Simultaneous determination of inorganic anions and cations by capillary electrophoresis with indirect UV detection I. Haumann, J. Boden, A. Mainka, U. Jegle Journal of Chromatography (2000) 895 , 269-277
Purpose: • Investigate the separation of low molecular mass anions and cations by capillary electrophoresis • Use of indirect UV detection to analyze anions or cations.
Background: • What is capillary electrophoresis (CE) ? • Advantages of CE • High separation efficiency • Small sample size required • Fast separation • Reproducibility • Automation
Electroosmosis is the fundamental process that drives CE. • Fused silica capillaries are used for separations • pI of fussed silica is about 1.5 • Degree of ionization is controlled by the pH of the buffer
Experiment/Instrumentation: • Chemicals used- Dimethyldiphenylphonium (DIPP) and trimesic acid • All solutions, electrolytes and standards were prepared using water purified with a Mill-Q System.
Instruments used- modular composed device combining a high-voltage power supply (P/ACE Capillary Electrophoresis System) • Fused Silica Capillary
Results/Discussion: • The article compares three different principles of simultaneous determination of anions and cations using CE. • Utilization of electroosmotic flow • Electrolyte flow generation by external pressure • Sample infection on both sides of the capillary
Utilization of the electroosmotic flow • Requires a strong electroosmotic flow (EOF) • Sample injection is taken from the side of capillary opposite of the detector • Cations reach the detector first • EOF transports neutral species inside sample zone. • Its necessary to increase the velocity of the EOF
Electrolyte flow generation by external pressure • Separation is carried out under conditions (pH of 6.0) • EOF is not strong enough to transport small anions towards the detector • Analysis of the cations is not possible • Additional flow added to the system • Optimum conditions concerning velocities can be adjusted
Sample injection on both sides of the capillary • Electric field is applied • Cations and anions migrate to the detector from opposite sides • Position of the detector is chosen in regards to the migration velocity • Hydrodynamic injection • Advantages- no strong EOF needed and no additional broadening caused by applying pressure
Selection and optimization of the electrolyte systems • Indirect UV detection is used if the analyte does not absorb in the UV-visible region of the spectrum. • In order to apply indirect UV detection the electrolyte system must contain UV active cations as well as UV active anions.
Imidazole-thiocyanate system • Imidazlole as cationic component • Thiocyanate as anionic component • 18-crown-6 was added to the electrolyte to resolve the ammonium • Citric acid was used to separate the alkali earth ions • To avoid overlapping of signals, slowest cation must leave point of detection before the first anion arrives at the detector. • Signals of the inorganic cations appear first in the electropherogram then less mobile cations, inorganic anions
DIPP-TMA system • Creates a strong EOF • Guarantees no overlapping of anion signals by cationic species if the EOF signal appears in front of the anion peaks • DIPP as cationic UV absorbent • TMA as anionic UV absorbent • HIBA as additive for optimization of the cation separtation
All analytes must be separated from each other and there must be enough space between the EOF signal and first anion peak. • Separation was developed by optimization of three parameters: pH, temperature and capillary length
pH: • Migration time of EOF, charge of the analytes and the electrolyte components depend on pH value • pH<4.5 (anionic analytes show peak broadening and smaller peaks. • Lower effective charge of the electrolyte anion TMA resulting in lower mobility. • Fronting of the anionic analyte peaks becomes stronger.
pH>5.5 • Velocity of EOF becomes higher • Resolution of the cations become lower • Longer migration times 4.8 was the most advantageous pH
Capillary Length: • Distance from EOF signal to anion peaks can be influenced by changing the capillary length between the cathodic end of the capillary and the detector • Migration time of the EOF signal and the first anion should be at least 30 sec to guarantee quality of the separation
Temperature: • 20 degrees Celsius- anion peaks were too small, which lead to overlapping • 25 degrees Celsius- velocity of EOF increases and distance between EOF and anions increases • Varying the temperature of the electrolyte system is a fast and easy method to choose the best conditions for different analytical tasks.
Conclusion: • Simultaneous determination of low molecular mass cation and anions can be carried out with different separation techniques • Most advantageous method utilized the sample injection at both sides of the capillary and opposite migration directions of anions and cations. • The electrolyte system has to be selected with respect to special requirements for separation and the indirect UV detection of both cationic and anionic analytes
Effective separation was developed by optimization of pH, capillary length and temperature