All QIBA PHANTOMS

All QIBA PHANTOMS. C52 Transducer held in ring stand – verasonics system Phantom on turntable – 8 rotation angles 7 transducer locations 3 focal depths (40, 60, 80 mm) SWS estimates: Bandpass filter 100-1000 Hz Ransac algorithm – fov ~ 2 cmx2cm

Download Presentation

All QIBA PHANTOMS

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. All QIBA PHANTOMS • C52 Transducer held in ring stand – verasonics system • Phantom on turntable – 8 rotation angles • 7 transducer locations • 3 focal depths (40, 60, 80 mm) • SWS estimates: • Bandpass filter 100-1000 Hz • Ransac algorithm – fov ~ 2 cmx2cm • Error bars mean +/- 1 std. dev from 56 measures

  2. Results 40 60 80 1786-X 1787-X

  3. Sim Results – Lateral Regression Window Location Window positions Blue 0-12 mm Red 1-12 Green 2-12 Black 4-12 • Conclusion – you must be outside the ROE (near field of shear wave) to get consistent speeds • Also- depth dependence here was within numerical error???