Research Update. Sukhi Basati PhD Candidate Laboratory for Product and Process Design Director: Prof. Andreas Linninger 3/31/2011. Microcontroller Unit. Shunt. Subcutaneous. 3.4 mm. 350 µm. Thesis aims. An implantable system for intracranial volume measurements in hydrocephalus.

naiara
Download Presentation

Research Update

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Research Update Sukhi Basati PhD Candidate Laboratory for Product and Process Design Director: Prof. Andreas Linninger 3/31/2011

  2. Microcontroller Unit Shunt Subcutaneous 3.4 mm 350 µm Thesis aims An implantable system for intracranial volume measurements in hydrocephalus • Develop sensor based on an impedance technique to measure volume • Computer aided design • Microfabricate sensor with biocompatible materials • Electronics design and fabrication • Bench-top analysis to optimize parameters • Validate measurements in an acute hydrocephalic animal model • Animal protocol to induce HC and implant sensor into hydrocephalic rat • Develop chronic, implantable measurement device • Wireless data communication and microcontroller implementation

  3. Spring 2011 research aims • Initiate HC experiments on west campus • Priority: High • Induce HC in weanling rats (only need n=10) (experiments should not last longer than end of February) • 2nd surgery involves implanting sensor into hydrocephalic ventricles and CSF removal • Submit acute volume measurement paper • Priority: High • JNS - pediatrics • Design and fabricate long term electronics consisting of surface mount components (microcontroller + wireless transmitter) • Priority: Medium • Write draft of HC simulation in weanling rats • Priority: Medium

  4. Instrumentation Pump CSF 4.8 mm 500 µm Acute HC Measurements Lambda Bregma? • Some problems… • Still cannot remove CSF. • We observe CSF leakage. • Is it possible that ventricle collapses, and we are trying to remove fluid from ECS?

  5. 3 week old rat, 21 day post kaolin – H1115 New Coordinates 7.3 mm 8.5 mm 2 mm Old Coordinates 5 mm 1 cm T2w; RARE (FSE); T2eff 54ms; TR 1640ms; NA 6; TT 5m14s FOV 3.0x3.5cm2; SLTH 1mm; SSDI 2mm; matrix 256x256

  6. 3 week old rat, 21 day post kaolin – H1115 9 mm 7 mm 1 cm 1 cm T2w; RARE (FSE); T2eff 54ms; TR 1640ms; NA 6; TT 5m14s FOV 3.0x3.0cm2; SLTH 1mm; SSDI 2mm; matrix 256x256

  7. Discussion with Pat McAllister • Works a lot with rat model of HC • Currently comparing kaolin induced HC with kaolin failures, however he usually does not get many failures • He injects Kaolin via surgical cutdowns • >90% successful induction • You also can see where your injecting • Advice given: • Percutaneous Injection • Keep needle bevel facing upwards. • We can try holding the animal, as to keep neck flexed. • Make sure we puncture antlanto-occipital membrane. • We can try only implanting the needle until CSF leaks out, and then we connect the kaolin-filled syringe. • A neurosurgeon in Cleveland only had 10% induction rate! • Just keep practicing Adjunct ProfessorProfessor of Neurosurgery Director of Basic Hydrocephalus Research • Acute Shunting • after implanting shunt in HC animal, you don’t see ventricular collapse. • at most 30% decrease in size • it takes 3-5 days for ventricular volume to decrease if shunt is successful. • and even then the size does not come back to normal • He has previously used a syringe pump in HC kittens • he even used a rate of 0.3 mL/hr over 20-25 minutes • Offered to send rats, wouldn’t mind collaborating…

  8. Location of Sensor • Shunt failure occurs often due to misplacement during surgery. • Surgery is often followed by MRI to confirm location within ventricles • It is a serious problem in animal models of HC. If shunt placement was successful, the ventricles drain. • Even though my device is designed for volume measurements, at least it indicates when you are in a fluid filled region… Ventricle location

  9. Rat Sensor Simulations Purpose: Sensor Simulations on HC rat geometry will provide design decisions such as electrode placement and catheter angle. This data will be input to optimization algorithms. • Model Geometry: • Rat HC ventricles, ranging from 21 day HC to 7 day HC. • Brain Parenchyma modeled as simple ellipsoid. • Next steps: • Work with Forrest on generating functions.

  10. Future Work • Purchase pressure transducer • Simulate Rat CSF dynamics and ventricular expansion • Develop power control circuit • Prepare DMD talk • Begin writing dissertation