Calibration Values for Dipole Validations at New RF Probe Separation Distance of 1.5cm

This document discusses the need to recalibrate the calibration values for dipole validations at a new RF probe separation distance of 1.5cm. It includes FDTD simulations, updated tables, measurements, and a conclusion on the findings.

Download Presentation

Calibration Values for Dipole Validations at New RF Probe Separation Distance of 1.5cm

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Calibration values for dipole validations at new RF probe separation distance of 1.5cmPINS-C item 5.k Jagadish Nadakuduti EME Research Engineer, Motorola, Inc. C63.19 SC8 WG3 meeting, March 28, 2006

  2. Outline • Background • C63.19-2006 Amendment: Change the RF measurement position from bottom of probe element to the center of element and change the test distance from 10mm to 15mm • FDTD simulations • Updated Table 4.2 and Table D.4 • Measurements • Conclusion

  3. Background: PINS-C item 5.k The RF probe separation distance is 15 mm and the calibration values in table 4.2 were made at 10 mm. These calculations need to be redone at 15 mm distance. 15 mm

  4. FDTD simulations: E- & H-field patterns Update field values from 10 mm to 15 mm distance

  5. FDTD simulations: Results at 10 mm • Created FDTD models as specified in “D.5.1.4 Dipole validation theoretical modeling”. • Comparison of FDTD simulations results to those specified in Table D.4. Table D-4. Results of FDTD modeling for ‘Thick dipoles’ (existing values vs. newsimulation results)

  6. FDTD simulations: Results at 10 mm contd. • CW peak E- and H-field values taken from table D-4 are converted into RMS values. These values are incorporated into Table 4.2. Table 4.2 – Illustrated dipole calculated values (existing and new simulation results) • “D.5.1.4 Dipole validation theoretical modeling” section describes simulations for ‘thick dipoles’ and has no information on modeling parameters for ‘planar dipoles’. The relatively high difference between values specified in table 4.2 and those from new simulations in case of ‘planar dipoles’ might be because of using different simulation parameters.

  7. Measurements:At 10 and 15 mm of probe separation distance with 100 mW of input power • Thick dipoles *898.5 MHz thick dipole is not available in the lab • Planar dipoles

  8. Updated Table D.4 at 15 mm Table D-4. Results of FDTD modeling at 15 mm

  9. Table 4.2 at 10 mm with measurement results Table 4.2 – Illustrated dipole calculated and measured values

  10. Updated Table 4.2 at 15 mm with measurement results Table 4.2 – Illustrated dipole calculated and measured values

  11. Conclusion • ‘Planar dipole’ FDTD simulations need to be repeated with the correct parameters. • The following sections in C63.19-2006 need to be updated with calibration values at 15 mm distance • In Section 4.4: • Table 4.2 • In Annex D: • Table D.4 • Figures D.3 - D.6, D.10 • Equations D.1 & D.2 • Include ‘Planar dipole’ simulation parameters in section ‘D.5.1.4 Dipole validation theoretical modeling’. • All of the above tasks can be completed in a time frame of 1-2 months.

  12. Thank you

  13. Table 4.2 at 10 mm

  14. Table D-4 at 10 mm