Polyimide joint under tension in 50% HF

Polyimide joint under tension in 50% HF. Goal: to observe effect of HF solution on PI-E joint under tension Things to look for: Destruction of polyimide tubing Stretching/deforming of joint Wire damaged or breaks or pulls out. HF + tension setup: 100 grams tension

nemo
Download Presentation

Polyimide joint under tension in 50% HF

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Polyimide joint under tension in 50% HF • Goal: to observe effect of HF solution on PI-E joint under tension • Things to look for: • Destruction of polyimide tubing • Stretching/deforming of joint • Wire damaged or breaks or pulls out

  2. HF + tension setup: • 100 grams tension • (not shown: polyethylene cover on HF) camera “tension plate” Wirejoint in pool of 50% HF Clip, pulley, weights Polyethylene block

  3. 76 hours 104 hours 120 hours 126hours 144 hours

  4. Closeup of joint:

  5. Conclusions: • HF attacks glue from ends, also forms bubbles • Wire slips out through softened glue, knot gets tighter • Possibly some stretching/shrinking