1 / 7

Magnetic Suspension Balance - Layout

Electromagnet. Permanent magnet. Δζ zp. Δζ mp. Measuring load decoupling. Sensor core. Sample. Measuring-point position. Zero-point position. Magnetic Suspension Balance - Layout. Magnetic system. Magnetic Suspension Balance. Position control system. Control system.

mcquaid
Download Presentation

Magnetic Suspension Balance - Layout

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Electromagnet Permanent magnet Δζzp Δζmp Measuring load decoupling Sensor core Sample Measuring-point position Zero-point position Magnetic Suspension Balance - Layout Magnetic system

  2. Magnetic Suspension Balance Position control system Control system Set-point controller PID controller Position trasducer PID Modulates voltage on electromagnet to held the suspension magnet in vertical position

  3. Magnetic Suspension Balance Temperature control system High temperature control: Up to 500 °C Set point Sample temperature Set point resistance Resistance value Cascade control Thermocouple value

  4. Magnetic Suspension Balance Temperature control system • Low temperature control: • Below Room temperature • Up to 100°C Jacket

  5. Sorption equipment Gas inlet : H2 CO N2 burner Gas chromatograph Vacuum pump 8 Thermocouple 9 Pressure transducer ( 150 bar) 10 Pressure transducer ( 1000 mbar)

  6. Example of Sorption Isotherm • PDMS-CO2 at 35°C Kinetics of Sorption Sorption Isotherm Legend: Exp. Data Fick law

  7. Other applications: Thermogravimetric analysis of polymers Study of water loss during the heating of Hyflon-Ion in hydrated state 5 Membrane fully hydrated 30°C 4 3 Weigth loss (%) molH2O/molSO3H 2 Heating 1 Evacuation at 30°C 0 Evacuation at 120°C Time (h) Temperature (°C) Legend: Desorption 30°C Heating up to 120°C

More Related