ABL800 FLEX Introduction. RTC, December 2004. Radiometer Medical ApS, Åkandevej 21, DK-2700 Brønshøj, Tel: +45 38 27 38 27, www.radiometer.com. Agenda. Front view Rear view Modular design Parameters Analyzer configurations Specifications Sampler types AutoCheck Screen design.
E N D
ABL800 FLEX Introduction RTC, December 2004 Radiometer Medical ApS, Åkandevej 21, DK-2700 Brønshøj, Tel: +45 38 27 38 27, www.radiometer.com
Agenda • Front view • Rear view • Modular design • Parameters • Analyzer configurations • Specifications • Sampler types • AutoCheck • Screen design
Front view • Aspiration inlet • Barcode reader • Color touch screen • Printer • Solutions (cleaning calibration, rinse and waste containers) • AutoCheck • CD-RW drive • Window to electrode modules 3 8 1 4 7 6 5 5 2
Rear view • Screen, rear • Gas containers • Communication ports • Power switch • Fan 1 2 2 3 5 4
Communication ports • The two USB communication ports provide connection to external devices, e.g. network, keyboard, USB storage device • The two other communication ports provide connection to mouse and keyboard • Furthermore there is a printer port, a network connector, a serial port and a port for an external CRT VG monitor 2 1 3 1USB 2Mouse/keyboard 3Network connections
Modular design • Allows flexibility in parameter availability 1 2 3 4 1Inlet 2pH/BG 3Met/Lyt 4 Oximetry
Parameters • There are 10 electrode slots in the two electrode modules • One electrode slot for the reference electrode • Nine other electrodes available for the following parameters: pH, pCO2, pO2, cK+, cNa+, cCa2+, cCl–, cGlucose, cLactate
Oximetry parameters ctHb = concentration of total hemoglobin sO2 = oxygen saturation % FO2Hb = fraction of oxyhemoglobin FCOHb = fraction of carboxyhemoglobin FHHb = fraction of deoxyhemoglobin FMetHb = fraction of methemoglobin FHbF = fraction of fetal hemoglobin ctBil = concentration of bilirubin
Analyzer configuration BG OXI ctHb/sO2 ctBIL/FHbF LYT MET 835 830 825 820 815 810 805 800 Models ABL835, ABL830 and ABL800 of the ABL800 FLEX analyzers are not available at the time of the release
Specifications – analysis details • Sample volumes • Macromodes 195 L (all parameters) • Micromodes 35-95 L (all parameters) • Measuring time • Macromodes 80 sec., cycle time 150 sec., approx. 24 samples/hr • Micromodes 80-135 sec., cycle time 145-200 sec., approx. 18 samples/hr • Derived parameters • 47 different derived parameters to choose from. See the types and definition in teh ABL800 FLEX Specification sheet
Specifications - communications • Computer specification • Intel Celeron Processor • Windows XP embedded • True multi-tasking system • Intel Celeron processor for DMS software • Dedicated 80386 CPU for wet section • Communication • LIS/HIS communication based on ASTM and HL7 • Network communication – Ethernet (TCP/IP protocol as communication protocol) • True two-way communication
Customizable • Very comprehensive setup programs • More functions than ever can be customized • Examples: operators and passwords, corrective actions, reference and critical ranges etc. • Automatic restart after replacements • On-line help systems • Video tutorials • Help • troubleshooting • Connection to RADIANCE • Especially analyzer remote control module
Sampler types Capillary tube QC with adapter Syringe Expired air Test tube
AutoCheck • Automated QC measurements • Eliminates user-to-user variance - results only reflect analyzer performance • Time-saving • Holds 20 ampoules
Color touch screen • Operation and management of the analyzer are performed via the color touch screen • The screen is divided into three sections • Top section/status bar varies slightly according to the status of the analyzer and setup • Center section contains the main information and interaction area • Bottom section contains navigation buttons and varies slightly according to the screen in question 1 2 3 1 Top section 2 Center section 3 Bottom section
Top section • The status bar describes the current task • E.g. calibration, rinse, aspiration • The time bar is present when the analyzer is performing a task • The parameter bar lists all the measured parameters available and activated • The traffic light gives an overview of various analyzer activities and the traffic light color shows the overall status of the analyzer – see the “02 What is what” presentation for a further description 3 1 2 1 Status bar 2 Parameter bar 3 Traffic light
Center section • The center section is the main information and interaction area • Each screen is identified by an icon and a name that matches the button that gives access to it • The screen may contain text, check buttons, scroll bars, a screen keypad or screen keyboard
Bottom section • The bottom section changes with the different options, depending on the screen in question • In the Ready screen • Access various information, e.g. the current time, next QC, if connected to RADIANCE • The buttons can be selected to give access to e.g. Logon/Log Off, the Menu • General navigation buttons • “Close” – close actual window • “Back” – go one step backwards • Other navigation buttons • Results, ID, messages, print, etc. 1 Quality Control Result screen: 2 3 1 Main menu 2 General navigation buttons 3 Other navigation buttons
Radiometer Training Center, December 2004 Radiometer Medical ApS, Åkandevej 21, DK-2700 Brønshøj, Tel: +45 38 27 38 27, www.radiometer.com