Contrast Echocardiography

Contrast Echocardiography. Introduction. US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents: Agitated saline with or without Indocyanine green. Agitated Saline. Agitating a solution of saline between two 10-mL syringes

orea
Download Presentation

Contrast Echocardiography

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. ContrastEchocardiography

  2. Introduction • US contrast agents first used- mid 1970 • Gas containing microbubbles. • First generation Contrast Agents: Agitated saline with or without Indocyanine green.

  3. Agitated Saline • Agitating a solution of saline between two 10-mL syringes • Each of which contains 5 mL of saline and 0.1 to 0.5 mL of room air • Forceful agitation through a three-way stopcock creates a population of microbubbles • ‘Dose’- 1- 5 ml

  4. Ideal contrast agent • Non-toxic • Intravenously injectable • Has to behave similarly to blood • Crosses pulmonary filter • Resistant to intravascular and intra-cardiac pressures • Stable throughout during the exam • Improve the Doppler signal-to-noise ratio

  5. Recent microbubble formulations

  6. Low surface tension. • Resistant to ultrasound destruction. • Slowly diffusing, insoluble, high molecular weight gases. • 1.1 – 8 µm size, 5х10⁸ to 1.2x 10¹⁰ microbubbles per millilitre • Single injection provide contrast effect for 3- 10 min. • Safe – 4 deaths after 2 million use • Contra indications • Known Rt to Lt shunts • Known hypersensitivity

  7. Ultrasound Interaction with Contrast Agent >0.3 MI <0.3 MI

  8. Machine settings • Dedicated contrast specific presets • Mechanical Index; Power of US beam Peak Negative acoustic pressure Transmitted Frequency Routine B mode uses – High MI - 0.9 to 1.4 Low MI < 0.3

  9. Contrast Destruction • High Mechanical Index • High Frame rate • Focal zone • Near field

  10. Fundamental Harmonic

  11. Continuous Imaging Low Mechanical Index High Mechanical Index

  12. Intermittent imaging • Triggered to ECG • In between imaging, no ultrasound energy is delivered. • Allows time for restitution of contrast effect. • Analysis of wall motion-not possible. • Evaluation of myocardial perfusion. Continuous low MI imaging. • Wall motion analysis in real time. • Used for cavity opacification. • Detection of very low concentration of myocardial contrast.

  13. Intermittent Triggered Imaging oIntermittent Imaging

  14. Power Spectrum Motion of the bubbles & their resonance in a stationary field

  15. Clinical Applications

  16. Detection and Utilization of Intracavitary contrast • Enhanced visualization of the LV endocardial borders • Improve reproducibility for wall motion analysis and volumetric measurements • Detection or exclusion of • Intracavitary thrombus • Ventricular noncompaction • Atypical forms of HCM (Apical) • Abnormal communication to the ventricular chamber

  17. Exclusion of Thrombus

  18. LV Thrombus

  19. Ventricular noncompaction

  20. Spectral Doppler Enhancement • Low concentrations of contrast agents • Enhancing the tricuspid regurgitation jet • Pulmonary vein flow • Increasing intensity of a relatively weak aortic stenosis jet

  21. Shunt Detection • Right-to-left shunts - agitated saline - agent of choice • Atrial septal defects of all types • Patent foramen ovale - Valsalva and cough • Pulmonary arteriovenous malformations - 5 to 15 cycles • Larger ventricular septal defects during diastole • Left SVC • Left-to-right shunt • Negative contrast effect

  22. Right-to-left shunt Negative contrast effect

  23. Atrial septal aneurysm with PFO

  24. Persistent Left SVC

  25. Myocardial Perfusion Contrast • First recognized in the 1980s • Preserved contrast effect in the myocardium - evidence of microvascular integrity and blood flow to the area • Analysis of myocardial flow - Time of appearance curve • Multiple time appearance curve analyses - necessary • Time of appearance curve requires a bolus effect • wait 10 minutes • purposeful destruction of the contrast agent - burst of high intensity (high mechanical index) ultrasound • Targeted to different regions of interest • Performed under basal conditions & after vasodilator stress

  26. Time of appearance curve

  27. α is directly related to myocardial blood volume • β is related to flow rate • The product of α and β- proportional to myocardial blood flow • Vasodilator results in an increase in flow velocity in those areas not perfused by a stenosed artery • Appearance of the contrast curves - differ in the normal and diseased beds

  28. Transcatheter alcohol septal ablation • Performed for the Rx of HOCM. • Catheter is placed in the 1st septal perforator of LAD. • Controlled myocardial infarction for reduction of proximal septal mass. • Before alcohol injection, diluted US contrast agent is injected to the selected artery. • To ensure- no contrast reflux. • To confirm the presence and size of vascular bed.

  29. Transcatheter alcohol septal ablation

  30. Attenuation & Shadowing

  31. Papillary Muscle Shadow

  32. Colour Artifact

  33. Competitive Flow May be confused with a true negative contrast effect due to an atrial septal defect

  34. Prominent eustachian valve and margination of contrast-enhanced blood flow May be confused with a true negative contrast effect due to an atrial septal defect

  35. Strain Rate Imaging

  36. Introduction • Evaluation of a myocardial region with reference to an adjacent myocardial segment. • Deformation analysis- analysis of ventricular mechanics or shapes during cardiac cycle. • Myocardial strain, strain rate, torsion. • Strain- percentage thickening or deformation of the myocardium during the cardiac cycle. • Change of strain per unit of time is referred to as strain rate

  37. Strain & Strain rate

  38. Strain calculated in three orthogonal planes- representing longitudinal, radial, circumferential contraction. • Negative strain- shortening of segment. • Positive strain- lengthening of segment

  39. Methods

  40. SR- Doppler tissue imaging

  41. Speckle tracking • ‘Speckles’ are small dots or groups of myocardial pixels that are created by the interaction of ultrasonic beams and the myocardium. • Considered as acoustic fingerprint for that region. • This enables to judge the direction of movement, the speed of such movement, and the distance of such movement of any points in the myocardium.