MR- Microscopy of small specimen on a Human 7T-Scanner Andreas Berg

Zentrum f. Medizinische Physik u. Biomedizinische Technik & MR- Hochfeld - Zentrum. MR- Microscopy of small specimen on a Human 7T-Scanner Andreas Berg. Methodology : hardware. Quality control spatial resolution.

stan
Download Presentation

MR- Microscopy of small specimen on a Human 7T-Scanner Andreas Berg

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Zentrum f. MedizinischePhysik u. BiomedizinischeTechnik & MR-Hochfeld-Zentrum MR-Microscopyofsmallspecimenon a Human 7T-ScannerAndreas Berg Methodology: hardware Quality control spatial resolution 1.) Paraplegia model in rat spine 2.) osteoarthritis: cartilage bone intervert. Disc 3.) teeth and dental composites A.Berg, MR-Center 30.11. 2010

  2. ResultsQuality control: Spatial Resolution Standard whole body scanner: Phantoms: Test structures d > 1000 µm MR-microscopyinsert on 7T scanner: 1,5 cm A. Berg, Deutsches Patentamt, München, DE19904635 , 19.08.1999 Phantoms: teststructures d < 100 µm

  3. First results: Quality Control: spatial Resolution Silicon grid in Silicon oil TM = 87 min SNR = 16 Voxelsize: 50,9 x 25,4 x 200 µm3 !!! = 2,6 10-4 mm3 = 0,26 10-9 l < 10-3 Clinical Voxel size (500 x 500 x 2000 µm) TM = 1,5 min SNR = 3,15 MR-microscopic images of Silicon grid sets in Silicon oil (finest: a/2 = 32 µm) Left: sagittal view at increased SNR (32 averages ) Top: enlarged area trace of profile is indicated Bottom: profile indicating modulation depth Right: sagittal view for a short protocol A. Berget al. MR-MICROSCOPY ON A HUMAN 7T-SCANNER; ISMRM/ESMRMB 2010, progrnr. 1048,

  4. MR-microscopy vs. histology: Osteoarthritis MR-microscopy: A. Berg; Samples andreferencing: S. Domayer, J. Hofstätter Univ. klinik f. Orthopädie

  5. Osteoarthritis cartilage-boneinterface: MR-microscopyvshistologicalstaining 1.) Healthycartilage(lateral condyleoftheknee): Homogeneous, thickcartilagelayer • MR-microscopyvshistology: • Intacttissue after measurement • Nodamage due tocutting • Nostainingnecessary • (T2, T1, Diffusivity) • Low effort, • Low preparation - andmeasurement time trabecularnetworkstructure MR-microscopy: TE = 8 ms Voxel: 96x 59 x 240 µm3 A. Berg, A. Potthast,  P. Starewicz, S. Domayer, J. G. Hofstaetter; MR-Microscopy on a human MR-7T Scanner: Methodologyandapplicationto medial osteoarthriticsamplesoftheKnee. ECR 2011 Vienna Histologicstaining: Safranin-O

  6. MR-Microscopyvs. histology osteoarthriticcartilage(medial condyleoftheknee): 8 mm 5 MR-microscopy: Voxelsize = 51 x 51 x 240 µm3 < 10-3clinical MRI voxelsize (500 x 500 x 2000 µm3 ) TSE: 3D-MTX: 256x 256 x 47 TE = 8.2 ms, TR = 2.3 s TM = 16 min Size: 250 µm - 50 µm Small degradationareasbelowthesurface Presumably: areasoffibrillationandlossofproteoglycane A. Berg, A. Potthast,  P. Starewicz, S. Domayer, J. G. Hofstaetter; MR-Microscopy on a human MR-7T Scanner: Methodologyandapplicationto medial osteoarthriticsamplesoftheKnee. ECR 2011 Vienna

  7. MR-MicroscopyvsHistology osteoarthriticcartilage(medial compartmentoftheknee): 4mm Small degradationareasbelowthesurface Presumably: areasoffibrillationandlossofproteoglycane A. Berg, A. Potthast,  P. Starewicz, S. Domayer, J. G. Hofstaetter; MR-Microscopy on a human MR-7T Scanner: Methodologyandapplicationto medial osteoarthriticsamplesoftheKnee. ECR 2011 Vienna

  8. 3D-multislicing 13 mm Pulse sequence: Berg_TSE Voxel: 51 x 51 x 240 µm3 3D-MTX: 256x 256 x 47 TM = 16 min A. Berg, A. Potthast,  P. Starewicz, S. Domayer, J. G. Hofstaetter; MR-Microscopy on a human MR-7T Scanner: Methodologyandapplicationto medial osteoarthriticsamplesoftheKnee. ECR 2011 Vienna