1 / 20

TCT measurements with irradiated strip detectors

TCT measurements with irradiated strip detectors Igor Mandić 1 , Vladimir Cindro 1 , Andrej Gorišek 1 , Gregor Kramberger 1 , Marko Milovanović 1 , Marko Mikuž 1,2 , Marko Zavrtanik 1 1 Jožef Stefan Institute, Ljubljana, Slovenia

roxy
Download Presentation

TCT measurements with irradiated strip detectors

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. TCT measurements with irradiated strip detectors Igor Mandić1, Vladimir Cindro1, Andrej Gorišek1,Gregor Kramberger1,Marko Milovanović1, Marko Mikuž1,2, Marko Zavrtanik1 1Jožef Stefan Institute, Ljubljana, Slovenia 2 Faculty of Mathematics and Physics,University of Ljubljana, Slovenia Most measurements made by diploma student MitjaKrnel

  2. Introduction: • TCT with focused IR laser light • light beam directed on the surface of strip detector: Top-TCT • “Spaghetti” detectors produced by Micron: • - p-type, FZ, 300 µm thick, 4x4 mm2 • - strip pitch: 80 µm • - implant width: 20 µm, DC coupled • Spaghetti: all strips connected on one side •  only one wire bond  faster work… •  E field as in strip detector, weighting field as in pad detector • Spaghetti Type 1 500 µm of implant not covered by metal • detectors irradiated to 1∙1015 , 2∙1015 and 5∙1015 n/cm2 • measurements after several annealing steps at 60°C up to total annealing of • 5120 minutes at 60°C • Motivation: check the uniformity of response •  withSpaghetti T1 laser measurement possible also under implants

  3. Setup detector HV Peltier controller beam diameter in the silicon FWHM~7 mm y table optical fiber & focusing system Cooled support Detector box width of light pulses ~ 100 ps , repetition rate 200 Hz cooling pipes z table x table Bias-T 1GHz oscilloscope fast current amplifier Laser Laser driver trigger line

  4. Detector Last 500 µm of implants not covered with metal 4 mm • All strips connected together • E field as in strip detector • weighting field as pad • detector •  signal is a sum of • currents induced on all • strips

  5. Signals Guardring • before irradiation • bias = 100 V (depleted) No metal metal scan laser spot across the surface Signals induced by laser beam at different locations: Before irradiation Bias = 100 V

  6. Before irradiation: scanoverdetectorsurface (step x: 2.5 µm, y: 50 µm) Bias = 100 V (depleted) End of implants Charge = integral of current pulse (offline) Guardring Metal

  7. Before irradiation Bias = 100 V Reflection from metal guard ring no metal Some signal also when beam on metal  tails of light spot No significant difference between metal and no-metal! metal

  8. Irradiated detector Φeq= 5·1015 n/cm2 Bias = 1000 V Annealed 5120 minutes at 60°C multiplication expected Large signals close to guard ring End of implants

  9. Irradiated detector Φeq= 5·1015 n/cm2; Bias = 1000 V; Annealed 5120 minutes at 60 C  high signals also near the guard-ring parallel to the strip strips Guard ring

  10. Irradiated detector Φeq= 5·1015 n/cm2 Bias = 1000 V Annealed 5120 minutes at 60°C Δt = 0.2 ns Δt = 0.6 ns Δt = 0.7 ns Step: 50 µm (dot colors correspond to line colors) guard ring v = 107cm/s Δt ~ 0.5 ns  v·Δt ~ 50 µm no metal metal The second peak from high field at the end of implant.

  11. Irradiated detector Φeq= 5·1015 n/cm2 ,Bias = 1000 V, annealed 5120 minutes at 60°C guard ring Pulses from scan over “no metal” part nometal metal Pulses from scan over “metal” Scan over “no metal” Scan over “metal” Complicated behavior near guardrings!

  12. Irradiated detector Φeq= 5·1015 n/cm2 , Bias = 1000 V Signals on edges increase with long term annealing  multiplication effects 80 min at 60°C 320 min at 60°C 1280 min at 60°C 5120 min at 60°C

  13. Irradiated detector: effect of annealing Φeq= 5·1015 n/cm2; Bias = 1000 V; implant Scan over non metalized implants (~ 0.5 mm from guard ring) 1000 V Scan over metalizedstrips ~ 1 mm from guard ring 1000 V •  more non-uniform with annealing multiplication at the edge of implants • effect increases closer to guard-ring

  14. Annealing Φeq= 5·1015 n/cm2 Bias = 1000 V implant • Annealing of CCE depends on • location •  Increase of charge with long term • annealing near implant edge

  15. Irradiated detector: annealing at lowerbiasvoltage 600 V Bias = 600 V; Φeq= 5·1015 n/cm2; implant Scan over non metalized implants ~ 0.5 mm from guard ring 600 V Scan over metalizedstrips ~ 1 mm from guard ring • smaller changes of charge variations with annealing • charge drops with reverse annealing  no (or less) multiplication

  16. Voltage scan 2e15, annealed 5120 min 5e15, annealed 5120 min Charge variations increase with bias voltage!

  17. Spaghetti detectors: • all strips connected  sum of signals from all strips measured • opposite polarity contributions from neighbor strips added •  variations of CCE dumped in spaghetti detectors! Hamamatsu detector irradiated to 1015 , ~5000 min at 60C, only one strip connected to amplifier: (see talk from Bari: https://indico.cern.ch/materialDisplay.py?contribId=5&sessionId=5&materialId=slides&confId=175330) Red: measured Blue: sum of signals from all strips at equivalent location: metalized strips to amplifier

  18. Summary • Top – TCT measurements with spaghetti detectors • large signals (multiplication) close to guard ring •  care should be taken to get realistic CCE (e.g. in test beam ….) • largest charge measured at edges of implants •  signs of charge multiplication (annealing behavior, bias dependence … ) • largest at the edge of implant • variations of collected charge across detector increase with multiplication •  variations are dumped in spaghetti diodes because of negative contributions • from neighbor strips •  test beam experiment can tell how problematic are these variations, if there • are dead regions where CCE falls below threshold …. • ( in HEP experiments most tracks are crossing the detector at an angle)

  19. AnnealingΦeq= 2·1015 n/cm2 Bias = 1000 V • larger non-uniformity after long • annealing • annealing of CCE depends on • location • increase of charge with long • annealing larger closer to • implant edge

  20. AnnealingΦeq= 2·1015 n/cm2 Bias = 1000 V metal • slightly different behavior at larger • distance from guard-ring

More Related