The Implications for Higher-Accuracy Absolute Gravity Measurements for NGS and its GRAV-D Project

The Implications for Higher-Accuracy Absolute Gravity Measurements for NGS and its GRAV-D Project. Vicki Childers, Daniel Winester, Mark Eckl, Dru Smith, Daniel Roman National Geodetic Survey. NGS Gravity Program (Pre-GRAV-D). 1976-77. 1980’s. etc.usf.edu. vulcan.wr.usgs.gov.

field
Download Presentation

The Implications for Higher-Accuracy Absolute Gravity Measurements for NGS and its GRAV-D Project

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. The Implications for Higher-Accuracy Absolute Gravity Measurements for NGS and its GRAV-D Project Vicki Childers, Daniel Winester, Mark Eckl, Dru Smith, Daniel Roman National Geodetic Survey

  2. NGS Gravity Program (Pre-GRAV-D) 1976-77 1980’s etc.usf.edu vulcan.wr.usgs.gov For NAVD 88 Orthometric Heights

  3. NGS Gravity Program (Pre-GRAV-D) 1995 to Present Superconducting Gravimeter & FG5 Absolute Gravimeter 1980’s to Present Table Mountain Geophysical Observatory Absolute Gravity Measurements

  4. NGS GRAV-D Project GRAV-D: Gravity for the Redefinition of the American Vertical Datum -> New datum by 2022 • Comprised of two parts: • Gravity field “Snapshot” baseline: Airborne gravity survey of all US-held territories • Temporal geoid change monitored for datum updates

  5. Role of Terrestrial Gravity in GRAV-D • New gravity tie for each airborne survey (absolute – A10) A10 Absolute Gravimeter

  6. Role of Terrestrial Gravity in GRAV-D • New gravity tie for each airborne survey (absolute – A10) • Re-survey problem areas identified by airborne data (relative, absolute – A10) • Monitor long-term geoid change via periodic re-measurement (relative, abs - A10 & FG5) • TMGO Intercomparisons for abs gravimeters • Geoid Slope Validation Surveys: Proof of Concept (Gravity for ortho hgts)

  7. How Would NGS Use a More Accurate Gravimeter? • Long-term monitoring of local or regional temporal geoid change • Replace FG5 (better speed, more portability, indoor and outdoor deployment, more stations per time) and A10 in all work (relative meters too!) • Deployment in less quiet and remote areas • Improved accuracy assessment for FG5s through intercomparisons Assuming….

  8. Assuming…. • Significant improvement to tides and ocean-loading corrections code to have accurate measurements at time intervals of 4 hours or less. • Total uncertainty ascribed to earth tide, ocean loading, and polar tide correctors is > 1 μGal (Technical Protocol for 8th ICAG-2009) • An efficient method of determining vertical gravity gradient

  9. The AOSense Atom Interferometric Absolute Gravimeter Mark Kasevich AOSense, Inc.

  10. 1991 Light-pulse atom interferometer

  11. Kinematic model for sensor operation Falling rock Falling atom • Distances measured in terms of phases (t1), (t2) and (t3) of optical laser field at position where atom interacts with laser beam • Atomic physics processes yield a ~ [(t1)-2(t2)+(t3)] • Determine trajectory curvature with three distance measurements (t1), (t2) and (t3) • For curvature induced by acceleration a, a ~ [(t1) - 2(t2) + (t3)]

  12. Why superb sensors? Gravimeter Laser • Atom = near perfect inertial reference. • Laser/atom interactions register relative motion between atom and sensor case. • Sensor accuracy derives from the exceptional stability of optical wavefronts. Sensor Case Atoms 408-735-9500 AOSense.com Sunnyvale, CA AOSense

  13. AOSense Commercial Compact Gravimeter • Commercial cold atom gravimeter • Noise < 0.7mg/Hz1/2 • 10 mGal resolution • > 12 Hz update rate • Shipped 11/22/10 • First commercial atom optic sensor 408-735-9500 AOSense.com Sunnyvale, CA AOSense

  14. Sensor output (blue) Instrument output (red dashed) model Interferometer fringe 408-735-9500 AOSense.com Sunnyvale, CA AOSense

  15. Next Generation Instrument (in development) • Fieldable • Improved noise performance • Improved accuracy • Improved vibration control 408-735-9500 AOSense.com Sunnyvale, CA AOSense

  16. AOSense, Inc. • Founded in 2004 to develop cold-atom sensors (Brent Young CEO). • Core capability is design, fabrication and testing of navigation and gravimetric sensors based on cold-atom technologies. • Staff of 40 • 20k sq. ft. R&D space (clean rooms, assembly, testing) 408-735-9500 AOSense.com Sunnyvale, CA AOSense