Summary of First Laser Measurements on Irradiated Detectors

Summary of First Laser Measurements on Irradiated Detectors. Nicholas Ptak Colin Parker 2/1/11. Introduction. So far, 7 irradiated detectors have been tested at different temperatures W17-BZ2 at –15C [4x10^12 p-stop,1x10^13 protons] W17-BZ4D at –10C [4x10^12 p-stop,1x10^13 protons]

swann
Download Presentation

Summary of First Laser Measurements on Irradiated Detectors

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Summary of First Laser Measurements on Irradiated Detectors Nicholas Ptak Colin Parker 2/1/11

  2. Introduction • So far, 7 irradiated detectors have been tested at different temperatures • W17-BZ2 at –15C [4x10^12 p-stop,1x10^13 protons] • W17-BZ4D at –10C [4x10^12 p-stop,1x10^13 protons] • W18-BZ2 at –15C [4x10^12 p-stop,1x10^13 protons] • W18-BZ4C at –10C and –15C [4x10^12 p-stop,1x10^13 protons] • W18-BZ4D at –15C [4x10^12 p-stop,1x10^13 protons] • W50-BZ2 at –15C [4x10^12 p-stop,1x10^13 protons] • W50-BZ4D at –10C [4x10^12 p-stop,1x10^13 protons]

  3. Farside Punchthrough • In most detectors, there seemed to be no punchthrough on the farside. The voltages on the near and far sides were very close to equal, and limits on the resolution of the measurements as well as other possible unaccounted for resistances made it difficult to show 0 current on the strip, and sometimes resulted in calculations of negative Reff. • Only 1 detector so far has shown punchthrough on the far side. The tests on the W17-BZ4D and W18-BZ4D detectors were done at two different temperatures and show the difference in Reff far with temperature • W50-BZ4D also showed a possibility for punchthrough

  4. Saturation • Some of the detectors tested seemed not to reach saturation at 200 volts • Including higher bias voltages should be considered for future tests

  5. Current vs Reff • It was also noted that the current vs Reff curves roughly followed a i^-0.5 dependence for the near side punchthrough

  6. Comparison with non-irradiated detectors • One test was done so far to compare irradiated with non-irradiated detectors • The current in the irradiated detector was lower and Reff higher • Neither detector reached saturation before 200V bias

  7. Summary • Farside punchthrough is only seen in some detectors • The detectors should be tested at higher bias voltages to look at saturation • The next tests will be done on the W50 zone 4s to compare punchthrough structures and in comparisons with non-irradiated detectors