Results on the 100 R&D EndCap crystalsmeasured in RCCERN on ACCOCE2 E. Auffray, P. Lecoq CERN EP-CMA
We have received 100 polished Endcaps crystals in June2000 from Bogoroditsk: • all have been tested in CERN RC • Visual Inspection, • Characterised with ACCOCe2 : 1st operation of the 2nd Accos machine in CERN RC
Longitudinal Transmission 360nm 420nm 620nm
Longitudinal Transmissionat 420nm 100 Endcaps Crystals Barrel Crystals Batch09
Light Yield Endcaps polished crystals Barrel polished crystals from Batch09
Light Yield Non uniformity FNUF Endcaps polished crystals Barrel polished crystals from Batch09
Lateral Irradiation at saturation Endcaps polished crystals Barrel polished crystals from Batch09
Considerations about Endcap crystals • After 2 years of R&D in Bogoroditsk • They have demonstrated that they can produce endcap crystals of similar quality than barrel crystals • Could be OK for at least 25% of endcap crystals • They have not yet achieved the rad. hardness specs for the most exposed endcap crystals