field with the cursor keys or the B/Field Function Key. Calibration: • The display area is divided into three windows during a calibration. • The first window displays data related to the instruments input signal. The second corresponding data related to the output signal. • The third window displays the error graph. The error graph is also seen among calibration result data. Instrumentation and Control

  49. Calibrator Ramping Configuration Window: • There are plenty of configuration windows in MC5-IS. The picture beside is the configuration window for Ramping settings. • The common thing for all configuration windows is that they reserve the whole display area for the configuration fields. • Use the cursor keys to move between fields. Instrumentation and Control

  50. Calibrator Table Tables: • Tables are used, e.g. when viewing calibration results in numeric format. Tables reserve the whole display area. • The tables are often larger than the display. In that case there are small arrows added to the table borders. They indicate that more information may be seen by using the arrow keys. • The up and down arrow keys scroll the list one line at a time. The 0 and 8 keys scroll the list one page at a time (if applicable). • Hint. • If the table has more columns than can be seen, use the numeric keys to quickly jump to corresponding column Instrumentation and Control