An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. YOĞUN BAKIM ÜNİTELERİNDEKİ YENİDOĞAN BEBEKLERİN RADYASYON DOZU Turan OLĞAR Ankara Üniversitesi Nükleer Bilimler Enstitüsü

  2. HASTA DOZU ÖLÇÜM YÖNTEMLERİ • Retrospektif Ölçüm Yöntem • Termolüminesans Dozimetre (TLD)

  3. RADYOLOJİK İNCELEME • AKCİĞER • KARIN MOBİL X-IŞIN SİSTEMLERİ • 2 GE VMX Plus • 1 GE AMX-4

  4. TÜP ÇIKIŞ DOZUNUN ÖLÇÜLMESİ Odak-Dedektör Mesafesi İyon Odası X-ışın tüpünden çıkan radyasyon;kVp, mA, ışınlama zamanınınfonksiyonu olarak ölçülmelidir.

  5. Tüp Voltajı (kVp) mAs Havadaki Doz (µGy) µGy/mAs (µGy/mAs)ort 50 mAs1 mAs2 mAs3 55 mAs1 mAs2 mAs3

  6. HASTA GİRİŞ DOZUNUN HAVAKERMA’DAN BULUNMASI OHM ODM İyon Odası ESD = HK x GSF x (ODM / OHM)2 x μd / μh HK : Hava Kerma (mGy) (ODM / OHM)2 : Ters Kare düzeltme faktörü μd / μh = 1.06 (Doku ve hava kütle azalım katsayıları oranı)

  7. HASTA ÜZERİNDE CİLT DOZUNUN SAPTANMASI TLD

  8. TLD KALİBRASYONU 1- Her fosforun farklı hassasiyette olması Tüm TLD’ler aynı miktarda ışınlanır, ortalama değerden belirli bir oranda sapanlar çalışmaya dahil edilmezler 2- Foto tüp akımının ışınlama ya da doz birimine çevrilmesi gerekir Belirli sayıda TLD bir iyon odası ile beraber ışınlanarak akım-doz (nC-mGy) dönüşümü bulunmalıdır

  9. APRUPA BİRLİĞİ KOMİSYONU TARAFINDAN ÖNERİLEN REFERANS DEĞERLER Ekran/Film Kombinasyonu Hızı Ek Filtrasyon kV Odak-Film Mesafesi (cm) Radyografi Başına Referans Doz Değeri (μGy) 200-400 1mmAl+0.1Cu yada 0,2mm Cu 60-65 80-100 80

  10. İnceleme Ortalama Radyografi Sayısı Ortalama kVp Ortalama mAs Radyografi Başına Ortalama Giriş Cilt Dozu (Gy) Radyografi Başına Ortalama Etkin Doz (µSv) Göğüs 14 (2-31) 49 (46-51) 1.9 (1.6-3.5) 67 (44-136) 15 (10-31) Karın 5 (1-24) 48 (45-51) 2 (1.6-2.5) 65 (25-197) 22 (7-63) BEBEK DOZU ÖLÇÜM SONUÇLARI

  11. GİRİŞ CİLT DOZU (Tüp Çıkış - TLD) Karın + Göğüs Radyografi Başına ESDTLD= 70 µGy Ortalama Gebelik Yaşı 28 hafta Klinikte Ortalama Tedavi Süresi 50 gün

  12. Referans İnceleme kVp/mAs Radyografi Başına Ortalama ESD (µGy) Radyografi Başına Ortalama E (µSv) Radyografi Başına Risk (x106) (E x 13x10-2) Bu Çalışma Göğüs 49/1.9 67 15 2 Karın 48/2.0 65 22 2.9 Brindhaban and Al-Khalifah Göğüs 57/1.6 60 26 - Karın 57/1.6 60 32 - Jones et. al. Göğüs 62/2.0 56.7 15.4 2 Karın 62/2.5 73.6 21.9 2.8 Armpilia et. al. Göğüs 53.1/2 36 7.8 1 Karın 53/2.0 39 10.2 1.3 Robinson and Dellagrammaticas Göğüs 60/1.0 53 - - Karın 60/1.0 57 - - Schneider et. al. Göğüs - 78 - - Lowe et. al. Göğüs - 16-160 - - Chapple et. al. Göğüs - 55 - - LİTERATÜR KARŞILAŞTIRMASI

  13. KAYNAKLAR • Armpilia CI, Fife IAJ, Croasdale PL. Radiation dose quantities and risk in neonates in a special care baby unit. Br J Radiol 2002; 75: 590-595 • Chapple CL, Faulkner K, Hunter EW. Energy imparted to neonates during X-ray examinations in a special care baby unit. Br J Radiol 1994; 67: 366-370 • European commission. European guidelines on quality criteria for diagnostic radiographic images in paediatrics. Report EUR 16261 EN. Brussels: CEC, 1996. • Hart D, Wall BF, Schrimpton PC, Bungay DR, Dance DR. Reference doses and patient size in paediatric radiology. NRPB R318. Chilton: HMSO, 2000 • Duggan L., Warren-Forward H, Smith T and Kron T. Investigation of dose reduction in neonatal radiography using specially designed phantoms and LiF:Mg,Cu,P TLDs.Br J Radiol 2003; 76: 232-237 • Brindhaban A and Al-Khalifah K. Radiation dose to premature infants in neonatal intensive care units in Kuwait Radiat Prot Dosimetry 2004; 111: 275-281 • Jones NF, Palarm TW and Negus IS. Neonatal chest and abdominal radiation dosimetry: a comparison of two radiographic techniques. Br J Radiol 2001; 74: 920-925 • Robinson A and Dellagrammaticas HD. Radiation doses to neonates requiring intensive care. Br J Radiol 1983; 56: 397-400 • Schneider K, Kohn MM and Ernst G. The Derivation of Reference Dose Values to Chest X Rays in Paediatric Radiography. Radiat Prot Dosimetry 1998; 80: 199 - 202 • Lowe A, Finch A, Boniface D, Chaudhuri R and Shekhda J. Diagnotic image quality of mobile neonatal chest X-rays and the radiation exposure incurred. Br J Radiol 1999; 72: 55-61

  14. TEŞEKKÜRLER