1 / 41

Ageing tests and analysis of organic compounds released from various detector materials

RD-51 WG2 meeting 10 th Dec. 2008. Ageing tests and analysis of organic compounds released from various detector materials. Kari Kurvinen on behalf of. H.Andersson d , T.Andersson d , J.Heino a , J.Huovelin c , K.Kurvinen a, * , R.Lauhakangas a ,

jacquelineg
Download Presentation

Ageing tests and analysis of organic compounds released from various detector materials

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. RD-51 WG2 meeting 10th Dec. 2008 Ageing tests and analysis of organic compounds released from various detector materials Kari Kurvinen on behalf of H.Anderssond, T.Anderssond, J.Heinoa, J.Huovelinc, K.Kurvinena,*, R.Lauhakangasa, S.Nenonend, A.Numminena, J.Ojalaa, R.Oravaa,b, J.Schultzc, H.Sipiläd, O.Vilhuc aHelsinki Institute of Physics, P.O.Box 64, FIN-00014 University of Helsinki, Finland bDepartment of Physical Sciences / Division of High Energy Physics, P.O.Box 64, FIN-00014 University of Helsinki, Finland cObservatory, P.O.Box 14, FIN-00014 University of Helsinki, Finland dMetorex International Oy, P.O.Box 85, FIN-02631 Espoo, Finland based on talks given in NSS 2003 and NSS2004 symposium (see conf.CDs and IEEE Trans. on Nucl. Sci 51 No.5, 2004)

  2. Outline • on wire chamber chemistry • classical methods of outgassing and ageing tests • compound oriented ageing tests • some materials analysed and organiccompounds found • ageing tests with an array of proportional counters

  3. Wire chamber chemistry: sample concentration by a cold trap

  4. Wire chamber chemistry: Avalanche compounds identified * “Potentially very explosive, it may be handled and transferred by low temperature distillation. It should be stored at -25 0C to prevent decomposition and formation of explosive polymers.” (Armitage, J.B. et al., J.Chem.Soc., 1951, 44)

  5. Introduction Standard outgassing test weight loss method • the only information obtained is the mass loss during the heating. • no information about the outgassed substances. • outgassing in the standard room temperature must be extrapolated.

  6. database of harmful materials Introduction Outgas study with an accelerated aging test radiation source gas mixture material under study gaseous radiation detector • monitoring current, gas gain, energy resolution, etc.

  7. Introduction Classical accelerated aging test • benefits • gives definite information of combatibility of the material with the detector • drawbacks • laborious and time consuming method • for each new material and new manufacturer of old materials tests must be repeated. • no information about the aging process itself.

  8. TD tube absorbant inert gas GC/MS TD tube compound X found in analysis gaseous radiation detector detector gas mixture database of harmful compounds Introduction Combined outgas analysis and accelerated aging test SAMPLING ANALYSIS TEST

  9. Introduction Sampling + Analysis + Test -method • weaknesses • even more laborous and time consuming than the traditional one in the beginning. • all compounds cannot be analyzed, some are not even commercially available. • all gas mixtures cannot be analyzed (?) • benefits • when the database is large enough, for introducing new materials only the sampling & analysis have to be performed (fast). • sampling, analysis and test may be done in different sites (labs?). • gas of a working detector may be monitoried (HEP experiments). • gives some information about polymerisation processes in the detector.

  10. Instruments - Sampling Sampling with thermo desorption (TD) method flow adjust 4 U-tubes for samples 4 TD tubes Peltier cooler old GC as an oven • Tenax TA used as an absorbant (range C5-C26) • days/weeks long sampling time possible -> good sensitivity

  11. Instruments - Analysis GC/MS with a TD device thermo desorption device GC/MS

  12. Instruments - Accelerated aging tests Irradiation chamber of 12 proportional counters • gas flow divided into four separate sections for inclusion of different impurities (3 detectors/section) • impurity monitored during the irradiation by regular GC/MS measurements. irradiation by X-ray device (Cu-target) monitoring by 55Fe X-ray source

  13. Results - Outgassing Analysis sampling by Tenax TA 260 min @ 150 ºC polyimide sample 8.9 g

  14. Results - Outgassing Analysis Peak areas in successive measurements

  15. Test setup Accelerated aging test with aromatic solvents P10 with no impurities P10 + Xylene as impurity P10 + Toluene as impurity P10 + Styrene as impurity

  16. Results - Accelerated Aging Test STYRENE added to P-10 gas mixture X-rays: 7 kVp, 190 μA Detector current: < 100 nA/(exposed wire cm) the reference detector

  17. Results - Accelerated Aging Test

  18. Results - Accelerated Aging Test Impurity: TOLUENE Impurity: XYLENE added

  19. Results - Accelerated Aging Test

  20. Results - Accelerated Aging Test

  21. Results - Accelerated Aging Test Aging of a proportional counter with different impurities of toluene in P-10 gas mixture X-rays: 7 kVp, 190 μA Detector current: < 100 nA/(exposed wire cm)

  22. Results - Accelerated Aging Test Aging of a proportional counter with different impurities of benzene in P-10gas mixture

  23. Confirmation of ageing with a different detector and a source relative gas amplification during irradiation by 244Am source a single wire proportional counter filled with P-10 gas mixture containing 250 ppm toluene Gas gain dropping Deposits on wire

  24. Plasma Chemistry H.Yasuda: Plasma Polymerization:

  25. Plasma Chemistry H.Yasuda: Plasma Polymerization:

  26. Plasma Chemistry H.Yasuda: Plasma Polymerization:

  27. Results – Analysis of avalanche compounds

  28. Results - Outgassing Analysis Outgassing analysis of some common detector materials • analyzed: • polyimides (4 different grades) • PET, PEEK, PA • FR4 • soldering tin • epoxies & glues • cable insulating materials • rubbers

  29. Results - Outgassing Analysis sampling by Tenax TA 260 min @ 150 ºC polyimide sample 3.1 g

  30. Results - Outgassing Analysis (Cu-cladding removed by wet etch) sampling by Tenax TA 26.5 h @ 120 ºC polyimide sample 4.6 g

  31. Results - Outgassing Analysis

  32. Results - Outgassing Analysis Signal wire (ETFE insulation) “space quality” by ESA sampling by Tenax TA 260 min @ 150 ºC sample 6.2 g

  33. Results - Outgassing Analysis Nylon tubing sampling by Tenax TA 26.5 h @ 70 ºC sample 12.4 g

  34. Results - Outgassing Analysis 24°C ! sample: 20 g Ar flow: 10 ml/min

  35. Results - Accelerated Aging Test Aging of a proportional counter with two different impurities of cyclohexane in P-10 gas mixture

  36. Results - Accelerated Aging Test Aging of a proportional counters with impurities of metylcyclohexane, benzaldehyde and ethylbenzene in P-10 gas mixture from Kapton from Kapton

  37. Results - Accelerated Aging Test Energy resolution of a proportional counter containing acetal (1,1-diethoxyethane) 1 ppm acetal

  38. Results - Accelerated Aging Test Some non-aromatic common solvents ethanol removed IPA removed aceton removed water removed

  39. Results – Starting point for a database? Outgassing compounds from various materials

  40. Summary • Aromatic compounds are observed to outgas from several detector materials. • All the tested aromatic compounds (benzene, xylene, toluene, styrene, benzaldehyde and ethylbenzene) caused aging in a proportional counter filled with P-10 gas mixture. • The original characteristics of the detector are recovered by irradiation after removing the impurities (except with styrene) • No aging was observed with five non-aromatic solvents (cyclohexane, ethanol, isopropanol, aceton and water).

More Related