X-Ray Webinar 3/27/2013. X-Ray update. Topics of the day. Online preventative maintenance. Calibration backup SUI updates Schick Update. Yearly PM now live on-line also for X-Ray units. Required for 3D units covered by service contracts. - Same log-in as for CAD/CAM units.

pleasance
Download Presentation

X-Ray Webinar 3/27/2013

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. X-Ray Webinar 3/27/2013 X-Ray update

  2. Topics of the day • Online preventative maintenance. • Calibration backup • SUI updates • Schick Update

  3. Yearly PM now live on-line also for X-Ray units • Required for 3D units covered by service contracts - Same log-in as for CAD/CAM units - Just choose “Imaging Products” on the left - Look for the correct manual for your unit. - Print out the checklist if you wish

  4. Yearly PM now live on-line also for X-Ray units

  5. Yearly maintenance (1) • Electrical/Mechanical check: (Follow the manual) • Check serial number of unit (bottom right of unit) • Check all buttons and lights and audible tone(s).(Dead-man switch release button).

  6. Yearly maintenance (2) • Electrical check: (Follow the manual) • Measure Power outlet adequacy (voltage drop). • Tip: Can be measured later while you do the dose check in calibration.

  7. Yearly maintenance (3) - Depending on machine there can be one (Galileos) to four (XG3D Ceph) lasers. • Attention: The Galileos laser can be obstructed • by headrest (if present). • (Take it out now, it must be removed for calibration • anyway). • Check lasers

  8. Yearly maintenance (4) All good so far ? Relax… … a minute

  9. Yearly maintenance (5) • Before you start: • Make sure you have all calibration phantoms !Pin Phantom (XG3D) • Geometry Phantom (XG3D, Galileos) Distortion Phantom (Galileos)

  10. Yearly maintenance (6) • Ceph Phantom (XG3D Ceph) • Constancy test (XG3D, Galileos) 2D !

  11. Yearly maintenance (7) • Go to sironasupport.comVideo page:

  12. Yearly maintenance (8) • Make a backup of the calibration data !!! • If you don’t and the calibration fails, the unit will not work anymore.

  13. Yearly maintenance (9) • Go through ALL steps start to finish, regardless if the symbols are green, yellow or red!(Except radiation field). Galileos: Tool can be downloaded from web-site: Sironasupport.com

  14. Yearly maintenance (10) • Perform the Acceptance or Constancy test: • Sidexis -> Utilities -> Acceptance/Constancy test

  15. Coming up: SUI updates (Software update image) • One Moment………... Ready?

  16. SUI updates (1) • “SUI update” refers to the installation/upgrade of the “firmware” of any XG/Galileos X-Ray system. Sidexis Manager -> SiXabcon • Attributes -> software update. • Select SW package you want to add. 4.07…4.12

  17. SUI updates (2) Do not pay too much attention to “red” and “green” BUT: Boards that show V00.00as their actual version might have an issue. There is a third way to update units: Modul-update which can be advantageous. ALWAYS IGNORE: Timout “vermutet” (expected). Wait until update is finished.

  18. SUI updates (2) • When do you need a SUI update? 3D machines: With (almost) every (PC) software update. Spareparts!- come with “random” software loaded, not necessarily matching the unit. Unit old – sparepart new.Unit brand new – sparepart older. Attention: Sometimes SUIs have to be upgraded in steps…

  19. SUI updates (3) • What to do? • Update/downgrade the spare part to unit Software? • Update the unit to (newer) spare part Software? CALL SIRONA Technical service!! This can both be dangerous!!!Older SW might not support newer units/parts, upgrading to newest SUI might lead to malfunctions with (older) PC software.

  20. Thank you !!