An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Temperature-Related Stability in Plasma Spectrometry Jerry Dulude, Ron Stux, and Vesna Dolic, Glass Expansion, jdulude@geicp.com

  2. Optical drift RF instability Sample related baseline shift Matrix variations Drift in sample intro temperature Profiling & thermostat Feedback loop Background correction Internal std. or matrix matching ???????? Causes of Signal Drift in ICP-OES Cause Solution Glass Expansion www.geicp.com 1

  3. IsoMist™ Programmable Temperature Spray Chamber (PTSC) Glass Expansion www.geicp.com 1

  4. IsoMist Characteristics • Programmable from -10 to 60C in 1 degree increments • Maintains temperature to within 0.1 degree • Built-in Peltier device • No external plumbing • Temperature measured near the chamber surface Glass Expansion www.geicp.com 1

  5. IsoMist on Optima 2100DV Glass Expansion www.geicp.com 1

  6. PE 2100DV Standard Parameters • Power: 1450 watts • Coolant gas: 15LPM • Aux. gas: 0.2LPM • Quartz injector, 1.8mm bore • Neb. Gas: 0.7LPM • SeaSpray concentric glass nebulizer (1ml/min) • IsoMist Programmable Temperature spray chamber with Twister Baffled Cyclonic Glass Expansion www.geicp.com 1

  7. Intensity vs. Temperature-PE 2100DV(1ml/min uptake) Glass Expansion www.geicp.com 1

  8. AVG INTENSITY Temperature Plots on PE 2100DV 31% increase from 10 to 25C or 2% per degree C 31% increase from 25 to 35C or 3% per degree C Glass Expansion www.geicp.com 1

  9. Transport Efficiency AVG INTENSITY Temperature Plots on PE 2100DV Glass Expansion www.geicp.com 1

  10. Transport Efficiency AVG INTENSITY Temperature Plots on PE 2100DV Avg. suppression of all lines Interference std: 1000ppm Na + K, 500ppm Ca Glass Expansion www.geicp.com 1

  11. Transport Efficiency AVG INTENSITY Temperature Plots on PE 2100DV Avg. suppression of all lines Avg. suppression of ion lines Interference std: 1000ppm Na + K, 500ppm Ca Glass Expansion www.geicp.com 1

  12. IsoMist on the Varian Vista MPX Glass Expansion www.geicp.com 1

  13. Varian Vista MPX Standard Parameters • Power: 1300 watts • Coolant gas: 15LPM • Aux. gas: 1.5LPM • Quartz injector, 2.3mm bore • Neb. Gas: 0.8LPM • SeaSpray concentric glass nebulizer (0.4ml/min) • IsoMist Programmable Temperature spray chamber with Twister Baffled Cyclonic Glass Expansion www.geicp.com 1

  14. Effect of Spray Chamber Temperature on Intensity (Varian Vista MPX) Glass Expansion www.geicp.com 1

  15. Effect of Temperature on IntensityVarian Vista MPX 10% increase from 20 to 25C or 2%/degree C Glass Expansion www.geicp.com 1

  16. Effect of Spray Chamber Temperature Varian Vista MPX Transport Efficiency Avg. Intensity Glass Expansion www.geicp.com 1

  17. Varian Vista MPX Parameters • Power: 1300 watts • Coolant gas: 15LPM • Aux. gas: 1.5LPM • Quartz injector, 1.8mm bore • Neb. Gas: 0.8LPM • SeaSpray concentric glass nebulizer (0.4ml/min) • IsoMist Programmable Temperature spray chamber with Twister Baffled Cyclonic Glass Expansion www.geicp.com 1

  18. Temperature vs. Intensity (Varian Vista MPX- 1.8mm bore) Glass Expansion www.geicp.com 1

  19. Temperature vs. Intensity (Varian Vista MPX) 1.8mm bore 2.3mm bore 10% decrease from 20 to 40C or 0.5% per degree C 29% increase from 10 to 40C or 1% per degree C Glass Expansion www.geicp.com 1

  20. Low Sample Uptake Applications • Biological applications • Intracellular fluids • Neonatal samples • Forensic applications • Speciation (HPLC-ICP-MS) • Radioactive wastes reduction • Low DL’s desirable Glass Expansion www.geicp.com 1

  21. Effect of Temperature on Normalized Intensity (20 uL/min uptake rate) PE Optima 2100DV Glass Expansion www.geicp.com 1

  22. Effect of Temperature on Intensity20ul/min uptake rate Avg. Intensity Estimated Transport Efficiency 133% increase from 21 to 40C or 7% per degree C Glass Expansion www.geicp.com 1

  23. Spray Chamber Temperature Impact • High impact with respect to intensity • Interdependent with other plasma parameters • Provides another dimension for optimization • Must remain constant during the experiment Glass Expansion www.geicp.com 1

  24. Constant temperature benefits(1ml/min sample uptake) • Higher accuracy • Higher productivity • Greater reproducibility • Better experiments Glass Expansion www.geicp.com 1