1 / 25

X Ray Diagnostics

X Ray Diagnostics. LCLS FAC Meeting Oct. 27, 2005. FEE Layout. Slit. Diagnostics Package. Be Mirrors 2 & 3. Solid Attenuator. Collimator 1. Fast close valve. Gas Attenuator. Ion Chamber. SiC Mirror 1. Ion Chamber. Diagnostics Package. SiC Mirror 2. Damage Considerations.

lynnea
Download Presentation

X Ray Diagnostics

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. X Ray Diagnostics LCLS FAC Meeting Oct. 27, 2005

  2. FEE Layout Slit Diagnostics Package Be Mirrors 2 & 3 Solid Attenuator Collimator 1 Fast close valve Gas Attenuator Ion Chamber SiC Mirror 1 Ion Chamber Diagnostics Package SiC Mirror 2

  3. Damage Considerations Dose along the beam line for different materials at Gaussian peak (under normal illumination) Shown is the maximum dose (over Ephoton=827 to 8267eV) SiC melt Si melt Dose (eV/atom) (maximum over 827-8267eV) B4C melt Be melt z (m from end of undulator)

  4. Dose obtained at TTF

  5. TTF VUV Beamline

  6. Damage experiment chamber TOF ion spectrometer Sample holder Visible spectrometer Microscope

  7. SiC thin film, 1 shot, 8 x melt

  8. SiC thin film, 10000 shots, ~ 1 x melt Emission spectra shows plasma was formed

  9. SiC thin film, 10000 shots, 10% melt No spectra can be distinguished above background

  10. Si/C mirror reflects for one shot First shot Second shot

  11. Desired FEL Measurements

  12. Desired Spontaneous Measurements

  13. FEE Cartoon Start of Experimental Hutches 5 mm diameter collimators Windowless Ion Chamber DiagnosticPackage Spectrometer / Indirect Imager mirror Solid Attenuator High-Energy Slit Total Energy Calorimeter e- WFOV Direct Imager Gas Attenuator FEL Offset mirror system Spectrometer camera Windowless Ion Chamber Muon Shield

  14. Redundant Commissioning Instrumentation

  15. Photoelectrons generated by 0.01% FEL Wide Field of View Direct Imager Single shot measurement of f(x,y), x, y ,u Camera Scintillators

  16. Indirect Imager Single shot measurement of f(x,y), x, y, u Multi shot measurement ofl Angle selects energy and attenuation

  17. B4C/SiC Test Multilayers Fabricated • 40 layer pairs • G = 0.7 • P = 6 nm Data at Cu Ka, 8 keV

  18. Thermal diffusion calculations performed Total Energy Calorimeter Single shot measurement of f(x,y), x, y, u t = 300 ms t = 100 ms Nd0.8Sr0.2MnO3 Cold Si substrate CMR Sensor array 100 pixels t = 0 Xray Beam T Cooling ring T, ms 5 0 Sample CMR at LLNL for etching and R(T) measurements

  19. Sensor Development: Nd0.8Sr0.2MnO3

  20. 10 cm Ion Chamber Single shot, non destructive, measurement of x’, y’, x, y ,u Segmented cathodes for position measurement 1 torr Imaging of optical emission for position measurement

  21. Measuring K with photons On axis: But Effect of changing K

  22. Near-Field calculations of detuned single undulator segments Mean, keV Photons/pulse/20 eV First and Last 2mm x 2mm First 8.2658 Last 8.2362 First 2 x 2 mm Last 1 x 1mm First 8.2658 Last 8.2651 Last has K(1+10-4) First 2 x 2 mm Last 1 x 1mm First 8.2658 Last 8.2645 Photon Energy, keV

  23. Single Shot 8 keV Spectrometer Single shot measurement of l 20 m downstream 1.4991 Å Sputter-sliced SiC / B4C multilayer P = 20 nm N = 1x104 D = 200 mm 1.4994 Å 33 mm thick 200 mm Problems: Don't have 20 m. Signal degradation due to grating thickness (coupled waveguides.) Only possible at high photon energies. Beam

  24. Only one diffracted order Other candidates for Single-Shot Spectrometers Mosaic Graphite crystal Asymmetric cut multilayer

  25. Summary • X-ray characterization for LCLS is challenging • Dose limits materials choice • Large range of signal levels • Some measurements require precision near r = 10-4 • Redundant diagnostics techniques will be used to measure pulse energy • Total energy calorimeter • Direct Imager • Indirect Imager • Ion chamber • Much work needed on single-shot spectral measurements • TTF damage experiment to be analyzed