General Idea. Normalized QDC for each pixel Data set s : PP158 from 2009 (~3.5 Mevents ) BeBe from 201 3 (13 +30 GeV , ~7М events) We have histo for every pixel´s count where we can check all suspicious pixels
E N D
General Idea • Normalized QDC for each pixel • Data sets: • PP158 from 2009 (~3.5 Mevents) • BeBe from 2013 (13+30GeV, ~7Мevents) • We have histo for every pixel´s count where we can check all suspicious pixels • Good pixel: number of hits has to be in range of 8 sigma from average hits of neighbor pixels
Determination of good pixels For counting average hits from neighbor pixels, only GOOD pixels are included; Zero-, Over- and Under-Count are excluded!
Conclusion • Right wall still decent: 20 dead and 41 suspicious pixels • Left wall seriously damaged: 75 dead and 232 (!) suspicious pixels • Most probably reason: broken or unglued PMTs • Cannot be justify without opening cassettes • Almost all cassettes in the left wall need repairing! • Serious job: material (PMTs, scintillators), men work… • Question: do we need readout upgrade in this situation?