1. Precision Xceed Pro First User Training
2. Medicines & Healthcare products Regulatory Agency MHRA state* that training should include :
Basic principles of testing
Demonstrate proper use of system
Consequences of improper use
Instruction in sample collection
Instruction in the importance of complete documentation of all results
Appropriate calibration & QC testing
Practical experience to satisfy instructor that the trainee is competent
*MDA Safety Notice SN 9616 ‘Extra-Laboratory Use of Blood Glucose Meters and Test Strips: Contra-Indications, Training and Advice to the Users’ (1996)
3. Precision Xceed Pro Overview
4. Precision Xceed Pro Blood Glucose Test Strips The Precision Xceed Pro Blood Glucose test strip :
0.6µL sample size
20-second result countdown
Assay Range 1.1 – 27.8 mmol/l
Haematocrit Range 20-70%
Individually foil wrapped test strips
Top or tip sample application
Visual confirmation of sufficient sample
Fill trigger ensures sufficient sample
Claim - Capillary, venous, arterial & neonatal (heel prick)
5. Precision Xceed Pro Meter
6. Precision Xceed Pro Meter – Port Cover The port cover prevents fluid entering the meter.
The meter should NOT be used without the yellow port cover in place.
Meters damaged by fluid in the strip port will be charged for.
7. Precision Xceed Pro Workstation The Workstation is Designed to House the :
Precision Xceed Pro Meter
Precision Xceed Pro Glucose and/or Ketone test strips
MediSense? High & Low Control Solutions
Quick Reference Guide
Single Patient Use Lancing Device & Cotton Pads for sample collection
When not in use, always store the Meter in the Workstation!
Do NOT store used test strips or lancets in the Workstation!
8. The Docking Station is Designed to :
Send data to the Network
Receive latest System Settings
Data transfer only takes approx 10-20 seconds
The Precision Xceed Pro can be set to prompt you when it needs Docking
Several Precision Xceed Pro Meters can use the same Docking Station
After 3 ~ 4 minutes of non use the Meter will turn off automatically ! Precision Xceed Pro Docking Station
9. Performing a Quality Control Test
10. Consequences of not performing QC Tests If QC is NOT performed, misleading results may lead to inappropriate treatment
Always follow your hospital protocol to avoid putting patients at risk
Under MHRA (Medicines and Healthcare products Regulatory Agency) guidelines it is your responsibility to perform/document QC results
The Precision Xceed Pro will prompt when this is due
QC solution is stable for 90 days after opening, so when opening a new bottle write the expiry date on the bottle.
11. Performing Quality Control (QC) Tests
Turn the Precision Xceed Pro on via the ‘On/Off’ button & select option ‘2’ (2-Control Test) via the keypad
When prompted Scan :
Your Operator Barcode
The Low Level solution Barcode
The Strip Lot Barcode
Follow the steps on Screen
Before applying solution, gently invert to mix & prevent air locks
Repeat the procedure for High Level Solution by :
Selecting option ‘1’ (1-Next Level) via the keypad
Following the on-screen prompts
12. Open the strip using the triangular notch in the foil tearing down & across the length of the strip
Ensure the 3 electrodes enter the port facing upward
A sample can be applied to the top or tip of the test strip
Apply a sample with the meter in a horizontal (flat) position to avoid it entering the strip port
Fill trigger electrode ensures a sufficient sample is applied
The result, or ‘PASS’ or ‘FAIL’, is displayed after 20-second countdown Performing Quality Control (QC) Tests
13. Performing Quality Control (QC) Tests If results are out-of-range or ‘FAIL’ :
Repeat test (option ‘2’)
Ensure you follow correct procedure & technique for QC testing
If repeated results are in-range or ‘PASS’ you can proceed
If either repeated result is out-of-range or ‘FAIL’ follow Hospital Protocol for out-of-range Results
If proceeding to a patient test, press the ‘Menu’ key and you will be returned to the main menu
14. External Quality Assessment (EQA)
Where QC is a means of self-checking reliable results are attained, EQA involves the analysis of samples with unknown values from an external source
These will be sent out on a regular basis from the laboratory
Results are then subject to analysis to compare results across different sites
15. Performing a Patient Test
16. Performing a Patient Test Turn the Precision Xceed Pro on via the ‘On/Off’ button & select option ‘1’ (1-Patient Test) via the keypad
When prompted Scan/Enter :
Your Operator Barcode
The Patient ID Barcode/Number
The Strip Lot Barcode
If you have to enter a patient ID number manually using the keypad you will be asked to enter the number correctly twice before being allowed to proceed to the next step
Follow the steps on Screen, ensuring you follow Hospital Protocol for sample collection
17. Performing a Patient Test Open the strip using the triangular notch in the foil tearing down & across the length of the strip
Ensure the 3 electrodes enter the port facing upward
A sample can be applied to the top or tip of the test strip
18. Limitations of Use
Particular care should be taken with those patients who are extremely unwell, particularly if they are :
Severely dehydrated
Severely hypotensive or in shock
In a hyperglycaemic-hyperosmolar state e.g. DKA
In these cases test results may be erroneously low
If in doubt, send a venous sample to the laboratory
19. Action Ranges Action Range
A result falling outside the action range will require some intervention from the operator, such as to repeat the result. A comment code can be scanned/entered into the meter.
For guidance on out of range results, please follow your Hospital protocol.
20. Recalling Patient Results
To recall patient results :
With the Precision Xceed Pro turned on :
Press the ‘Menu’ button
Press button ‘1’ (1-Data Review)
Enter your Operator ID, either via keypad or scan
In Data Review you can choose to view :
All Patient Data, to scroll through all stored patient results
Patient by Patient ID, to scroll through a particular patient’s results
Patient by Operator ID, to scroll through all patient results you have taken
21. Data Upload ‘Docking the Precision Xceed Pro’
Data Upload :
Place the Precision Xceed Pro in the Docking Station every 24 hours, as per Hospital Protocol or when the meter prompts you
Data transfer only takes approx 10-20 seconds
Store the Precision Xceed Pro in its Workstation when data transfer is complete
Several Precision Xceed Pro Meters can use the same Docking Station
After 3 ~ 4 mins of non use the Meter will turn off automatically
22. Health & Safety
Always follow Hospital Protocol for Health & Safety !
Never place used Sharps or Test Strips in the workstation
Read & Understand the ‘Limitation of Use’ section, clearly printed on the insert that comes with every box of test strips
Clean the Precision Xceed Pro & Workstation thoroughly as instructed.
23. Maintenance
24. Maintenance Maintenance of the Precision Xceed Pro System :
Clean the Precision Xceed Pro & Workstation with a damp cloth and a mild detergent. Acceptable cleaning solutions include alcohol and ammonia based cleaners. Bleach is not recommended.
Replace ‘AA’ Alkaline batteries when ‘Low Battery Power’ message is displayed.
Perform QC Test as per Hospital Protocol, when ‘QC Due Now’ or ‘Low, High Control Tests Required’ is displayed.
Dock the Precision Xceed Pro as per Hospital Protocol, when ‘Upload Due Now’ or ‘Upload Required’ is displayed
‘Error nnn’, refer to Trouble shooting guide in Precision Xceed Pro Operator’s Manual.
When not in use, always store the Precision Xceed Pro Meter in it’s Workstation.