Precision Xceed Pro First User Training

Medicines

jana
Download Presentation

Precision Xceed Pro First User Training

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


    1. Precision Xceed Pro First User Training

    2. Medicines & Healthcare products Regulatory Agency MHRA state* that training should include : Basic principles of testing Demonstrate proper use of system Consequences of improper use Instruction in sample collection Instruction in the importance of complete documentation of all results Appropriate calibration & QC testing Practical experience to satisfy instructor that the trainee is competent *MDA Safety Notice SN 9616 ‘Extra-Laboratory Use of Blood Glucose Meters and Test Strips: Contra-Indications, Training and Advice to the Users’ (1996)

    3. Precision Xceed Pro Overview

    4. Precision Xceed Pro Blood Glucose Test Strips The Precision Xceed Pro Blood Glucose test strip : 0.6µL sample size 20-second result countdown Assay Range 1.1 – 27.8 mmol/l Haematocrit Range 20-70% Individually foil wrapped test strips Top or tip sample application Visual confirmation of sufficient sample Fill trigger ensures sufficient sample Claim - Capillary, venous, arterial & neonatal (heel prick)

    5. Precision Xceed Pro Meter

    6. Precision Xceed Pro Meter – Port Cover The port cover prevents fluid entering the meter. The meter should NOT be used without the yellow port cover in place. Meters damaged by fluid in the strip port will be charged for.

    7. Precision Xceed Pro Workstation The Workstation is Designed to House the : Precision Xceed Pro Meter Precision Xceed Pro Glucose and/or Ketone test strips MediSense? High & Low Control Solutions Quick Reference Guide Single Patient Use Lancing Device & Cotton Pads for sample collection When not in use, always store the Meter in the Workstation! Do NOT store used test strips or lancets in the Workstation!

    8. The Docking Station is Designed to : Send data to the Network Receive latest System Settings Data transfer only takes approx 10-20 seconds The Precision Xceed Pro can be set to prompt you when it needs Docking Several Precision Xceed Pro Meters can use the same Docking Station After 3 ~ 4 minutes of non use the Meter will turn off automatically ! Precision Xceed Pro Docking Station

    9. Performing a Quality Control Test

    10. Consequences of not performing QC Tests If QC is NOT performed, misleading results may lead to inappropriate treatment Always follow your hospital protocol to avoid putting patients at risk Under MHRA (Medicines and Healthcare products Regulatory Agency) guidelines it is your responsibility to perform/document QC results The Precision Xceed Pro will prompt when this is due QC solution is stable for 90 days after opening, so when opening a new bottle write the expiry date on the bottle.

    11. Performing Quality Control (QC) Tests Turn the Precision Xceed Pro on via the ‘On/Off’ button & select option ‘2’ (2-Control Test) via the keypad When prompted Scan : Your Operator Barcode The Low Level solution Barcode The Strip Lot Barcode Follow the steps on Screen Before applying solution, gently invert to mix & prevent air locks Repeat the procedure for High Level Solution by : Selecting option ‘1’ (1-Next Level) via the keypad Following the on-screen prompts

    12. Open the strip using the triangular notch in the foil tearing down & across the length of the strip Ensure the 3 electrodes enter the port facing upward A sample can be applied to the top or tip of the test strip Apply a sample with the meter in a horizontal (flat) position to avoid it entering the strip port Fill trigger electrode ensures a sufficient sample is applied The result, or ‘PASS’ or ‘FAIL’, is displayed after 20-second countdown Performing Quality Control (QC) Tests

    13. Performing Quality Control (QC) Tests If results are out-of-range or ‘FAIL’ : Repeat test (option ‘2’) Ensure you follow correct procedure & technique for QC testing If repeated results are in-range or ‘PASS’ you can proceed If either repeated result is out-of-range or ‘FAIL’ follow Hospital Protocol for out-of-range Results If proceeding to a patient test, press the ‘Menu’ key and you will be returned to the main menu

    14. External Quality Assessment (EQA) Where QC is a means of self-checking reliable results are attained, EQA involves the analysis of samples with unknown values from an external source These will be sent out on a regular basis from the laboratory Results are then subject to analysis to compare results across different sites

    15. Performing a Patient Test

    16. Performing a Patient Test Turn the Precision Xceed Pro on via the ‘On/Off’ button & select option ‘1’ (1-Patient Test) via the keypad When prompted Scan/Enter : Your Operator Barcode The Patient ID Barcode/Number The Strip Lot Barcode If you have to enter a patient ID number manually using the keypad you will be asked to enter the number correctly twice before being allowed to proceed to the next step Follow the steps on Screen, ensuring you follow Hospital Protocol for sample collection

    17. Performing a Patient Test Open the strip using the triangular notch in the foil tearing down & across the length of the strip Ensure the 3 electrodes enter the port facing upward A sample can be applied to the top or tip of the test strip

    18. Limitations of Use Particular care should be taken with those patients who are extremely unwell, particularly if they are : Severely dehydrated Severely hypotensive or in shock In a hyperglycaemic-hyperosmolar state e.g. DKA In these cases test results may be erroneously low If in doubt, send a venous sample to the laboratory

    19. Action Ranges Action Range A result falling outside the action range will require some intervention from the operator, such as to repeat the result. A comment code can be scanned/entered into the meter. For guidance on out of range results, please follow your Hospital protocol.

    20. Recalling Patient Results To recall patient results : With the Precision Xceed Pro turned on : Press the ‘Menu’ button Press button ‘1’ (1-Data Review) Enter your Operator ID, either via keypad or scan In Data Review you can choose to view : All Patient Data, to scroll through all stored patient results Patient by Patient ID, to scroll through a particular patient’s results Patient by Operator ID, to scroll through all patient results you have taken

    21. Data Upload ‘Docking the Precision Xceed Pro’ Data Upload : Place the Precision Xceed Pro in the Docking Station every 24 hours, as per Hospital Protocol or when the meter prompts you Data transfer only takes approx 10-20 seconds Store the Precision Xceed Pro in its Workstation when data transfer is complete Several Precision Xceed Pro Meters can use the same Docking Station After 3 ~ 4 mins of non use the Meter will turn off automatically

    22. Health & Safety Always follow Hospital Protocol for Health & Safety ! Never place used Sharps or Test Strips in the workstation Read & Understand the ‘Limitation of Use’ section, clearly printed on the insert that comes with every box of test strips Clean the Precision Xceed Pro & Workstation thoroughly as instructed.

    23. Maintenance

    24. Maintenance Maintenance of the Precision Xceed Pro System : Clean the Precision Xceed Pro & Workstation with a damp cloth and a mild detergent. Acceptable cleaning solutions include alcohol and ammonia based cleaners. Bleach is not recommended. Replace ‘AA’ Alkaline batteries when ‘Low Battery Power’ message is displayed. Perform QC Test as per Hospital Protocol, when ‘QC Due Now’ or ‘Low, High Control Tests Required’ is displayed. Dock the Precision Xceed Pro as per Hospital Protocol, when ‘Upload Due Now’ or ‘Upload Required’ is displayed ‘Error nnn’, refer to Trouble shooting guide in Precision Xceed Pro Operator’s Manual. When not in use, always store the Precision Xceed Pro Meter in it’s Workstation.